BACKGROUND AND PURPOSE: Intracranial aneurysms with a wide-neck or an unfavorable dome-to-neck ratio may be difficult to treat properly and safely. Our aim was to evaluate the TriSpan neck-bridge device to assist coiling of wide-neck bifurcation aneurysms in the anterior circulation. MATERIALS AND METHODS: In 14 patients, we performed 16 TriSpan-assisted coil embolizations with wide-neck bifurcation aneurysms of the anterior circulation. Eleven procedures were indicated for acutely ruptured aneurysms. Five were performed electively for the following: recurrent aneurysm after coil only (n = 1) or after TriSpan-assisted embolization (n = 2), aneurysm remnant after clipping (n = 1), and aneurysm incidentally found in = 1). Procedural and clinical complications were recorded. Follow-up angiography was performed, and clinical outcomes were assessed by using the modified Rankin Scale score. RESULTS: TriSpan-assisted embolization was successful in 15/16 (93.8%) procedures, with complete occlusion in 2/16 (12.5%), near-complete occlusion in 10/16 (62.5%), and incomplete occlusion in 3/16 (18.75%). There were 6 (37.5%) intraprocedural complications: thrombus formation (n = 3), protrusion of a TriSpan loop in the parent artery (n = 1), TriSpan displacement in the aneurysm (n = 1), and tangling of a coil loop in the device (n = 1). Three patients died in the hospital (21.4%). Follow-up angiography or MR angiography was available in 8 (57.1%) patients and showed complete (n 2), near-complete (n = 2), and incomplete occlusion (n = 4). Long-term clinical outcome was no (n = 4) or minor symptoms (n = 1) and moderate (n 2), moderately severe (n = 2), or severe handicap (n 2). CONCLUSION: The use of the TriSpan device is feasible in the anterior circulation and can assist treatment of difficult wide-neck bifurcation aneurysms.
引用
收藏
页码:73 / 78
页数:6
相关论文
共 22 条
[1]   Reconstruction of the basilar tip with T stent configuration for treatment of a wide-neck aneurysm [J].
Akpek, S ;
Morsi, H ;
Benndorf, G ;
Strother, CM ;
Mawad, ME .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (09) :1024-1026
[2]  
Arat A, 2005, AM J NEURORADIOL, V26, P1768
[3]  
Baldi S, 2003, AM J NEURORADIOL, V24, P1222
[4]  
Baxter BW, 1998, AM J NEURORADIOL, V19, P1176
[5]  
Brisman JL, 2005, AM J NEURORADIOL, V26, P1237
[6]  
Cavanagh Stephen J, 2002, J Neurosci Nurs, V34, P288
[7]  
Cloft HJ, 2000, AM J NEURORADIOL, V21, P1312
[8]   SURGICAL RISK AS RELATED TO TIME OF INTERVENTION IN REPAIR OF INTRACRANIAL ANEURYSMS [J].
HUNT, WE ;
HESS, RM .
JOURNAL OF NEUROSURGERY, 1968, 28 (01) :14-&
[9]  
Lubicz B, 2004, AM J NEURORADIOL, V25, P1381
[10]   International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial [J].
Molyneux, A ;
Kerr, R ;
Stratton, I ;
Sandercock, P ;
Clarke, M ;
Shrimpton, J ;
Holman, R .
LANCET, 2002, 360 (9342) :1267-1274