Monoblock versus modular polyethylene insert in uncemented total knee arthroplasty: A randomized RSA study with 2-year follow-up of 53 patients

被引:11
作者
Andersen, Mikkel Rathsach [1 ,2 ]
Winther, Nikolaj [1 ]
Lind, Thomas [2 ]
Schroder, Henrik [1 ]
Flivik, Gunnar [3 ]
Petersen, Michael Mork [1 ]
机构
[1] Univ Copenhagen, Dept Orthoped, Rigshosp, Copenhagen, Denmark
[2] Univ Copenhagen, Dept Orthoped, Herlev Gentofte Hosp, Copenhagen, Denmark
[3] Lund Univ, Skane Univ Hosp, Dept Orthoped, Clin Sci Lund, Lund, Sweden
关键词
CEMENTED TIBIAL COMPONENTS; TRABECULAR METAL; RADIOSTEREOMETRIC ANALYSIS; PERIPROSTHETIC OSTEOLYSIS; FIXATION; MECHANISMS; WEAR; REVISION; REPLACEMENTS; PROSTHESIS;
D O I
10.1080/17453674.2016.1233654
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose Backside wear of the polyethylene insert in total knee arthroplasty (TKA) can produce clinically significant levels of polyethylene debris, which can lead to loosening of the tibial component. Loosening due to polyethylene debris could theoretically be reduced in tibial components of monoblock polyethylene design, as there is no backside wear. We investigated the effect of 2 different tibial component designs, monoblock and modular polyethylene, on migration of the tibial component in uncemented TKA.Patients and methods In this randomized study, 53 patients (mean age 61 years), 32 in the monoblock group and 33 in the modular group, were followed for 2 years. Radiostereometric analysis (RSA) was done postoperatively after weight bearing and after 3, 6, 12, and 24 months. The primary endpoint of the study was comparison of the tibial component migration (expressed as maximum total point motion (MTPM)) of the 2 different implant designs.Results We did not find any statistically significant difference in MTPM between the groups at 3 months (p = 0.2) or at 6 months (p = 0.1), but at 12 and 24 months of follow-up there was a significant difference in MTPM of 0.36mm (p = 0.02) and 0.42mm (p = 0.02) between groups, with the highest amount of migration (1.0mm) in the modular group. The difference in continuous migration (MTPM from 12 and 24 months) between the groups was 0.096mm (p = 0.5), and when comparing MTPM from 3-24 months, the difference between the groups was 0.23mm (p = 0.07).Interpretation In both study groups, we found the early migration pattern expected, with a relatively high initial amount of migration from operation to 3 months of follow-up, followed by stabilization of the implant with little migration thereafter. However, the modular implants had a statistically significantly higher degree of migration compared to the monoblock. We believe that the greater stiffness of the modular implants was the main reason for the difference in migration, but an initial creep in the polyethylene metal-back locking mechanism of the modular group could also be a possible explanation for the observed difference in migration between the 2 study groups.
引用
收藏
页码:607 / 614
页数:8
相关论文
共 39 条
[1]   Low-conforming all-polyethylene tibial component not inferior to metal-backed component in cemented total knee arthroplasty -: Prospective, randomized radiostereometric analysis study of the AGC total knee prosthesis [J].
Adalberth, G ;
Nilsson, KG ;
Byström, S ;
Kolstad, K ;
Milbrink, J .
JOURNAL OF ARTHROPLASTY, 2000, 15 (06) :783-792
[2]   THE EFFECT OF A STEM ON THE TIBIAL COMPONENT OF KNEE ARTHROPLASTY - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY OF UNCEMENTED TIBIAL COMPONENTS IN THE FREEMAN-SAMUELSON KNEE ARTHROPLASTY [J].
ALBREKTSSON, BEJ ;
RYD, L ;
CARLSSON, LV ;
FREEMAN, MAR ;
HERBERTS, P ;
REGNER, L ;
SELVIK, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (02) :252-258
[3]   The effectiveness of isolated tibial insert exchange in revision total knee arthroplasty [J].
Babis, GC ;
Trousdale, RT ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (01) :64-68
[4]   The stability of the cemented tibial component of total knee arthroplasty [J].
Catani, F ;
Leardini, A ;
Ensini, A ;
Cucca, G ;
Bragonzoni, L ;
Toksvig-Larsen, S ;
Giannini, S .
JOURNAL OF ARTHROPLASTY, 2004, 19 (06) :775-782
[5]   Evaluating the Progression of Osteolysis After Total Knee Arthroplasty [J].
Dalling, Jason G. ;
Math, Kevin ;
Scuderi, Giles R. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2015, 23 (03) :173-180
[6]   PATIENT-REPORTED OUTCOME MEASURES IN THE NHS: NEW METHODS FOR ANALYSING AND REPORTING EQ-5D DATA [J].
Devlin, Nancy J. ;
Parkin, David ;
Browne, John .
HEALTH ECONOMICS, 2010, 19 (08) :886-905
[7]   Fixation of a Trabecular Metal Knee Arthroplasty Component A Prospective Randomized Study [J].
Dunbar, M. J. ;
Wilson, D. A. J. ;
Hennigar, A. W. ;
Amirault, J. D. ;
Gross, M. ;
Reardon, G. P. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (07) :1578-1586
[8]   Clinical results of modular polyethylene insert exchange with retention of total knee arthroplasty components [J].
Engh, GA ;
Koralewicz, LM ;
Pereles, TR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (04) :516-523
[9]  
FENG EL, 1994, CLIN ORTHOP RELAT R, P60
[10]   Osteolysis around total knee arthroplasty: A review of pathogenetic mechanisms [J].
Gallo, J. ;
Goodman, S. B. ;
Konttinen, Y. T. ;
Wimmer, M. A. ;
Holinka, M. .
ACTA BIOMATERIALIA, 2013, 9 (09) :8046-8058