A Comparison of the Immunochemical Methods, PETINIA and EMIT, With That of HPLC-UV for the Routine Monitoring of Mycophenolic Acid in Heart Transplant Patients

被引:1
作者
Kunicki, Pawel K. [1 ]
Pawinski, Tomasz [2 ]
Boczek, Aleksandra [1 ]
Was, Joanna [1 ]
Bodnar-Broniarczyk, Magdalena [2 ]
机构
[1] Inst Cardiol, Dept Clin Biochem, Clin Pharmacol Unit, PL-04628 Warsaw, Poland
[2] Med Univ Warsaw, Fac Pharm, Dept Drug Chem, Warsaw, Poland
关键词
mycophenolic acid; PETINIA; EMIT; HPLC-UV; therapeutic drug monitoring; PERFORMANCE LIQUID-CHROMATOGRAPHY; ENZYME-MULTIPLIED IMMUNOASSAY; TANDEM MASS-SPECTROMETRY; SOLID-ORGAN TRANSPLANTATION; GLUCURONIDE METABOLITES; HUMAN PLASMA; ASSAY; RECIPIENTS; MOFETIL; MPA;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The aim of this study was to evaluate particle enhanced turbidimetric inhibition immunoassay (PETINIA) recently developed for mycophenolic acid (MPA) determination in plasma and to compare it with a reference high-performance liquid chromatography (HPLC) method, using samples from heart transplant recipients. The results are presented in the context of PETINIA being compared with enzyme multiplied immunoassay technique (EMIT). Methods: PETINIA evaluation was performed using 194 routine trough plasma samples at steady state. EMIT was evaluated using 677 samples from 61 steady-state 12-hour profiles obtained from 35 heart transplant patients. Evaluation was undertaken on a Dimension EXL 200 analyzer (PETINIA) and on a Viva-E analyzer (EMIT). Results: The mean MPA concentration measured by PETINIA was significantly higher than that measured by high-performance liquid chromatography combined with UV detector (2.36 +/- 1.30 mcg/mL versus 1.82 +/- 1.23 mcg/mL, respectively, P < 0.0001). Bland-Altman analysis revealed a mean bias of 0.54 mcg/mL [95% confidence interval (CI), 0.49-0.59] comprising 33.48% (95% CI, 30.34-36.61). Passing-Bablok regression was: y = 1.100x + 0.38 (95% CI for slope: 1.044-1.154 and for intercept: 0.30-0.47). Regardless of a significant observed correlation (r(2) = 0.9230, P, 0.0001), the statistical analyses showed a significant difference between PETINIA and the reference chromatographic method. The mean MPA concentration measured by EMIT was significantly higher than that measured by HPLC (7.48 +/- 8.34 mcg/mL versus 5.57 +/- 6.61 mcg/mL, respectively, P < 0.0001) with a mean bias of 1.91 mcg/mL (95% CI, 1.75-2.07) comprising 35.91% (95% CI, 34.37-37.45). The significant difference between EMIT and HPLC was confirmed by Passing-Bablok regression: y = 1.300x + 0.24 (95% CI for slope: 1.279-1.324 and for intercept: 0.18-0.29). The analysis of the determinations, grouped by sampling time, revealed positive bias between EMIT and HPLC ranging from 24.54% to 42.77% and inversely proportional to MPA concentrations with r(2) = 0.9122 (P < 0.001). Conclusions: The new immunochemical PETINIA method was associated with significantly higher MPA concentrations in routine therapeutic drug monitoring samples from heart transplant patients. The magnitude of the MPA overestimation was similar to that observed by use of the EMIT method.
引用
收藏
页码:311 / 318
页数:8
相关论文
共 48 条
  • [1] Altman DG, 2002, CLIN CHEM, V48, P801
  • [2] Bartoli AN, 2011, THER DRUG MONIT, V33, P537
  • [3] Evaluation of an immunoassay (EMIT) for mycophenolic acid in plasma from renal transplant recipients compared with a high-performance liquid chromatography assay
    Beal, JL
    Jones, CE
    Taylor, PJ
    Tett, SE
    [J]. THERAPEUTIC DRUG MONITORING, 1998, 20 (06) : 685 - 690
  • [4] Comparison of a New Enzymatic Assay with a High-Performance Liquid Chromatography/Ultraviolet Detection Method for Therapeutic Drug Monitoring of Mycophenolic Acid in Adult Liver Transplant Recipients
    Blanchet, Benoit
    Taieb, Fabrice
    Conti, Filomena
    Abbas, Halim
    Seydi, Idrissa
    Harcouet, Laura
    Dauphin, Alain
    Calmus, Yvon
    Tod, Michel
    [J]. LIVER TRANSPLANTATION, 2008, 14 (12) : 1745 - 1751
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] Automated monitoring of C2 and C0 blood levels of mycophenolic acid and cyclosporine on the Abbott Architect c8000
    Boer, Klas
    Brehmer-Streck, Sybille
    Deufel, Thomas
    Schmidt, Dirk
    Kiehntopf, Michael
    [J]. CLINICAL BIOCHEMISTRY, 2007, 40 (15) : 1163 - 1167
  • [7] Brandhorst G, 2008, THER DRUG MONIT, V30, P428, DOI 10.1097/FTD.0b013e31817fd590
  • [8] An Investigation Into the Bias Between Liquid Chromatography-Tandem Mass Spectrometry and an Enzyme Multiplied Immunoassay Technique for the Measurement of Mycophenolic Acid
    Brown, Nigel W.
    Franklin, Michael E.
    Einarsdottir, Eyrun N.
    Gonde, Christopher E.
    Pires, Maria
    Taylor, Paul J.
    Tredger, J. Michael
    [J]. THERAPEUTIC DRUG MONITORING, 2010, 32 (04) : 420 - 426
  • [9] Mycophenolic acid monitoring:: Evaluation of the EMIT MPA immunoassay in kidney and lung transplantation
    Brunet, M
    Oppenheimer, F
    Martorell, J
    Vilardell, J
    Carreño, MC
    Carrillo, M
    Corbella, J
    [J]. TRANSPLANTATION PROCEEDINGS, 1999, 31 (06) : 2275 - 2276
  • [10] Establishment of High-Performance Liquid Chromatography and Enzyme Multiplied Immunoassay Technology Methods for Determination of Free Mycophenolic Acid and Its Application in Chinese Liver Transplant Recipients
    Chen, Bing
    Gu, ZhiDong
    Chen, Hao
    Zhang, WeiXia
    Fen, XiaoJing
    Cai, WeiMin
    Fan, QiShi
    [J]. THERAPEUTIC DRUG MONITORING, 2010, 32 (05) : 653 - 660