High-density CMOS Microelectrode Array System for Impedance Spectroscopy and Imaging of Biological Cells

被引:0
作者
Vijay, Viswam [1 ]
Raziyeh, Bounik [1 ]
Amir, Shadmani [1 ]
Jelena, Dragas [1 ]
Alicia, Boos Julia [1 ]
Axel, Birchler [1 ]
Jan, Muller [1 ]
Chen Yihui [1 ,2 ]
Andreas, Hierlemann [1 ]
机构
[1] Swiss Fed Inst Technol, Dept Biosyst Sci & Engn, Basel, Switzerland
[2] Analog Devices Shanghai Co Ltd, Shanghai, Peoples R China
来源
2016 IEEE SENSORS | 2016年
基金
欧洲研究理事会;
关键词
CMOS Microelectrode Array (MEA); Electrochemical Impedance Spectroscopy; Electrophysiology; Lock-in Amplifier; CHANNELS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A monolithic measurement platform was implemented to enable label-free in-vitro electrical impedance spectroscopy measurements of cells on multi-functional CMOS microelectrode array. The array includes 59,760 platinum microelectrodes, densely packed within a 4.5 mm 2.5 mm sensing region at a pitch of 13.5 m. The 32 on-chip lock-in amplifiers can be used to measure the impedance of any arbitrarily chosen electrodes on the array by applying a sinusoidal voltage, generated by an on-chip waveform generator with a frequency range from 1 Hz to 1 MHz, and measuring the respective current. Proof-ofconcept measurements of impedance sensing and imaging are shown in this paper. Correlations between cell detection through optical microscopy and electrochemical impedance scanning were established.
引用
收藏
页数:3
相关论文
共 7 条
  • [1] CMOS high density electrical impedance biosensor array for tumor cell detection
    Chen, Yu
    Wong, Chee Chung
    Pui, Tze Sian
    Nadipalli, Revanth
    Weerasekera, Roshan
    Chandran, Jegatha
    Yu, Hao
    Rahman, Abdur R. A.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2012, 173 : 903 - 907
  • [2] A Multi-Modality CMOS Sensor Array for Cell-Based Assay and Drug Screening
    Chi, Taiyun
    Park, Jong Seok
    Butts, Jessica C.
    Hookway, Tracy A.
    Su, Amy
    Zhu, Chengjie
    Styczynski, Mark P.
    McDevitt, Todd C.
    Wang, Hua
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2015, 9 (06) : 801 - 814
  • [3] Label-free impedimetric immunosensor for ultrasensitive detection of cancer marker Murine double minute 2 in brain tissue
    Elshafey, Reda
    Tlili, Chaker
    Abulrob, Abedelnasser
    Tavares, Ana C.
    Zourob, Mohammed
    [J]. BIOSENSORS & BIOELECTRONICS, 2013, 39 (01) : 220 - 225
  • [4] Guo JJ, 2015, IEEE WCNC, P1, DOI 10.1109/WCNC.2015.7127435
  • [5] A CMOS Electrochemical Impedance Spectroscopy (EIS) Biosensor Array
    Manickam, Arun
    Chevalier, Aaron
    McDermott, Mark
    Ellington, Andrew D.
    Hassibi, Arjang
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2010, 4 (06) : 379 - 390
  • [6] Nguemo F., 2015, LABEL FREE BIOSENSOR, P283
  • [7] Viswam V, 2016, ISSCC DIG TECH PAP I, V59, P394, DOI 10.1109/ISSCC.2016.7418073