Monte-Carlo simulations of the full-energy-peak efficiency for a broad-energy germanium detector - a comparison between GEANT4 and FLUKA
被引:7
|作者:
Aviv, O.
[1
]
Elia, P.
[2
]
机构:
[1] Soreq Nucl Res Ctr, IL-81800 Yavne, Israel
[2] Ben Gurion Univ Negev, Dept Biomed Engn, POB 653, IL-84105 Beer Sheva, Israel
关键词:
Detector modelling and simulations I (interaction of radiation with matter;
interaction of photons with matter;
interaction of hadrons with matter;
etc);
Simulation methods and programs;
Gamma detectors (scintillators;
CZT;
HPGe;
HgI etc);
Detector alignment and calibration methods (lasers;
sources;
particle-beams);
GAMMA;
CODE;
D O I:
10.1088/1748-0221/15/05/P05016
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
The detection efficiency of a broad-energy germanium detector was determined empirically and compared to predictions by the Monte-Carlo software GEANT4 and FLUKA. The considered source geometries were point-like, filter paper and a water beaker. Calculations were performed in the photon energy range of 30-1836 keV. Simulated detection efficiencies obtained from both Monte-Carlo software were found to be in a good agreement (typically < 5%) with experimental results for the considered source geometries. A comparison between the detection efficiencies obtained with GEANT4 and FLUKA codes exhibited a good agreement with a robust average of approximately 1%.