We describe a new approach to the problem of collimator optimization in nuclear medicine; our methodology is illustrated for the challenging case of gallium-67 imaging. Collimator-design methods based on empirical rules, such as specification of an allowable level of single-septal penetration (SSP) at a fixed energy, are especially inappropriate for radionuclides characterized by an abundance of high-energy contaminant photons that scatter in the patient, collimator, and/or detector before detection within one of a few photopeak energy windows. Lead X-rays produced in the collimator are an additional source of contamination. We designed optimal collimation for Ga-67 based on relevant clinical imaging tasks and a realistic simulation of photon transport in a phantom, collimator, and detector. Collimator designs were compared on the basis of performance in lesion detection, as predicted by a three-channel Hotelling observer (CHO), as well as in tumor and background activity estimation (EST), quantified by task-specific signal-to-noise ratios (SNRs). The optimal values of collimator lead content were 22.0 and 23.8 g/cm(2), respectively, for CHO and EST, while the optimal geometric resolution values were 1.8 and 1.6 cm full-width at half-maximum (FWHM), respectively, at a distance of 23.5 cm. The resolution of a commercially available medium-energy low-penetration collimator (MELP) is 1.9 cm FWHM at this distance. The optimal values for SSP at 300 keV were 7.3% and 5.8% based on CHO and EST, respectively, compared to 5.2% for the MELP collimator. Compared with the commercial MELP collimator, the Ga-67 collimator optimized for tumor detection or activity estimation tasks provided improved geometric spatial resolution with reduced geometric efficiency and, surprisingly, allowed an increased level of single-septal penetration.
引用
收藏
页码:1347 / 1356
页数:10
相关论文
共 25 条
  • [1] Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability
    Abbey, CK
    Barrett, HH
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (03): : 473 - 488
  • [2] ABBEY CK, 1995, INT C INF PROC MED I
  • [3] MODEL OBSERVERS FOR ASSESSMENT OF IMAGE QUALITY
    BARRETT, HH
    YAO, J
    ROLLAND, JP
    MYERS, KJ
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1993, 90 (21) : 9758 - 9765
  • [4] Barrett HH., 2004, FDN IMAGE SCI
  • [5] Cherry Simon R., 2012, Physics in Nuclear Medicine, Vforth
  • [6] Approximation of hexagonal holes by square holes in Monte Carlo simulation of gamma-camera collimation
    de Vries, DJ
    Moore, SC
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) : 2186 - 2195
  • [7] DEVELOPMENT AND VALIDATION OF A MONTE-CARLO SIMULATION OF PHOTON TRANSPORT IN AN ANGER CAMERA
    DEVRIES, DJ
    MOORE, SC
    ZIMMERMAN, RE
    MUELLER, SP
    FRIEDLAND, B
    LANZA, RC
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (04) : 430 - 438
  • [8] The effect of image compression in model and human performance
    Eckstein, MP
    Abbey, CK
    Bochud, FO
    Bartroff, JL
    Whiting, JS
    [J]. MEDICAL IMAGING 1999: IMAGE PERCEPTION AND PERFORMANCE, 1999, 3663 : 243 - 252
  • [9] Optimization of Ga-67 imaging for detection and estimation tasks: Dependence of imaging performance on spectral acquisition parameters
    El Fakhri, G
    Moore, SC
    Kijewski, MF
    [J]. MEDICAL PHYSICS, 2002, 29 (08) : 1859 - 1866
  • [10] HOTELLING TRACE CRITERION AND ITS CORRELATION WITH HUMAN-OBSERVER PERFORMANCE
    FIETE, RD
    BARRETT, HH
    SMITH, WE
    MYERS, KJ
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05): : 945 - 953