The effect of breath termination criterion on breathing patterns and the work of breathing during pressure support ventilation

被引:5
作者
Tokioka, H [1 ]
Tanaka, T [1 ]
Ishizu, T [1 ]
Fukushima, T [1 ]
Iwaki, T [1 ]
Nakamura, Y [1 ]
Kosogabe, Y [1 ]
机构
[1] Okayama Rosai Hosp, Dept Anesthesiol, Okayama 7028055, Japan
关键词
D O I
暂无
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
With pressure support ventilation (PSV), each PSV breath is flow-cycled, and the breath termination criterion (TC) is usually nonadjustable. When TC does not match the interaction between the patient's inspiratory-expiratory efforts to the opening and closing of the inspiratory and expiratory valves, patient-ventilator asynchrony may occur; and the work of breathing (WOB) may increase. Therefore, we studied the effect of TC on breathing patterns and WOE during PSV in eight patients with acute respiratory distress syndrome or acute lung injury. We studied five levels of TC during PSV-1%, 5%, 20%, 35%, and 45% of the peak inspiratory flow. With increasing levels of TC, the tidal volume decreased and respiratory frequency increased, along with a decrease in duty cycle. WOE markedly increased with increasing levels of TC from 0.31 +/- 0.12 J/L with TC 1% to 0.51 +/- 0.11 J/L with TC 45%. Premature termination with double breathing occurred in one patient with TC 35% and four patients with TC 45%. Delayed termination with a duty cycle of >0.5 occurred in two patients with TC 1%. In conclusion, the proper adjustment of TC improves patient-ventilator synchrony and decreases WOE during PSV.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 18 条
[1]  
AGOSTINI E, 1970, RESPIRATORY MUSCLES, P115
[2]  
BAYDUR A, 1982, AM REV RESPIR DIS, V126, P788
[3]   REPORT OF THE AMERICAN-EUROPEAN CONSENSUS CONFERENCE ON ARDS - DEFINITIONS, MECHANISMS, RELEVANT OUTCOMES AND CLINICAL-TRIAL COORDINATION [J].
BERNARD, GR ;
ARTIGAS, A ;
BRIGHAM, KL ;
CARLET, J ;
FALKE, K ;
HUDSON, L ;
LAMY, M ;
LEGALL, JR ;
MORRIS, A ;
SPRAGG, R ;
DHAINAUT, JF ;
MATTHAY, M ;
MANCEBO, J ;
MEYRICK, B ;
PAYEN, D ;
PERRET, C ;
FOWLER, AA ;
SCHALLER, MD ;
VANASBECK, BS ;
COCHIN, B ;
LANKEN, PN ;
LEEPER, KV ;
MARINI, J ;
MURRAY, JF ;
OPPENHEIMER, L ;
PESENTI, A ;
REID, L ;
RINALDO, J ;
VILLAR, J ;
Hyers, T ;
Knaus, W ;
Matthay, R ;
Pinsky, M ;
Bone, RC ;
Bosken, C ;
Johanson, WG ;
Lewandowski, K ;
Repine, J ;
Rodriguez-Roisin, R ;
Roussos, C .
INTENSIVE CARE MEDICINE, 1994, 20 (03) :225-232
[4]   Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients [J].
Bonmarchand, G ;
Chevron, V ;
Ménard, JF ;
Girault, C ;
Moritz-Berthelot, F ;
Pasquis, P ;
Leroy, J .
CRITICAL CARE MEDICINE, 1999, 27 (04) :715-722
[5]   Increased initial flow rate reduces inspiratory work of breathing during pressure support ventilation in patients with exacerbation of chronic obstructive pulmonary disease [J].
Bonmarchand, G ;
Chevron, V ;
Chopin, C ;
Jusserand, D ;
Girault, C ;
Moritz, F ;
Leroy, J ;
Pasquis, P .
INTENSIVE CARE MEDICINE, 1996, 22 (11) :1147-1154
[6]   Pressure-support ventilation: Still a simple mode? [J].
Brochard, L .
INTENSIVE CARE MEDICINE, 1996, 22 (11) :1137-1138
[7]   INSPIRATORY PRESSURE SUPPORT PREVENTS DIAPHRAGMATIC FATIGUE DURING WEANING FROM MECHANICAL VENTILATION [J].
BROCHARD, L ;
HARF, A ;
LORINO, H ;
LEMAIRE, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (02) :513-521
[8]   AN ANALYSIS OF DESYNCHRONIZATION BETWEEN THE SPONTANEOUSLY BREATHING PATIENT AND VENTILATOR DURING INSPIRATORY PRESSURE SUPPORT [J].
FABRY, B ;
GUTTMANN, J ;
EBERHARD, L ;
BAUER, T ;
HABERTHUR, C ;
WOLFF, G .
CHEST, 1995, 107 (05) :1387-1394
[9]  
Kacmarek RM, 1994, RESP CARE, V39, P881
[10]   THE NAGOYA CONFERENCE ON SYSTEM-DESIGN AND PATIENT-VENTILATOR INTERACTIONS DURING PRESSURE SUPPORT VENTILATION [J].
MACINTYRE, N ;
NISHIMURA, M ;
USADA, Y ;
TOKIOKA, H ;
TAKEZAWA, J ;
SHIMADA, Y .
CHEST, 1990, 97 (06) :1463-1466