Motion Correction in MR Thermometry of Abdominal Organs: a Comparison of the Referenceless vs. the Multibaseline Approach

被引:49
作者
de Senneville, Baudouin Denis [1 ]
Roujol, Sebastien [1 ,2 ]
Moonen, Chrit [1 ]
Ries, Mario [1 ]
机构
[1] Univ Bordeaux 2, UMR 5231, Lab Mol & Funct Imaging Physiol Therapy, CNRS, F-33076 Bordeaux, France
[2] Univ Bordeaux 1, LaBRI, UMR 5800, CNRS, F-33405 Talence, France
关键词
artifacts; technique development; technical research; correction; susceptibility; FOCUSED ULTRASOUND; LIVER;
D O I
10.1002/mrm.22514
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reliable temperature and thermal-dose measurements using proton resonance frequency shift-based magnetic resonance (MR) thermometry for MR-guided ablation of abdominal organs require a robust correction of artefacts induced by the target displacement through an inhomogeneous and time-variant magnetic field. Two correction approaches emerged recently as promising candidates to allow continuous real-time MR-thermometry under free-breathing conditions: The multibaseline correction method, which relies on a pre-recorded correction table allowing to correct for periodic phase changes, and the referenceless method, which depends on a background phase estimation in the target area based on the assumption of a smooth spatial variation of the phase across the organ. This study combines both methods with real-time in-plane motion correction to permit both temperature and thermal-dose calculations on the fly. Subsequently, the practical aspects of both methods are compared in two application scenarios, a radio frequency-ablation and a high-intensity focused ultrasound ablation. A hybrid approach is presented that exploits the strong points of both methods, allowing accurate and precise proton resonance frequency-thermometry measurements during periodical displacement, even in the presence of spontaneous motion and strong susceptibility variations in the target area. Magn Reson Med 64:1373-1381, 2010. (C)2010 Wiley-Liss, Inc.
引用
收藏
页码:1373 / 1381
页数:9
相关论文
共 18 条
[1]   Understanding phase maps in MRI: A new cutline phase unwrapping method [J].
Chavez, S ;
Xiang, QS ;
An, L .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (08) :966-977
[2]   SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS [J].
CONTURO, TE ;
SMITH, GD .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :420-437
[3]  
CORNELIUS N, 1983, P ACM SIGGRAPH SIGAR
[4]   Real-time adaptive methods for treatment of mobile organs by MRI-controlled high-intensity focused ultrasound [J].
de Senneville, Baudouin Denis ;
Mougenot, Charles ;
Moonen, Chrit T. W. .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (02) :319-330
[5]  
de Zwart JA, 2001, MAGNET RESON MED, V45, P128, DOI 10.1002/1522-2594(200101)45:1<128::AID-MRM1017>3.0.CO
[6]  
2-M
[7]  
DESENNEVILLE BD, 2007, IEEE INT C IM PROC S, V3, P137
[8]   Monitoring and correcting spatio-temporal variations of the MR scanner's static magnetic field [J].
El-Sharkawy, Abdel Monem ;
Schar, Michael ;
Bottomley, Paul A. ;
Atalar, Ergin .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2006, 19 (05) :223-236
[9]  
Friston K., 1995, HUM BRAIN MAP, V2, P165
[10]   Real-Time MR Thermometry for Monitoring HIFU Ablations of the Liver [J].
Holbrook, Andrew B. ;
Santos, Juan M. ;
Kaye, Elena ;
Rieke, Viola ;
Pauly, Kim Butts .
MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (02) :365-373