Repetitively Mode-Locked Cavity-Enhanced Absorption Spectroscopy (RML-CEAS) for Near-Infrared Gas Sensing

被引:10
作者
He, Qixin [1 ,2 ]
Lou, Minhan [1 ]
Zheng, Chuantao [1 ,2 ]
Ye, Weilin [1 ,3 ]
Wang, Yiding [2 ]
Tittel, Frank K. [1 ]
机构
[1] Rice Univ, Dept Elect & Comp Engn, 6100 Main St, Houston, TX 77005 USA
[2] Jilin Univ, Coll Elect Sci & Engn, State Key Lab Integrated Optoelect, 2699 Qianjin St, Changchun 130012, Jilin, Peoples R China
[3] Shantou Univ, Coll Engn, 243 Daxue Rd, Shantou 515063, Peoples R China
基金
美国国家科学基金会; 中国国家自然科学基金;
关键词
infrared absorption spectroscopy; cavity enhanced absorption spectroscopy; gas sensor; SPECTROMETER; BAND; NOISE;
D O I
10.3390/s17122792
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A Pound-Drever-Hall (PDH)-based mode-locked cavity-enhanced sensor system was developed using a distributed feedback diode laser centered at 1.53 mu m as the laser source. Laser temperature scanning, bias control of the piezoelectric ceramic transducer (PZT) and proportional-integral-derivative (PID) feedback control of diode laser current were used to repetitively lock the laser modes to the cavity modes. A gas absorption spectrum was obtained by using a series of absorption data from the discrete mode-locked points. The 15 cm-long Fabry-Perot cavity was sealed using an enclosure with an inlet and outlet for gas pumping and a PZT for cavity length tuning. The performance of the sensor system was evaluated by conducting water vapor measurements. A linear relationship was observed between the measured absorption signal amplitude and the H2O concentration. A minimum detectable absorption coefficient of 1.5 x 10(-8) cm(-1) was achieved with an averaging time of 700 s. This technique can also be used for the detection of other trace gas species by targeting the corresponding gas absorption line.
引用
收藏
页数:12
相关论文
共 30 条
[11]   Detection of HO2 in an atmospheric pressure plasma jet using optical feedback cavity-enhanced absorption spectroscopy [J].
Gianella, Michele ;
Reuter, Stephan ;
Aguila, Ana Lawry ;
Ritchie, Grant A. D. ;
van Helden, Jean-Pierre H. .
NEW JOURNAL OF PHYSICS, 2016, 18
[12]   Cavity-Enhanced Near-Infrared Laser Absorption Spectrometer for the Measurement of Acetonitrile in Breath [J].
Gianella, Michele ;
Ritchie, Grant A. D. .
ANALYTICAL CHEMISTRY, 2015, 87 (13) :6881-6889
[13]   A quantum cascade laser-based optical feedback cavity-enhanced absorption spectrometer for the simultaneous measurement of CH4 and N2O in air [J].
Hamilton, D. J. ;
Orr-Ewing, A. J. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2011, 102 (04) :879-890
[14]   Cavity enhanced absorption spectroscopy using a broadband prism cavity and a supercontinuum source [J].
Johnston, Paul S. ;
Lehmann, Kevin K. .
OPTICS EXPRESS, 2008, 16 (19) :15013-15023
[15]   High pressure Cavity Ring Down Spectroscopy: Application to the absorption continuum of CO2 near 1.7 μm [J].
Kassi, S. ;
Campargue, A. ;
Mondelain, D. ;
Tran, H. .
JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2015, 167 :97-104
[16]   Off-axis continuous-wave cavity-enhanced absorption spectroscopy of narrow-band and broadband absorbers using red diode lasers [J].
Kasyutich, VL ;
Canosa-Mas, CE ;
Pfrang, C ;
Vaughan, S ;
Wayne, RP .
APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (6-7) :755-761
[17]   LASER BEAMS AND RESONATORS [J].
KOGELNIK, H ;
LI, T .
PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1966, 54 (10) :1312-+
[18]   Self-referenced, accurate and sensitive optical frequency comb spectroscopy with a virtually imaged phased array spectrometer [J].
Kowzan, Grzegorz ;
Lee, Kevin F. ;
Paradowska, Magdalena ;
Borkowski, Mateusz ;
Ablewski, Piotr ;
Wojtewicz, Szymon ;
Stec, Kamila ;
Lisak, Daniel ;
Fermann, Martin E. ;
Trawinski, Ryszard S. ;
Maslowski, Piotr .
OPTICS LETTERS, 2016, 41 (05) :974-977
[19]   Sensitive CH4 detection applying quantum cascade laser based optical feedback cavity-enhanced absorption spectroscopy [J].
Lang, N. ;
Macherius, U. ;
Wiese, M. ;
Zimmermann, H. ;
Roepcke, J. ;
van Helden, J. H. .
OPTICS EXPRESS, 2016, 24 (06) :A536-A543
[20]   Cavity enhanced absorption spectroscopy of multiple trace gas species using a supercontinuum radiation source [J].
Langridge, J. M. ;
Laurila, T. ;
Watt, R. S. ;
Jones, R. L. ;
Kaminski, C. F. ;
Hult, J. .
OPTICS EXPRESS, 2008, 16 (14) :10178-10188