A multiwire proportional counter for XAS fluorescence experiments

被引:2
|
作者
Barrea, RA
Tamura, E
Tolentino, HCN
机构
[1] Lab Nacl Luz Sincrotron, BR-13083970 Campinas, SP, Brazil
[2] Univ Nacl Cordoba, Fac Matemat Astron & Fis, RA-5010 Cordoba, Argentina
关键词
multiwire proportional counters; x-ray detectors; diluted samples; fluorescence EXAFS;
D O I
10.1107/S0909049500014746
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A multiwire proportional counter was used in fluorescence X-ray absorption measurements and a comparison to a Si( Li) and NaI(Tl) detectors was done. The main features of the multiwire proportional counter are its high counting rate capability (10(7) counts.s(-1)) and large active area (6 x 6 cm(2)). It was shown that the MWPC is suitable for fluorescence absorption. Although the maximum capability was not reached in the present experiments, it was found that as the counting rate increase the MWPC performance became better than Si(Li) detectors and shows a similar response to the scintillator counter at medium counting rates (up to 10(5) counts.s(-1)).
引用
收藏
页码:381 / 383
页数:3
相关论文
共 50 条
  • [1] MULTIWIRE PROPORTIONAL COUNTER
    TAIT, WH
    JOURNAL OF THE INSTITUTION OF NUCLEAR ENGINEERS, 1974, 15 (02): : 47 - 52
  • [2] PRESSURIZED MULTIWIRE PROPORTIONAL COUNTER FOR ELECTRONS
    KITAHARA, T
    ISOZUMI, Y
    ITO, S
    NUCLEAR INSTRUMENTS & METHODS, 1977, 140 (02): : 263 - 267
  • [3] A MULTIWIRE PROPORTIONAL COUNTER BEAM SCANNER
    VANDERGRAFF, H
    DIRKER, J
    VANEIJK, CWE
    LOURENS, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1981, 188 (03): : 495 - 496
  • [4] HIGH COUNT RATE MULTIWIRE PROPORTIONAL COUNTER
    SEMPER, RJ
    GUARNIERI, CR
    WALKER, JC
    NUCLEAR INSTRUMENTS & METHODS, 1975, 129 (02): : 447 - 450
  • [5] DEVELOPMENT OF A MULTIWIRE POSITION SENSITIVE PROPORTIONAL COUNTER
    Kakavand, Tayeb
    Nasseri, Mohammad M.
    Sharifi, Bahador
    Razavi, Rohallah
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2011, 26 (03): : 209 - 213
  • [6] MULTIWIRE PROPORTIONAL COUNTER DESIGNED FOR MICRODOSIMETRIC STUDIES
    BAILY, NA
    STEIGERWALT, JE
    RADIATION RESEARCH, 1977, 71 (02) : 284 - 299
  • [7] THE APPLICATION OF THE MULTIWIRE PROPORTIONAL COUNTER TO BONE DENISTOMETRY
    HORSMAN, A
    METABOLIC BONE DISEASE & RELATED RESEARCH, 1983, 4 (06): : 363 - 363
  • [8] CWRU MULTIWIRE PROPORTIONAL COUNTER READOUT SYSTEM
    BEVINGTON, PR
    LESKOVEC, RA
    NUCLEAR INSTRUMENTS & METHODS, 1977, 147 (02): : 431 - 442
  • [9] STUDY OF A GAS SCINTILLATION MULTIWIRE PROPORTIONAL COUNTER
    ALDARGAZELLI, SS
    ARIYARATNE, TR
    BREARE, JM
    NANDI, BC
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1978, 11 (13) : 1773 - &
  • [10] MULTIWIRE PROPORTIONAL AND SEMIPROPORTIONAL COUNTER WITH A VARIABLE SENSITIVE VOLUME
    GRUNBERG, C
    COHEN, L
    MATHIEU, L
    NUCLEAR INSTRUMENTS & METHODS, 1970, 78 (01): : 102 - &