摘要
Protective coatings were successfully formed on biomedical NiTi alloy by micro-arc oxidation (MAO) using pulsed bipolar power supply. The coating surface exhibits a typical MAO porous structure, and the coating mainly consists of O, Al, Ti and Ni, with the atomic concentration of 65.11%, 27.77%, 2.20% and 2.8%, respectively. The thickness of MAO coating is about 24.0 mu m when the duration time of the MAO treatment was 90 min at 400 V constant voltage treatment. XRD analysis showed that micro-arc oxidation coating is composed of gamma-Al2O3 and alpha-Al2O3. The wear resistance of the coatings was investigated by ball-on-disk friction test. The microhardness of the NiTi alloy is greatly enhanced due to the formation of Al2O3 coating after micro-arc oxidation treatment. The friction coefficient of the coated NiTi is stable at 0.85 and the wear resistance is increased by 9 times compared with uncoated NiTi. The wear mechanism transforms from abrasive-dominant for the uncoated sample to adhesive-dominant for the coated sample. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:391 / 394
页数:4
相关论文
共 23 条
[1]  
Brailovski V, 1996, BIO-MED MATER ENG, V6, P291
[2]  
CHEN FL, 2007, J ALLOY COMPD, V100, P238
[3]   The corrosion and nickel release behavior of laser surface-melted NiTi shape memory alloy in Hanks, solution [J].
Cui, ZD ;
Man, HC ;
Yang, XJ .
SURFACE & COATINGS TECHNOLOGY, 2005, 192 (2-3) :347-353
[4]   Biomimetic deposition of apatite coating on surface-modified NiTi alloy [J].
Gu, YW ;
Tay, BY ;
Lim, CS ;
Yong, MS .
BIOMATERIALS, 2005, 26 (34) :6916-6923
[5]   Spark anodizing of β-Ti alloy for wear-resistant coating [J].
Habazaki, H. ;
Onodera, T. ;
Fushimi, K. ;
Konno, H. ;
Toyotake, K. .
SURFACE & COATINGS TECHNOLOGY, 2007, 201 (21) :8730-8737
[6]   Improved biological performance of Ti implants due to surface modification by micro-arc oxidation [J].
Li, LH ;
Kong, YM ;
Kim, HW ;
Kim, YW ;
Kim, HE ;
Heo, SJ ;
Koak, JY .
BIOMATERIALS, 2004, 25 (14) :2867-2875
[7]   Improvement of corrosion properties of microarc oxidation coating on magnesium alloy by optimizing current density parameters [J].
Liang, Jun ;
Hu, Litian ;
Hao, Jingcheng .
APPLIED SURFACE SCIENCE, 2007, 253 (16) :6939-6945
[8]   Formation of hydroxyapatite on Ti-6Al-4V alloy by microarc oxidation and hydrothermal treatment [J].
Liu, F ;
Wang, FP ;
Shimizu, T ;
Igarashi, K ;
Zhao, LC .
SURFACE & COATINGS TECHNOLOGY, 2005, 199 (2-3) :220-224
[9]   Sol-gel deposited TiO2 film on NiTi surgical alloy for biocompatibility improvement [J].
Liu, JX ;
Yang, DZ ;
Shi, F ;
Cai, YJ .
THIN SOLID FILMS, 2003, 429 (1-2) :225-230
[10]   Surface characteristics, biocompatibility, and mechanical properties of nickel-titanium plasma-implanted with nitrogen at different implantation voltages [J].
Liu, X. M. ;
Wu, S. L. ;
Chan, Y. L. ;
Chu, Paul K. ;
Chung, C. Y. ;
Chu, C. L. ;
Yeung, K. W. K. ;
Lu, W. W. ;
Cheung, K. M. C. ;
Luk, K. D. K. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2007, 82A (02) :469-478