Collimator optimization for detection and quantitation tasks: Application to gallium-67 imaging

被引:18
|
作者
Moore, SC
Kijewski, MF
El Fakhri, G
机构
[1] Brigham & Womens Hosp, Dept Radiol, Div Nucl Med, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Boston, MA 02115 USA
关键词
collimator; detection; estimation; Ga-67; scintigraphy; SPECT;
D O I
10.1109/TMI.2005.857211
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We describe a new approach to the problem of collimator optimization in nuclear medicine; our methodology is illustrated for the challenging case of gallium-67 imaging. Collimator-design methods based on empirical rules, such as specification of an allowable level of single-septal penetration (SSP) at a fixed energy, are especially inappropriate for radionuclides characterized by an abundance of high-energy contaminant photons that scatter in the patient, collimator, and/or detector before detection within one of a few photopeak energy windows. Lead X-rays produced in the collimator are an additional source of contamination. We designed optimal collimation for Ga-67 based on relevant clinical imaging tasks and a realistic simulation of photon transport in a phantom, collimator, and detector. Collimator designs were compared on the basis of performance in lesion detection, as predicted by a three-channel Hotelling observer (CHO), as well as in tumor and background activity estimation (EST), quantified by task-specific signal-to-noise ratios (SNRs). The optimal values of collimator lead content were 22.0 and 23.8 g/cm(2), respectively, for CHO and EST, while the optimal geometric resolution values were 1.8 and 1.6 cm full-width at half-maximum (FWHM), respectively, at a distance of 23.5 cm. The resolution of a commercially available medium-energy low-penetration collimator (MELP) is 1.9 cm FWHM at this distance. The optimal values for SSP at 300 keV were 7.3% and 5.8% based on CHO and EST, respectively, compared to 5.2% for the MELP collimator. Compared with the commercial MELP collimator, the Ga-67 collimator optimized for tumor detection or activity estimation tasks provided improved geometric spatial resolution with reduced geometric efficiency and, surprisingly, allowed an increased level of single-septal penetration.
引用
收藏
页码:1347 / 1356
页数:10
相关论文
共 50 条
  • [1] DETECTION OF INFECTIONS WITH GALLIUM-67 AND SCINTIGRAPHIC IMAGING
    SILVA, J
    HARVEY, WC
    JOURNAL OF INFECTIOUS DISEASES, 1974, 130 (02): : 125 - 131
  • [2] GALLIUM-67 TUMOR IMAGING
    WAXMAN, AD
    SIEMSEN, JK
    WESTERN JOURNAL OF MEDICINE, 1975, 123 (05): : 388 - 389
  • [3] A MULTICHANNEL SYSTEM FOR IMAGING WITH GALLIUM-67
    ROSS, DA
    MCCLAIN, WJ
    EAST, JK
    BELL, PR
    SOUTHERN MEDICAL JOURNAL, 1970, 63 (11) : 1374 - &
  • [4] Superiority of SPECT to planar imaging in the detection of myocardial inflarnation with Gallium-67
    Correa, PL
    Felix, RCM
    Azevedo, JC
    Dohmann, HFR
    Montera, MW
    Avila, DA
    Mesquita, ET
    Mesquita, CT
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S321 - S321
  • [5] Gallium-67 imaging of pericardial lymphoma in AIDS
    Prvulovich, EM
    Costa, DC
    Bomanji, J
    Clarke, GA
    Townsend, CE
    Miller, RF
    Ell, PJ
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (06) : 995 - 996
  • [6] DETECTION OF EXPERIMENTAL OSTEOMYELITIS WITH GALLIUM-67 CITRATE
    RAFKIN, H
    JOURNAL DE CHIRURGIE, 1975, 109 (5-6): : 656 - 656
  • [7] GALLIUM-67 IN DETECTION OF METASTATIC BREAST CARCINOMA
    RICHMAN, SD
    INGLE, JN
    LEVENSON, SM
    TORMEY, DC
    JONES, AE
    JOHNSTON, GS
    JOURNAL OF NUCLEAR MEDICINE, 1975, 16 (06) : 560 - 561
  • [8] Gallium-67 scintigraphy: a cornerstone in functional imaging of lymphoma
    Einat Even-Sapir
    Ora Israel
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : S65 - S81
  • [9] Does Gadopentate interfere with gallium-67 tumour imaging?
    Baker, RJ
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (05) : 904 - 904
  • [10] Revisiting gallium-67 imaging: investigation of the energy photopeaks
    Ali, Layla
    Khalil, Magdy
    Hadi, Nadia
    NUCLEAR MEDICINE COMMUNICATIONS, 2010, 31 (12) : 1068 - 1074