Delayed migration of a screw into the gastrointestinal tract after anterior cervical spine plating

被引:49
作者
Gazzeri, Roberto [1 ]
Tamorri, Marco [2 ]
Faiola, Andrea [2 ]
Gazzeri, Giovanni [2 ]
机构
[1] San Giovanni Addolorata Hosp, Dept Neurosurg, Rome, Italy
[2] San Filippo Neri Hosp, Rome, Italy
关键词
anterior fusion; instrumentation failure; cervical spine; cervical plating; complication; screw migration;
D O I
10.1097/BRS.0b013e31816b8831
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Case report and clinical discussion. Objective. To describe a rare case of spinal cervical screw migration into the gastrointestinal tract 11 years after surgery. Summary of Background Data. Anterior cervical spine fusion and stabilization with plating is a well-established procedure for cervical myelopathy, cervical spinal trauma, and spinal infectious disease. Esophageal injury has been related to screw or plate extrusion. Methods. We present a 45-year-old white man suffering from severe quadriparesis and neck pain. Cervical spine magnetic resonance images showed spondylodiscitis and spinal fracture of C4 and C5 vertebral bodies. He underwent anterior surgical decompression consisting of C4 and C5 corpectomies and fusion with fibular allograft fixated with a cervical plate. The patient's neurologic examination gradually improved during his follow-up. Results. The patient returned 11 years after cervical spine graft and plating, complaining of severe dysphagia and high fever. Radiographs of the cervical spine showed a screw back-out. Three days later, a new radiograph of the cervical spine revealed a progression of the screw's extrusion. In a new cervical spine radiograph, obtained immediately before surgical intervention, the displaced screw was no more visible. Abdominal radiograph showed the missed screw in the right lower abdominal quadrant. Barium meal swallowing test showed no leakage on the cervical area. At 6 months follow-up, the patient is in good condition and symptom free. Conclusion. In this case, the delayed esophageal perforation occurred 11 years after initial surgery: the rapid progression of the screw extrusion after initial pull-out (6 days) was documented by several radiographs and spontaneous closure of the fistula was achieved.
引用
收藏
页码:E268 / E271
页数:4
相关论文
共 18 条
[1]   Management of esophageal perforations after therapeutic upper gastrointestinal endoscopy [J].
Adamek, HE ;
Jakobs, R ;
Dorlars, D ;
Martin, WR ;
Kromer, MU ;
Riemann, JF .
SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1997, 32 (05) :411-414
[2]  
Chataigner H, 1997, REV CHIR ORTHOP, V83, P78
[3]  
CLOWARD RB, 1971, SURGERY, V69, P175
[4]   Extrusion of a screw into the gastrointestinal tract after anterior cervical spine plating [J].
Fountas, Kostas N. ;
Kapsalaki, Eftychia Z. ;
Machinis, Theofilos ;
Robinson, Joe Sam .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (03) :199-203
[5]   Missing anterior cervical plate and screws - A case report [J].
Fujibayashi, S ;
Shikata, J ;
Kamiya, N ;
Tanaka, C .
SPINE, 2000, 25 (17) :2258-2261
[6]   Esophageal perforations after anterior cervical surgery [J].
Gaudinez, RF ;
English, GM ;
Gebhard, JS ;
Brugman, JL ;
Donaldson, DH ;
Brown, CW .
JOURNAL OF SPINAL DISORDERS, 2000, 13 (01) :77-84
[7]   Oral extrusion of a screw after anterior cervical spine plating [J].
Geyer, TE ;
Foy, MA .
SPINE, 2001, 26 (16) :1814-1816
[8]   The use of plate fixation in anterior surgery of the degenerative cervical spine: a comparative prospective clinical study [J].
Grob, D ;
Peyer, JV ;
Dvorak, J .
EUROPEAN SPINE JOURNAL, 2001, 10 (05) :408-413
[9]   Descending cervical mediastinitis [J].
Kiernan, PD ;
Hernandez, A ;
Byrne, WD ;
Bloom, R ;
Dicicco, B ;
Hetrick, V ;
Graling, P ;
Vaughan, B .
ANNALS OF THORACIC SURGERY, 1998, 65 (05) :1483-1488
[10]   The significance of hardware failure in anterior cervical plate fixation - Patients with 2- to 7-year follow-up [J].
Lowery, GL ;
McDonough, RF .
SPINE, 1998, 23 (02) :181-186