Deep learning based spectral extrapolation for dual-source, dual-energy x-ray computed tomography

被引:10
作者
Clark, Darin P. [1 ]
Schwartz, Fides R. [2 ]
Marin, Daniele [2 ]
Ramirez-Giraldo, Juan C. [3 ]
Badea, Cristian T. [1 ]
机构
[1] Duke Univ, Dept Radiol, Ctr Vivo Microscopy, Durham, NC 27710 USA
[2] Duke Univ, Dept Radiol, Durham, NC 27710 USA
[3] Siemens Healthineers, CT R&D Collaborat, Cary, NC USA
关键词
data completion; deep learning; dual-energy CT; dual-source CT; CT; SCAN;
D O I
10.1002/mp.14324
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Data completion is commonly employed in dual-source, dual-energy computed tomography (CT) when physical or hardware constraints limit the field of view (FoV) covered by one of two imaging chains. Practically, dual-energy data completion is accomplished by estimating missing projection data based on the imaging chain with the full FoV and then by appropriately truncating the analytical reconstruction of the data with the smaller FoV. While this approach works well in many clinical applications, there are applications which would benefit from spectral contrast estimates over the larger FoV (spectral extrapolation)-e.g. model-based iterative reconstruction, contrast-enhanced abdominal imaging of large patients, interior tomography, and combined temporal and spectral imaging. Methods To document the fidelity of spectral extrapolation and to prototype a deep learning algorithm to perform it, we assembled a data set of 50 dual-source, dual-energy abdominal x-ray CT scans (acquired at Duke University Medical Center with 5 SiemensFlashscanners; chain A: 50 cm FoV, 100 kV; chain B: 33 cm FoV, 140 kV + Sn; helical pitch: 0.8). Data sets were reconstructed usingReconCT(v14.1, Siemens Healthineers): 768 x 768 pixels per slice, 50 cm FoV, 0.75 mm slice thickness, "Dual-Energy - WFBP" reconstruction mode with dual-source data completion. A hybrid architecture consisting of a learned piecewise linear transfer function (PLTF) and a convolutional neural network (CNN) was trained using 40 scans (five scans reserved for validation, five for testing). The PLTF learned to map chain A spectral contrast to chain B spectral contrast voxel-wise, performing an image domain analog of dual-source data completion with approximate spectral reweighting. The CNN with its U-net structure then learned to improve the accuracy of chain B contrast estimates by copying chain A structural information, by encoding prior chain A, chain B contrast relationships, and by generalizing feature-contrast associations. Training was supervised, using data from within the 33-cm chain B FoV to optimize and assess network performance. Results Extrapolation performance on the testing data confirmed our network's robustness and ability to generalize to unseen data from different patients, yielding maximum extrapolation errors of 26 HU following the PLTF and 7.5 HU following the CNN (averaged per target organ). Degradation of network performance when applied to a geometrically simple phantom confirmed our method's reliance on feature-contrast relationships in correctly inferring spectral contrast. Integrating our image domain spectral extrapolation network into a standard dual-source, dual-energy processing pipeline for SiemensFlashscanner data yielded spectral CT data with adequate fidelity for the generation of both 50 keV monochromatic images and material decomposition images over a 30-cm FoV for chain B when only 20 cm of chain B data were available for spectral extrapolation. Conclusions Even with a moderate amount of training data, deep learning methods are capable of robustly inferring spectral contrast from feature-contrast relationships in spectral CT data, leading to spectral extrapolation performance well beyond what may be expected at face value. Future work reconciling spectral extrapolation results with original projection data is expected to further improve results in outlying and pathological cases.
引用
收藏
页码:4150 / 4163
页数:14
相关论文
共 50 条
[31]   Dual-source multienergy CT with triple or quadruple x-ray beams [J].
Yu, Lifeng ;
Ren, Liqiang ;
Li, Zhoubo ;
Leng, Shuai ;
McCollough, Cynthia H. .
JOURNAL OF MEDICAL IMAGING, 2018, 5 (03)
[32]   Deep learning approach for automatic segmentation of ulna and radius in dual-energy X-ray imaging [J].
Yang, Fan ;
Weng, Xin ;
Miao, Yuehong ;
Wu, Yuhui ;
Xie, Hong ;
Lei, Pinggui .
INSIGHTS INTO IMAGING, 2021, 12 (01)
[33]   Diagnosing cervical lymph node metastasis in oral squamous cell carcinoma based on third-generation dual-source, dual-energy computed tomography [J].
Yong-Heng Luo ;
Xi-Long Mei ;
Qin-Ru Liu ;
Bo Jiang ;
Sheng Zhang ;
Ke Zhang ;
Xia Wu ;
Yong-Mei Luo ;
Ya-Jun Li .
European Radiology, 2023, 33 :162-171
[34]   Diagnosing cervical lymph node metastasis in oral squamous cell carcinoma based on third-generation dual-source, dual-energy computed tomography [J].
Luo, Yong-Heng ;
Mei, Xi-Long ;
Liu, Qin-Ru ;
Jiang, Bo ;
Zhang, Sheng ;
Zhang, Ke ;
Wu, Xia ;
Luo, Yong-Mei ;
Li, Ya-Jun .
EUROPEAN RADIOLOGY, 2023, 33 (01) :162-171
[35]   Variation in Hounsfield unit calculated using dual-energy computed tomography: comparison of dual-layer, dual-source, and fast kilovoltage switching technique [J].
Ohira, Shingo ;
Mochizuki, Junji ;
Niwa, Tatsunori ;
Endo, Kazuyuki ;
Minamitani, Masanari ;
Yamashita, Hideomi ;
Katano, Atsuto ;
Imae, Toshikazu ;
Nishio, Teiji ;
Koizumi, Masahiko ;
Nakagawa, Keiichi .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) :458-466
[36]   A Characterization of Deep Learning Reconstruction Applied to Dual-Energy Computed Tomography Monochromatic and Material Basis Images [J].
Nikolau, Ethan P. ;
Toia, Giuseppe V. ;
Nett, Brian ;
Tang, Jie ;
Szczykutowicz, Timothy P. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2023, 47 (03) :437-444
[37]   Radiation Optimized Dual-source Dual-energy Computed Tomography Pulmonary Angriography: Intra-individual and Inter-individual Comparison [J].
Li, Xie ;
Chen, Guo Zhong ;
Zhao, Yan E. ;
Schoepf, U. Joseph ;
Albrecht, Moritz H. ;
Bickford, Matthew W. ;
Gu, Hai Feng ;
Zhang, Long Jiang ;
Lu, Guang Ming .
ACADEMIC RADIOLOGY, 2017, 24 (01) :13-21
[38]   Pre-operative Percutaneous Nephrolithotripsy Characterisation of Kidney Stones with Second-Generation Dual-Source Dual-Energy Computed Tomography [J].
Ali, D. K. Mella Mohd ;
Mahmud, Mohd Hafizi ;
Mohamad, Noor Shafini .
MALAYSIAN JOURNAL OF MEDICAL SCIENCES, 2020, 27 (05) :43-52
[39]   Comparison of dual-source computed tomography (CT) dual-energy angiography with conventional 3D imaging towards caudomedial vessels [J].
Li, Daowei ;
Liu, Na ;
Ma, Qiang ;
He, Dan .
BIOMEDICAL RESEARCH-INDIA, 2017, 28 (12) :5445-5449
[40]   Dual-Energy Computed Tomography: Technology, Applications, and Advantages [J].
Tran, Julia ;
Yaghmai, Vahid ;
Shi, James .
CONTEMPORARY DIAGNOSTIC RADIOLOGY, 2025, 48 (01)