Potential value of automated daily screening of cardiac resynchronization therapy defibrillator diagnostics for prediction of major cardiovascular events: results from Home-CARE (Home Monitoring in Cardiac Resynchronization Therapy) study

被引:38
作者
Sack, Stefan [1 ]
Wende, Christian Michael [2 ]
Naegele, Herbert [3 ]
Katz, Amos [4 ]
Bauer, Wolfgang Rudolf [5 ]
Barr, Craig Scott [6 ]
Malinowski, Klaus [7 ]
Schwacke, Harald [8 ]
Leyva, Francisco [9 ]
Proff, Jochen [10 ]
Berdyshev, Sergey [10 ]
Paul, Vincent [11 ]
机构
[1] Univ Klinikum Essen, Dept Cardiol, Essen, Germany
[2] St Elisabeth Klin, Saarlouis, Germany
[3] St Adolf Stift, Reinbek, Germany
[4] Barzilai Govt Hosp, Dept Cardiol, Ashqelon, Israel
[5] Univ Klinikum Wurzburg, Dept Cardiol, Wurzburg, Germany
[6] Russels Hall Hosp, Dept Cardiol, Dudley, England
[7] Helios Klinikum Aue GmbH, Dept Cardiol, Aue, Germany
[8] Allgemeines Krankenhaus Altona, Innere Abt, Hamburg, Germany
[9] Good Hope Hosp, Sutton Coldfield, England
[10] Biotron SE & Co KG, Berlin, Germany
[11] St Peters Hosp St Georges, Dept Cardiol, Chertsey, Surrey, England
关键词
Remote device monitoring; Cardiac resynchronization therapy defibrillator; Multiparameter predictor; Cardiovascular hospitalizations; Heart failure; Home monitoring; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; HEART-FAILURE PATIENTS; INTRATHORACIC IMPEDANCE; DEVICE DIAGNOSTICS; IDENTIFY PATIENTS; INCREASED RISK; FOLLOW-UP; MORTALITY; HOSPITALIZATIONS; MANAGEMENT;
D O I
10.1093/eurjhf/hfr089
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim To investigate whether diagnostic data from implanted cardiac resynchronization therapy defibrillators (CRT-Ds) retrieved automatically at 24 h intervals via a Home Monitoring function can enable dynamic prediction of cardiovascular hospitalization and death. Methods and results Three hundred and seventy-seven heart failure patients received CRT-Ds with Home Monitoring option. Data on all deaths and hospitalizations due to cardiovascular reasons and Home Monitoring data were collected prospectively during 1-year follow-up to develop a predictive algorithm with a predefined specificity of 99.5%. Seven parameters were included in the algorithm: mean heart rate over 24 h, heart rate at rest, patient activity, frequency of ventricular extrasystoles, atrial-atrial intervals (heart rate variability), right ventricular pacing impedance, and painless shock impedance. The algorithm was developed using a 25-day monitoring window ending 3 days before hospitalization or death. While the retrospective sensitivities of the individual parameters ranged from 23.6 to 50.0%, the combination of all parameters was 65.4% sensitive in detecting cardiovascular hospitalizations and deaths with 99.5% specificity (corresponding to 1.83 false-positive detections per patient-year of follow-up). The estimated relative risk of an event was 7.15-fold higher after a positive predictor finding than after a negative predictor finding. Conclusion We developed an automated algorithm for dynamic prediction of cardiovascular events in patients treated with CRTD devices capable of daily transmission of their diagnostic data via Home Monitoring. This tool may increase patients' quality of life and reduce morbidity, mortality, and health economic burden, it now warrants prospective studies.
引用
收藏
页码:1019 / 1027
页数:9
相关论文
共 37 条
[1]  
Abraham WT, 2006, REV CARDIOVASC MED, V7, pS33
[2]   Continuous autonomic assessment in patients with symptomatic heart failure - Prognostic value of heart rate variability measured by an implanted cardiac resynchronization device [J].
Adamson, PB ;
Smith, AL ;
Abraham, WT ;
Kleckner, KJ ;
Stadler, RW ;
Shih, A ;
Rhodes, MM .
CIRCULATION, 2004, 110 (16) :2389-2394
[3]   Ongoing right ventricular hemodynarnics in heart failure -: Clinical value of measurements derived from an implantable monitoring system [J].
Adamson, PB ;
Magalski, A ;
Braunschweig, F ;
Böhm, M ;
Reynolds, D ;
Steinhaus, D ;
Luby, A ;
Linde, C ;
Ryden, L ;
Cremers, B ;
Takle, T ;
Bennett, T .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (04) :565-571
[4]   Telecardiology and remote monitoring of implanted electrical devices: The potential for fresh clinical care perspectives [J].
Boriani, Giuseppe ;
Diemberger, Igor ;
Martignani, Cristian ;
Biffi, Mauro ;
Valzania, Cinzia ;
Bertini, Matteo ;
Domenichini, Giulia ;
Saporito, Davide ;
Ziacchi, Matteo ;
Branzi, Angelo .
JOURNAL OF GENERAL INTERNAL MEDICINE, 2008, 23 (Suppl 1) :73-77
[5]   Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure - The COMPASS-HF study [J].
Bourge, Robert C. ;
Abraham, William T. ;
Adamson, Philip B. ;
Aaron, Mark F. ;
Aranda, Juan M., Jr. ;
Magalski, Anthony ;
Zile, Michael R. ;
Smith, Andrew L. ;
Smart, Frank W. ;
O'Shaughnessy, Mark A. ;
Jessup, Mariell L. ;
Sparks, Brandon ;
Naftel, David L. ;
Stevenson, Lynne Warner .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (11) :1073-1079
[6]   Incidence and prognostic significance of sustained ventricular tachycardias in heart failure patients implanted with biventricular pacemakers without a back-up defibrillator: results from the prospective, multicentre, Mona Lisa cohort study [J].
Boveda, Serge ;
Marijon, Eloi ;
Jacob, Sophie ;
Defaye, Pascal ;
Winter, Jobst B. ;
Bulava, Alan ;
Gras, Daniel ;
Albenque, Jean Paul ;
Combes, Nicolas ;
Pavin, Dominique ;
Delarche, Nicolas ;
Teubl, Alexander ;
Lambiez, Marie ;
Chevalier, Philippe .
EUROPEAN HEART JOURNAL, 2009, 30 (10) :1237-1244
[7]   Can monitoring of intrathoracic impedance reduce morbidity and mortality in patients with chronic heart failure? Rationale and design of the Diagnostic Outcome Trial in Heart Failure (DOT-HF) [J].
Braunschweig, Frieder ;
Ford, Ian ;
Conraads, Viviane ;
Cowie, Martin R. ;
Jondeau, Guillaume ;
Kautzner, Josef ;
Lunati, Maurizio ;
Aguilera, Roberto Munoz ;
Yu, Cheuk Man ;
Marijianowskii, Monique ;
Borggrefe, Martin ;
van Veldhuisen, Dirk J. .
EUROPEAN JOURNAL OF HEART FAILURE, 2008, 10 (09) :907-916
[8]   Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure [J].
Bristow, MR ;
Saxon, LA ;
Boehmer, J ;
Krueger, S ;
Kass, DA ;
De Marco, T ;
Carson, P ;
DiCarlo, L ;
DeMets, D ;
White, BG ;
DeVries, DW ;
Feldman, AM .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (21) :2140-2150
[9]   Prevalence and prognostic impact of comorbidities in heart failure patients with implantable cardioverter-defibrillator [J].
Bruch, Christian ;
Bruch, Chahrebanu ;
Sindermann, Juergen ;
Breithardt, Guenter ;
Gradaus, Rainer .
EUROPACE, 2007, 9 (08) :681-686
[10]   Remote monitoring and follow-up of pacemakers and implantable cardioverter defibrillators [J].
Burri, Haran ;
Senouf, David .
EUROPACE, 2009, 11 (06) :701-709