Quantification techniques to minimize the effects of native T1 variation and B1 inhomogeneity in dynamic contrast-enhanced MRI of the breast at 3 T

被引:4
作者
Azlan, Che A. [1 ,2 ]
Ahearn, Trevor S. [1 ]
Di Giovanni, Pierluigi [1 ]
Semple, Scott I. K. [3 ]
Gilbert, Fiona J. [1 ]
Redpath, Thomas W. [1 ]
机构
[1] Univ Aberdeen, Aberdeen Biomed Imaging Ctr, Aberdeen AB25 2ZD, Scotland
[2] Univ Malaya, Dept Biomed Imaging, Kuala Lumpur, Malaysia
[3] Univ Edinburgh, Clin Res Imaging Ctr, Edinburgh, Midlothian, Scotland
关键词
DCE-MRI; high-field MRI; native T1; B1; inhomogeneity; AGAROSE GELS; DCE-MRI; ACCURACY; MAMMOGRAPHY; SENSITIVITY; TURBOFLASH; PULSES;
D O I
10.1002/mrm.23021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The variation of the native T1 (T10) of different tissues and B1 transmission-field inhomogeneity at 3 T are major contributors of errors in the quantification of breast dynamic contrast-enhanced MRI. To address these issues, we have introduced new enhancement indices derived from saturation-recovery snapshot-FLASH (SRSF) images. The stability of the new indices, i.e., the SRSF enhancement factor (EFSRSF) and its simplified version (EF'SRSF) with respect to differences in T10 and B1 inhomogeneity was compared against a typical index used in breast dynamic contrast-enhanced MRI, i.e., the enhancement ratio (ER), by using computer simulations. Imaging experiments with Gd-DTPA-doped gel phantoms and a female volunteer were also performed. A lower error was observed in the new indices compared to enhancement ratio in the presence of typical T10 variation and B1 inhomogeneity. At changes of relaxation rate (R1) of 8 s-1, the differences between a T10 of 1266 and 566 ms are <1, 12, and 58%, respectively, for EFSRSF, EF'SRSF, and ER, whereas differences of 20, 8, and 51%, respectively, result from a 50% B1 field reduction at the same R1. These quantification techniques may be a solution to minimize the effect of T10 variation and B1 inhomogeneity on dynamic contrast-enhanced MRI of the breast at 3 T. Magn Reson Med, 2012. (C) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:531 / 540
页数:10
相关论文
共 26 条
[1]  
AZLAN CA, 2010, P JOINT ANN M ISMRM, P2492
[2]   B1 Transmission-Field Inhomogeneity and Enhancement Ratio Errors in Dynamic Contrast-Enhanced MRI (DCE-MRI) of the Breast at 3T [J].
Azlan, Che A. ;
Di Giovanni, Pierluigi ;
Ahearn, Trevor S. ;
Semple, Scott I. K. ;
Gilbert, Fiona J. ;
Redpath, Thomas W. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (01) :234-239
[3]   MR CHARACTERIZATION OF SUSPICIOUS BREAST-LESIONS WITH A GADOLINIUM-ENHANCED TURBOFLASH SUBTRACTION TECHNIQUE [J].
BOETES, C ;
BARENTSZ, JO ;
MUS, RD ;
VANDERSLUIS, RF ;
VANERNING, LJTO ;
HENDRIKS, JHCL ;
HOLLAND, R ;
RUYS, SHJ .
RADIOLOGY, 1994, 193 (03) :777-781
[4]  
Brookes JA, 1999, JMRI-J MAGN RESON IM, V9, P163, DOI 10.1002/(SICI)1522-2586(199902)9:2<163::AID-JMRI3>3.0.CO
[5]  
2-L
[6]   The accuracy of pharmacokinetic parameter measurement in DCE-MRI of the breast at 3 T [J].
Di Giovanni, P. ;
Azlan, C. A. ;
Ahearn, T. S. ;
Semple, S. I. ;
Gilbert, F. J. ;
Redpath, T. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :121-132
[7]  
GRANT DM, 1996, ENCY NUCL MAGNETIC R, P1396
[8]   Contrast-enhanced MRI of the breast: Accuracy, value, controversies, solutions [J].
HeywangKobrunner, SH ;
Viehweg, P ;
Heinig, A ;
Kuchler, C .
EUROPEAN JOURNAL OF RADIOLOGY, 1997, 24 (02) :94-108
[9]   METHOD FELL THE QUANTITATIVE ASSESSMENT OF CONTRAST AGENT UPTAKE IN DYNAMIC CONTRAST-ENHANCED MRT [J].
HITTMAIR, K ;
GOMISCEK, G ;
LANGENBERGER, K ;
RECHT, M ;
IMHOF, H ;
KRAMER, J .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (05) :567-571
[10]   PHARMACOKINETIC MAPPING OF THE BREAST - A NEW METHOD FOR DYNAMIC MR MAMMOGRAPHY [J].
HOFFMANN, U ;
BRIX, G ;
KNOPP, MV ;
HESS, T ;
LORENZ, WJ .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (04) :506-514