Use of 2-Octyl Cyanoacrylate Adhesive and Polyester Mesh for Wound Closure in Primary Knee Arthroplasty

被引:13
作者
Holte, Andrew J. [1 ]
Tofte, Josef N. [2 ]
Dahlberg, Greta J. [1 ]
Noiseux, Nicolas [2 ]
机构
[1] Univ Iowa, Roy J & Lucille A Carver Coll Med, Iowa City, IA USA
[2] Univ Iowa Hosp & Clin, Dept Orthopaed & Rehabil, 200 Hawkins Dr,01008 JPP, Iowa City, IA 52242 USA
关键词
SKIN CLOSURE; TRADITIONAL SUTURES; JOINT ARTHROPLASTY; DEVICE;
D O I
10.3928/01477447-20170531-03
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Wound closure in primary knee arthroplasty is important both for surgical success, by minimizing the risk of wound complications and infection, and for patient satisfaction with wound care and cosmesis. This article reports the use of a topical self-adhering, pressure-sensitive polyester mesh combined with 2-octyl cyanoacrylate adhesive for final skin closure of primary knee arthroplasty without external staples or sutures. During a 2.5-year period, 360 consecutive primary knee arthroplasty procedures (321 total knee arthroplasty procedures and 39 partial knee arthroplasty procedures) performed by 1 surgeon at 1 institution were reviewed. For all primary arthroplasty procedures, closure was performed with a standardized method. Final skin closure was performed with 2-octyl cyanoacrylate adhesive and polyester mesh. Patients returned at 6 weeks and 1 year postoperatively. The study found a 0.8% rate of superficial wound complications, no cases of dehiscence, no wound complications requiring operative irrigation and debridement, and no prosthetic joint infections. One patient had a reaction consistent with previously reported localized contact dermatitis that was believed to be related to the 2-octyl cyanoacrylate adhesive. In 2 cases, treatment with cephalexin was required, in 1 case for 2 small suture abscesses and in the other case for peri-incisional cellulitis. All wound complications resolved by 6 weeks. The results showed that 2-octyl cyanoacrylate adhesive and polyester mesh can be used successfully in the setting of primary knee arthroplasty for closure of high-tension knee arthroplasty wounds. This series found a 0.8% rate of superficial wound complications, no cases of dehiscence, and no infections.
引用
收藏
页码:E784 / E787
页数:4
相关论文
共 14 条
[1]   Barbed Suture Is Associated With Increased Risk of Wound Infection After Unicompartmental Knee Arthroplasty [J].
Chawla, Harshvardhan ;
van der List, Jelle P. ;
Fein, Nicole B. ;
Henry, Michael W. ;
Pearle, Andrew D. .
JOURNAL OF ARTHROPLASTY, 2016, 31 (07) :1561-1567
[2]  
Dobbelaere A, 2015, ACTA ORTHOP BELG, V81, P454
[3]   Allergic Contact Dermatitis to Dermabond (2-Octyl Cyanoacrylate) After Total Knee Arthroplasty [J].
Durando, Dunnett ;
Porubsky, Caitlin ;
Winter, Suzanne ;
Kalymon, JoEllen ;
O'Keefe, Thomas ;
LaFond, Ann Ammond .
DERMATITIS, 2014, 25 (02) :99-100
[4]   A Comparison of Wound Closure Techniques for Total Knee Arthroplasty [J].
Eggers, Mitchell D. ;
Fang, Li ;
Lionberger, David R. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (08) :1251-1258
[5]   The Use of Dermabond® as an Adjunct to Wound Closure After Total Knee Arthroplasty: Examining Immediate Post-Operative Wound Drainage [J].
El-Gazzar, Yaser ;
Smith, Daniel C. ;
Kim, Sun Jin ;
Hirsh, David M. ;
Blum, Yossef ;
Cobelli, Marcie ;
Cohen, Hillel W. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (04) :553-556
[6]   Effective Wound Closure With a New Two-Component Wound Closure Device (Prineo™) in Excisional Body-Contouring Surgery: Experience in Over 200 Procedures [J].
Huemer, Georg M. ;
Schmidt, Manfred ;
Helml, Gudrun H. ;
Shafighi, Maziar ;
Dunst-Huemer, Karin M. .
AESTHETIC PLASTIC SURGERY, 2012, 36 (02) :382-386
[7]   Use of 2-Octyl Cyanoacrylate Together with a Self-Adhering Mesh (Dermabond™ Prineo™) for Skin Closure Following Abdominoplasty: An Open, Prospective, Controlled, Randomized, Clinical Study [J].
Parvizi, D. ;
Friedl, H. ;
Schintler, M. V. ;
Rappl, T. ;
Laback, C. ;
Wiedner, M. ;
Vasiljeva, A. ;
Kamolz, L. P. ;
Spendel, S. .
AESTHETIC PLASTIC SURGERY, 2013, 37 (03) :529-537
[8]   Wound Complications in Joint Arthroplasty: Comparing Traditional and Modern Methods of Skin Closure [J].
Patel, Ronak M. ;
Cayo, Max ;
Patel, Arpan ;
Albarillo, Marie ;
Puri, Lalit .
ORTHOPEDICS, 2012, 35 (05) :E641-E646
[9]  
Peersman G, 2006, HSS J, V2, P70, DOI 10.1007/s11420-005-0130-2
[10]   A Comparison of a New Skin Closure Device and Intradermal Sutures in the Closure of Full-Thickness Surgical Incisions [J].
Richter, Dirk ;
Stoff, Alexander ;
Ramakrishnan, Venkat ;
Exner, Klaus ;
Jernbeck, Jan ;
Blondeel, Phillip N. .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2012, 130 (04) :843-850