Geometrical considerations for the design of liquid-phase biochemical sensors using a cantilever's fundamental in-plane mode

被引:23
作者
Beardslee, L. A. [1 ]
Josse, F. [2 ]
Heinrich, S. M. [3 ]
Dufour, I. [4 ]
Brand, O. [1 ]
机构
[1] Georgia Inst Technol, Dept Elect & Comp Engn, Elect & Comp Engn Fac, Atlanta, GA 30332 USA
[2] Marquette Univ, Dept Elect & Comp Engn, Milwaukee, WI 53233 USA
[3] Marquette Univ, Dept Civil & Environm Engn, Milwaukee, WI 53233 USA
[4] Univ Bordeaux, IMS, Talence, France
基金
美国国家科学基金会;
关键词
Liquid-phase; Cantilevers; Bio-chemical sensors; MICROCANTILEVERS; ENVIRONMENT; INTEGRATION; SYSTEMS; CHIP;
D O I
10.1016/j.snb.2012.01.035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The influence of the beam geometry on the quality factor and resonance frequency of resonant silicon cantilever beams vibrating in their fundamental in-plane flexural mode in water has been investigated. Compared to cantilevers vibrating in their first out-of-plane flexural mode, utilizing the in-plane mode results in reduced damping and reduced mass loading by the surrounding fluid. Quality factors as high as 86 have been measured in water for cantilevers with a 20 mu m thick silicon layer. Based on the experimental data, design guidelines are established for beam dimensions that ensure maximal Q-factors and minimal mass loading by the surrounding fluid, thus improving the limit-of-detection of mass-sensitive biochemical sensors. Elementary theory is also presented to help explain the observed trends. Additional discussion focuses on the tradeoffs that exist in designing liquid-phase biochemical sensors using in-plane cantilevers. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 28 条
[1]  
Agache V, 2011, PROC IEEE MICR ELECT, P157, DOI 10.1109/MEMSYS.2011.5734385
[2]  
Allan D.W., 1981, 35 ANN FREQ CONTR S, P470
[3]  
Beardslee L.A., TRANSD 2011 BEIJ CHI, P1112
[4]   Thermal Excitation and Piezoresistive Detection of Cantilever In-Plane Resonance Modes for Sensing Applications [J].
Beardslee, Luke Armitage ;
Addous, Assim M. ;
Heinrich, Stephen ;
Josse, Fabien ;
Dufour, Isabelle ;
Brand, Oliver .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2010, 19 (04) :1015-1017
[5]   Microsensor integration into systems-on-chip [J].
Brand, Oliver .
PROCEEDINGS OF THE IEEE, 2006, 94 (06) :1160-1176
[6]   Weighing of biomolecules, single cells and single nanoparticles in fluid [J].
Burg, Thomas P. ;
Godin, Michel ;
Knudsen, Scott M. ;
Shen, Wenjiang ;
Carlson, Greg ;
Foster, John S. ;
Babcock, Ken ;
Manalis, Scott R. .
NATURE, 2007, 446 (7139) :1066-1069
[7]   Multimode thermoelastic dissipation [J].
Chandorkar, Saurabh A. ;
Candler, Robert N. ;
Duwel, Amy ;
Melamud, Renata ;
Agarwal, Manu ;
Goodson, Kenneth E. ;
Kenny, Thomas W. .
JOURNAL OF APPLIED PHYSICS, 2009, 105 (04)
[8]   Ethanol vapor detection in saline solution using piezoresistive microcantilevers [J].
Cheney, C. Parks ;
Wig, A. ;
Hedden, D. L. ;
Gehl, A. ;
Lereu, A. L. ;
Farahi, R. H. ;
Hunter, S. R. ;
Ferrell, T. L. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (09)
[9]   Generalized model of resonant polymer-coated microcantilevers in viscous liquid media [J].
Cox, Russell ;
Josse, Fabien ;
Wenzel, Michael J. ;
Heinrich, Stephen M. ;
Dufour, Isabelle .
ANALYTICAL CHEMISTRY, 2008, 80 (15) :5760-5767
[10]  
Heinrich S.M., 2010, P IEEE SENS C WAIK H