In this article, a novel x-ray image intensifier is introduced. It is mainly composed of an x-ray intensifying screen and a low-light-level (L-3) image intensifier. In order to obtain a bright enough image, the spectral compatibility among three different combinations of the x-ray intensifying screen and the photocathode is analyzed. The comparison indicates that a (Zn,Cd)S:Ag screen is superior to either a CaWO4 screen or a Gd2O2S:Tb screen when combined with the Super S-25 photocathode and spectral compatibility analysis is a useful guide when designing an optoelectronic imaging device. (c) 2006 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Novel combined X-ray image intensifier
    Yu, Chun-Yu
    Chang, Ben-Kang
    Fu, Rong-Guo
    Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis, 2007, 27 (03): : 610 - 613
  • [2] The novel combined X-ray image intensifier
    Yu Chun-yu
    Chang Ben-kang
    Fu Rong-guo
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27 (03) : 610 - 613
  • [3] X-ray image intensifier
    Balter, S
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 1999, 46 (02) : 238 - 244
  • [4] X-RAY IMAGE INTENSIFIER
    不详
    BMJ-BRITISH MEDICAL JOURNAL, 1954, 2 (4895): : 1044 - 1044
  • [5] X-RAY IMAGE INTENSIFIER
    MCDONALD, RE
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1978, 23 (05): : 691 - 691
  • [6] NOVEL PROXIMITY X-RAY IMAGE INTENSIFIER TUBE
    WANG, SP
    ROBBINS, CD
    BATES, CW
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 130 (02) : 384 - 384
  • [7] A novel x-ray intensifier
    Yu, CY
    Liu, TB
    Wu, LL
    Chang, BK
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS: DIAGNOSTICS AND TREATMENT II , PTS 1 AND 2, 2005, 5630 : 141 - 145
  • [8] MEASUREMENT OF X-RAY IMAGE INTENSIFIER SHARPNESS IN THE X-RAY DEPARTMENT
    LEHERON, JC
    POLETTI, JL
    PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (01): : 93 - 104
  • [9] The novel combined x-ray image intensifier and its imaging performance
    Yu Chunyu
    Chang Benkang
    Fu Rongguo
    Yao Xiao
    OPTOELETRONIC MATERIALS AND DEVICES, PTS 1 AND 2, 2006, 6352
  • [10] Novel algorithm for distortion correction of image intensifier in X-ray angiography system
    Li, Yingchao
    Wang, Yongtian
    Yang, Jian
    Li, Yanhui
    BMEI 2008: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOL 2, 2008, : 203 - +