Postimplantation MRI with cylindric and cubic intervertebral test implants: evaluation of implant shape, material, and volume in MRI artifacting - an in vitro study

被引:17
作者
Ernstberger, Thorsten
Heidrich, Gabert
Buchhorn, Gottfried
机构
[1] Univ Gottingen, Dept Orthopaed Surg, D-37075 Gottingen, Germany
[2] Univ Gottingen, Dept Diagnost Radiol, D-37075 Gottingen, Germany
[3] Univ Gottingen, Dept Orthoped Surg, Biomat Lab, D-37075 Gottingen, Germany
关键词
interbody test implants; MRI artifacting; implant parameters; total artifact volume;
D O I
10.1016/j.spinee.2006.03.016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND CONTEXT: Interbody spacers for anterior spine fusion are made of different materials, which can affect the postfusion magnetic resonance imaging (MRI) scans. Susceptibility artifacts specially for metallic implants can decrease the image quality. PURPOSE: This study focused on the influence of determined implant parameters like shape, implant volume, and implant material in MRI artifacting with regard to solid geometrical titanium and carbon test spacers. STUDY DESIGN/SETTING: A comparative study on an in vitro spine model. The study was performed at a university research facility. METHODS: In this study we evaluated the postimplantation MRI scans of three determined cuboids and six cylinders, which were made of titanium alloy and carbon fiber reinforced polymer. MRI scans were carried out by using T1 turbo spin echo (TSE) sequences. The total artifact volume (TAV) as well as median artifact area (MAA) were calculated. Additionally the implant volume (IV)/TAV and cross-sectional area (CSA)/MAA relation were determined. Statistical analyses were calculated with the t test and Newman-Keuls test for multiple comparisons. RESULTS: Considering all test implants with an increasing implant size, the TAV and the MAA became significantly larger (p <.001) with simultaneous reduction of the IV/TAV and CSA/MAA relation. In contrast to the carbon implant group, for titanium cylinders with an implant volume equivalent to the cuboids significant differences in MRI artifacting (p <.05) could be demonstrated. CONCLUSIONS: Susceptibility artifacts were clearly affected by the implant material, shape, and implant volume. Independent of the implant material, with regard to a more advantageous IV/TAV and CSA/MAA relation, for larger implants the artifact rate was more limited to the implant's direct surroundings. On the basis of a high magnetizability for titanium implants with an equal IV, the range of implant-related susceptibility artifacts was influenced by the material as well as implant shape in favor of cylindric implants. According to the very low MRI artifact rate of carbon implants, the implant shape did not have any significant effect on the artifact behavior. (c) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:353 / 359
页数:7
相关论文
共 19 条
[1]   MAGNETIC-RESONANCE-IMAGING - CARDIAC EJECTION FRACTION MEASUREMENTS - PHANTOM STUDY COMPARING 4 DIFFERENT METHODS [J].
DEBATIN, JF ;
NADEL, SN ;
SOSTMAN, HD ;
SPRITZER, CE ;
EVANS, AJ ;
GRIST, TM .
INVESTIGATIVE RADIOLOGY, 1992, 27 (03) :198-204
[2]  
ERNSTBERGER T, 2006, EUR SPINE J, V7, P1
[3]   Artifacts in MRI of the temporomandibular joint caused by dental alloys: A phantom study at 1.5 T [J].
Fellner, C ;
Behr, M ;
Fellner, F ;
Held, P ;
Handel, G ;
Feuerbach, S .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 166 (05) :421-428
[4]   REDUCTION OF ARTIFACTS IN MAGNETIC-RESONANCE IMAGES BY USING OPTIMIZED MATERIALS FOR DIAGNOSTIC DEVICES AND IMPLANTS [J].
FRITZSCHE, S ;
THULL, R ;
HAASE, A .
BIOMEDIZINISCHE TECHNIK, 1994, 39 (03) :42-46
[5]  
Goulet JA, 1997, CLIN ORTHOP RELAT R, P76
[6]  
Henk C B, 1999, Top Magn Reson Imaging, V10, P247, DOI 10.1097/00002142-199908000-00006
[7]   Influence of sequence type on the extent of the susceptibility artifact in MRI - a shoulder specimen study after suture anchor repair [J].
Herold, T ;
Caro, WC ;
Heers, G ;
Perlick, L ;
Grifka, J ;
Feuerbach, S ;
Nitz, W ;
Lenhart, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (09) :1296-1301
[8]   A comparative study of MR imaging profile of titanium pedicle screws [J].
Malik, AS ;
Boyko, O ;
Aktar, N ;
Young, WF .
ACTA RADIOLOGICA, 2001, 42 (03) :291-293
[9]   Postoperative magnetic resonance imaging with titanium implants of the thoracic and lumbar spine [J].
Ortiz, O ;
Pait, TG ;
McAllister, P ;
Sauter, K .
NEUROSURGERY, 1996, 38 (04) :741-745
[10]   Optimizing imaging parameters for MR evaluation of the spine with titanium pedicle screws [J].
Petersilge, CA ;
Lewin, JS ;
Duerk, JL ;
Yoo, JU ;
Ghaneyem, AJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (05) :1213-1218