The Influence of Kinematic Conditions and Variations in Component Positioning on the Severity of Edge Loading and Wear of Ceramic-on-Ceramic Hip Bearings

被引:2
作者
Ali, Murat [1 ]
Al-Hajjar, Mazen [1 ,2 ]
Fisher, John [1 ]
Jennings, Louise M. [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] DePuySynthes Joint Reconstruct, Leeds LS11 8DT, W Yorkshire, England
来源
CERAMICS-SWITZERLAND | 2019年 / 2卷 / 03期
基金
英国工程与自然科学研究理事会; 英国生物技术与生命科学研究理事会;
关键词
hip replacement; hip simulator; ceramic; wear; edge loading; JOINT SIMULATOR; CUP ANGLE; METAL; MICROSEPARATION; PROSTHESES; TRIBOLOGY; FRICTION; DEBRIS;
D O I
10.3390/ceramics2030037
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Dynamic separation and direct edge loading of hip replacement bearings can be caused by many factors, including implant positioning, implant design, changes in device over time, surgical variations and patient variations. Such dynamic separation and direct edge loading can lead to increased wear. Different input kinematic conditions have been used for experimental hip simulator studies to produce clinically relevant elliptical contact wear paths between the bearings during gait. The aim of this study was to investigate the influence of input kinematics (two axes of rotation simulation conditions (without abduction/adduction) and three axes of rotation simulation conditions (with abduction/adduction and different loading profiles) and variations in component positioning (different levels of medial-lateral translational mismatch at standard and steep cup inclination angles) on the occurrence, severity of edge loading, dynamic separation and wear of size 36 mm ceramic-on-ceramic hip bearings on an electromechanical hip joint simulator. The results showed that, overall, either of the two axes or three axes input profiles were equally valid in providing a suitable preclinical testing method for assessing the occurrence and severity of edge loading and wear under edge loading conditions. In terms of component positioning, as cup inclination and medial-lateral translational mismatch increased, so did dynamic separation, axial load at the rim, severity of edge loading and wear.
引用
收藏
页码:488 / 501
页数:14
相关论文
共 31 条
  • [1] Hip joint simulators: State of the art
    Affatato, S.
    Leardini, W.
    Zovalloni, M.
    [J]. BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2006, : 171 - +
  • [2] Tribology and total hip joint replacement: Current concepts in mechanical simulation
    Affatato, S.
    Spinelli, A.
    Zavalloni, M.
    Mazzega-Fabbro, C.
    Viceconti, A.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2008, 30 (10) : 1305 - 1317
  • [3] Combined wear and ageing of ceramic-on-ceramic bearings in total hip replacement under edge loading conditions
    Al-Hajjar, Mazen
    Gremillard, Laurent
    Begand, Sabine
    Oberbach, Thomas
    Hans, Karen
    Delfosse, Daniel
    Chevalier, Jerome
    Jennings, Louise M.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 98 : 40 - 47
  • [4] Ali M., 2019, INFLUENCE KINEMATIC
  • [5] Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings
    Ali, Murat
    Al-Hajjar, Mazen
    Partridge, Susan
    Williams, Sophie
    Fisher, John
    Jennings, Louise M.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 389 - 397
  • [6] A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates
    Barbour, PSM
    Stone, MH
    Fisher, J
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) : 455 - 467
  • [7] A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups
    Barbour, PSM
    Stone, MH
    Fisher, J
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) : 569 - 576
  • [8] Friction of total hip replacements with different bearings and loading conditions
    Brockett, Claire
    Williams, Sophie
    Jin, Zhongmin
    Isaac, Graham
    Fisher, John
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (02) : 508 - 515
  • [9] Slide track analysis of eight contemporary hip simulator designs
    Calonius, O
    Saikko, V
    [J]. JOURNAL OF BIOMECHANICS, 2002, 35 (11) : 1439 - 1450
  • [10] Bioengineering reasons for the failure of metal-on-metal hip prostheses AN ENGINEER'S PERSPECTIVE
    Fisher, J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (08): : 1001 - 1004