Variability of the pullout strength of cancellous bone screws with cement augmentation

被引:19
作者
Procter, P. [1 ,2 ]
Bennani, P. [2 ]
Brown, C. J. [2 ]
Arnoldi, J. [3 ]
Pioletti, D. P. [4 ]
Larsson, S. [5 ]
机构
[1] Stryker Osteosynth, Selzach, Switzerland
[2] Brunel Univ, Sch Engn & Design, Uxbridge UB8 3PH, Middx, England
[3] Univ Hosp Bern, Clin Plast & Hand Surg, CH-3010 Bern, Switzerland
[4] Ecole Polytech Fed Lausanne, Lab Biomech Orthoped, CH-1015 Lausanne, Switzerland
[5] Univ Uppsala Hosp, Dept Orthopaed, Uppsala, Sweden
关键词
Cancellous bone; Bone screws; Pullout testing; Human cadaver; Finite element modelling; CALCIUM-PHOSPHATE CEMENT; FRACTURE FIXATION; HIP SCREW; STABILITY; FAILURE; TENSILE;
D O I
10.1016/j.clinbiomech.2015.03.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Orthopaedic surgeons often face clinical situations where improved screw holding power in cancellous bone is needed. Injectable calcium phosphate cements are one option to enhance fixation. Methods: Paired screw pullout tests were undertaken in which human cadaver bone was augmented with calcium phosphate cement. A finite element model was used to investigate sensitivity to screw positional placement. Findings: Statistical analysis of the data concluded that the pullout strength was generally increased by cement augmentation in the in vitro human cadaver tests. However, when comparing the individual paired samples there were surprising results with lower strength than anticipated after augmentation, in apparent contradiction to the generally expected conclusion. Investigation using the finite element model showed that these strength reductions could be accounted for by small screw positional changes. A change of 0.5 mm might result in predicted pullout force changes of up to 28%. Interpretation: Small changes in screw position might lead to significant changes in pullout strength sufficient to explain the lower than expected individual pullout values in augmented cancellous bone. Consequently whilst the addition of cement at a position of low strength would increase the pullout strength at that point it might not reach the pullout strength of the un-augmented paired test site. However, the overall effect of cement augmentation produces a significant improvement at whatever point in the bone the screw is placed. The use of polymeric bone-substitute materials for tests may not reveal the natural variation encountered in tests using real bone structures. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:500 / 506
页数:7
相关论文
共 45 条
[1]   Odontoid fractures: high complication rate associated with anterior screw fixation in the elderly [J].
Andersson, S ;
Rodrigues, M ;
Olerud, C .
EUROPEAN SPINE JOURNAL, 2000, 9 (01) :56-59
[2]   Use of a synthetic bone void filler to augment screws in osteopenic ankle fracture fixation [J].
Andreassen, GS ;
Hoiness, PR ;
Skraamm, I ;
Granlund, O ;
Engebretsen, L .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2004, 124 (03) :161-165
[3]   The influence of cracks, notches and holes on the tensile strength of cellular solids [J].
Andrews, EW ;
Gibson, LJ .
ACTA MATERIALIA, 2001, 49 (15) :2975-2979
[4]  
[Anonymous], 2007, F54307E1 ASTM
[5]   Cancellous bone screw thread design and holding power [J].
Asnis, SE ;
Ernberg, JJ ;
Bostrom, MPG ;
Wright, TM ;
Harrington, RM ;
Tencer, A ;
Peterson, M .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (07) :462-469
[6]   A modified hip screw incorporating injected cement for the fixation of osteoporotic trochanteric fractures [J].
Augat, P ;
Rapp, S ;
Claes, L .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (05) :311-316
[7]   Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue [J].
Bayraktar, HH ;
Morgan, EF ;
Niebur, GL ;
Morris, GE ;
Wong, EK ;
Keaveny, TM .
JOURNAL OF BIOMECHANICS, 2004, 37 (01) :27-35
[8]   Dens fractures in the elderly - Results of anterior screw fixation in 19 elderly patients [J].
Berlemann, U ;
Schwarzenbach, O .
ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (04) :319-324
[9]   An approximate model for cancellous bone screw fixation [J].
Brown, C. J. ;
Sinclair, R. A. ;
Day, A. ;
Hess, B. ;
Procter, P. .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2013, 16 (04) :443-450
[10]   Mechanical performance of standard and cannulated 4.0-mm cancellous bone screws [J].
Brown, GA ;
McCarthy, T ;
Bourgeault, CA ;
Callahan, DJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2000, 18 (02) :307-312