Magnetic resonance temperature imaging

被引:149
作者
De Senneville, BD
Quesson, B
Moonen, CTW
机构
[1] Univ Bordeaux 2, Lab Mol & Funct Imaging, ERT CNRS, F-33076 Bordeaux, France
[2] Image Guided Therapy, Pessac, France
关键词
magnetic resonance imaging; temperature imaging; thermal therapies; motion artifacts; susceptibility artifacts;
D O I
10.1080/02656730500133785
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Continuous, real-time, 3D temperature mapping during a hyperthermic procedure may provide (i) enhanced safety by visualizing temperature maps in and around the treated region, (ii) improved efficiency by adapting local energy deposition with feedback coupling algorithms and (iii) therapy end-points based on the accumulated thermal dose. Non-invasive mapping of temperature changes can be achieved with MRI and may be based on temperature dependent MRI parameters. The excellent linearity of the temperature dependency of the proton resonance frequency (PRF) and its near-independence with respect to tissue type make the PRF-based methods the preferred choice for many applications, in particular at mid- to-high field strength (>= 0.5 T). The PRF methods employ RF-spoiled gradient echo imaging methods and incorporate fat suppression techniques for most organs. A standard deviation of less than 1 degrees C, for a temporal resolution below 1 s and a spatial resolution of similar to 2 mm is feasible for immobile tissues. Special attention is paid to methods for reducing artifacts in MR temperature mapping caused by intra-scan and inter-scan motion and motion and temperature-induced susceptibility effects in mobile tissues. Real-time image processing and visualization techniques, together with accelerated MRI acquisition techniques, are described because of their potential for therapy guidance.
引用
收藏
页码:515 / 531
页数:17
相关论文
共 83 条
  • [1] A new ytterbium chelate as contrast agent in chemical shift imaging and temperature sensitive probe for MR spectroscopy
    Aime, S
    Botta, M
    Fasano, M
    Terreno, E
    Kinchesh, P
    Calabi, L
    Paleari, L
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) : 648 - 651
  • [2] BARTHOLET A, 2000, SPIN TRANSITION MOL
  • [3] Temperature quantification using the proton frequency shift technique: In vitro and in vivo validation in an open 0.5 tesla interventional MR scanner during RF ablation
    Botnar, RM
    Steiner, P
    Dubno, B
    Erhart, P
    von Schulthess, GK
    Debatin, JF
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (03) : 437 - 444
  • [4] BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
  • [5] THE ESTIMATION OF LOCAL BRAIN TEMPERATURE BY IN-VIVO H-1 MAGNETIC-RESONANCE SPECTROSCOPY
    CADY, EB
    DSOUZA, PC
    PENRICE, J
    LOREK, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (06) : 862 - 867
  • [6] A NEW BLOCK-MATCHING CRITERION FOR MOTION ESTIMATION AND ITS IMPLEMENTATION
    CHEN, MJ
    CHEN, LG
    CHIUEH, TD
    LEE, YP
    [J]. IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS FOR VIDEO TECHNOLOGY, 1995, 5 (03) : 231 - 236
  • [7] Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam
    Chung, AH
    Hynynen, K
    Colucci, V
    Oshio, K
    Cline, HE
    Jolesz, FA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) : 745 - 752
  • [8] Chung YC, 1999, MAGNET RESON MED, V42, P864, DOI 10.1002/(SICI)1522-2594(199911)42:5<864::AID-MRM5>3.0.CO
  • [9] 2-8
  • [10] CONTOUR-BASED MOTION ESTIMATION
    DAVIS, LS
    WU, ZQ
    SUN, HF
    [J]. COMPUTER VISION GRAPHICS AND IMAGE PROCESSING, 1983, 23 (03): : 313 - 326