摘要
X-ray lenses and X-ray tubes that are produced at IRO in collaboration with other companies should be certified in accordance with standards. In the report, five different measurement techniques for experimental estimation of focal spot sizes in X-ray tubes are considered. They are: the pinhole, the slit, the edge, the scanning and the grid- wire methods. The same techniques can be used for experimental estimation of focal spot sizes in X-ray lenses. Some experiments and numerical simulation have been performed to estimate accuracy of the above methods. We are developing a simple and reliable wire method. The arrangement consists of an X-ray tube under investigation (or an X-ray tube and an X-ray lens under investigation), a set of W-wires with diameters from 0.03mm to 0.3mm, and X-ray detectors (the CCD- camera, scintillation or semiconductor detector). Agreement between the wire and edge methods of measurements is rather good, while three others give only rough estimations when we used them for measuring focal spots in the range from 0.01mm to 0.05mm.
引用
收藏
页码:326 / 335
页数:10
相关论文
共 50 条
  • [1] Investigation of micro-focus X-ray tube for radiography
    Shovkun, VY
    Kumakhov, MA
    Pirshin, IV
    Gelever, VD
    X-Ray and Neutron Capillary Optics II, 2005, 5943 : 255 - 259
  • [2] MEASUREMENT OF FOCAL SPOT SIZE IN MAMMOGRAPHY X-RAY TUBES
    LAW, J
    BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (781) : 44 - 50
  • [3] X-Ray Refractive Large Aperture Rolled Prism Lenses as Condensers for X-Ray Tubes
    Vogt, H.
    Simon, M.
    Last, A.
    Marschall, F.
    Mohr, J.
    Nazmov, V.
    Eisenhower, R.
    Mettendorf, K. U.
    OPTICAL DESIGN AND ENGINEERING IV, 2011, 8167
  • [4] Measurement of the Parameters of the Focal Spot of an x-Ray Tube Using Kumakhov Optics
    A. Yu. Romanov
    Measurement Techniques, 2004, 47 : 670 - 674
  • [5] Comparison of different methods for determining the size of a focal spot of microfocus X-ray tubes
    Salamon, M.
    Hanke, R.
    Krueger, P.
    Sukowski, F.
    Uhlmann, N.
    Voland, V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 591 (01) : 54 - 58
  • [6] Fabrication and evaluation of variable focus X-ray lenses
    Khounsary, Ali
    Dufresne, Eric M.
    Kewish, Cameron M.
    Qian, Jun
    Assoufid, Lahsen
    Conley, Ray
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 582 (01) : 117 - 119
  • [7] Evaluation of the effective focal spot size of x-ray tubes by utilizing edge response analysis
    Nishiki, Masayuki
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [8] Quantitative method of measuring spot size of microfocus X-ray generator
    Uchida, Hiroyuki
    Hasuike, Kazuto
    Torii, Ken'ichi
    Tsunemi, Hiroshi
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2006, 45 (6A): : 5277 - 5279
  • [9] Partial volume correction for approximating crack opening displacements in CFRP material obtained from micro-focus X-ray CT scans
    Bull, D. J.
    Sinclair, I.
    Spearing, S. M.
    COMPOSITES SCIENCE AND TECHNOLOGY, 2013, 81 : 9 - 16
  • [10] Research on Cylinder Method for Measuring Focal Spot Size of X-ray Tube
    Li W.
    Wang L.
    Zheng J.
    Xu S.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2022, 56 (10): : 2155 - 2164