CT-guided percutaneous lung biopsy: Comparison of conventional CT fluoroscopy to CT fluoroscopy with electromagnetic navigation system in 60 consecutive patients

被引:32
作者
Grand, David Justin [1 ]
Atalay, Michael A. [1 ]
Cronan, John J. [1 ]
Mayo-Smith, William W. [1 ]
Dupuy, Damian E. [1 ]
机构
[1] Brown Univ, Dept Diagnost Imaging, Warren Alpert Sch Med, Providence, RI 02903 USA
关键词
CT fluoroscopy; Electromagnetic Navigation; Lung biopsy; RADIOFREQUENCY ABLATION;
D O I
10.1016/j.ejrad.2011.05.030
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine if use of an electromagnetic navigation system (EMN) decreases radiation dose and procedure time of CT fluoroscopy guided lung biopsy in lesions smaller than 2.5 cm. Materials/methods: 86 consecutive patients with small lung masses (<2.5 cm) were approached. 60 consented and were randomized to undergo biopsy with CT fluoroscopy (CTF) (34 patients) or EMN (26 patients). Technical failure required conversion to CTF in 8/26 EMN patients; 18 patients completed biopsy with EMN. Numerous biopsy parameters were compared as described below. Results: Average fluoroscopy time using CTF was 28.2 s compared to 35.0 s for EMN (p = 0.1). Average radiation dose was 117 mGy using CTF and 123 mGy for EMN (p = 0.7). Average number of needle repositions was 3.7 for CTF and 4.4 for EMN (p = 0.4). Average procedure time was 15 min for CTF and 20 min for EMN (p = 0.01). There were 7 pneumothoracesin the CTF group and 6 pneumothoraces in the EMN group (p = 0.7). One pneumothorax in the CTF group and 3 pneumothoraces in the EMN group required chest tube placement (p = 0.1). One pneumothorax patient in each group required hospital admission. Diagnostic specimens were obtained in 31/34 patients in the CTF group and 22/26 patients in the EMN group (p = 0.4). Conclusions: EMN was not statistically different than CTF for fluoroscopy time, radiation dose, number of needle repositions, incidence of pneumothorax, need for chest tube, or diagnostic yield. Procedure time was increased with EMN. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:E133 / E136
页数:4
相关论文
共 10 条
  • [1] Electromagnetic Navigation for Thoracic Aortic Stent-graft Deployment: A Pilot Study in Swine
    Abi-Jaoudeh, Nadine
    Glossop, Neil
    Dake, Michael
    Pritchard, William F.
    Chiesa, Alberto
    Dreher, Matthew R.
    Tang, Thomas
    Karanian, John W.
    Wood, Bradford J.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (06) : 888 - 895
  • [2] Electromagnetic Navigation System for CT-Guided Biopsy of Small Lesions
    Appelbaum, Liat
    Sosna, Jacob
    Nissenbaum, Yizhak
    Benshtein, Alexander
    Goldberg, S. Nahum
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (05) : 1194 - 1200
  • [3] Benefits and safety of CT fluoroscopy in interventional radiologic procedures
    Carlson, SK
    Bender, CE
    Classic, KL
    Zink, FE
    Quam, JP
    Ward, EM
    Oberg, AL
    [J]. RADIOLOGY, 2001, 219 (02) : 515 - 520
  • [4] Analysis of the factors associated with radiofrequency ablation-induced pneumothorax
    Gillams, A. R.
    Lees, W. R.
    [J]. CLINICAL RADIOLOGY, 2007, 62 (07) : 639 - 644
  • [5] Electromagnetic catheter navigation during bronchoscopy - Validation of a novel method by conventional fluoroscopy
    Hautmann, H
    Schneider, A
    Pinkau, T
    Peltz, F
    Feussner, H
    [J]. CHEST, 2005, 128 (01) : 382 - 387
  • [6] ITO E, 2010, NEUROSURGERY, V16
  • [7] Electromagnetic tracking for thermal ablation and biopsy guidance:: Clinical evaluation of spatial accuracy
    Kruecker, Jochen
    Xu, Sheng
    Glossop, Neil
    Viswanathan, Anand
    Borgert, Joern
    Schulz, Heinrich
    Wood, Bradford J.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 18 (09) : 1141 - 1150
  • [8] Electromagnetic Navigation to Aid Radiofrequency Ablation and Biopsy of Lung Tumors
    Santos, Ricardo S.
    Gupta, Avneesh
    Ebright, Michael I.
    DeSimone, Michael
    Steiner, Gregory
    Estrada, Mary-Jane
    Daly, Benedict
    Fernando, Hiran C.
    [J]. ANNALS OF THORACIC SURGERY, 2010, 89 (01) : 265 - 268
  • [9] Experimental nonfluoroscopic placement of inferior vena cava filters: Use of an electromagnetic navigation system with previous CT data
    Solomon, SB
    Magee, CA
    Acker, DE
    Venbrux, AC
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (01) : 92 - 95
  • [10] Patient and staff dose during CT guided biopsy, drainage and coagulation
    Teeuwisse, WM
    Geleijns, J
    Broerse, JJ
    Obermann, WR
    Van Meerten, ELV
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (884) : 720 - 726