Design and Rolling Locomotion of a Magnetically Actuated Soft Capsule Endoscope

被引:291
作者
Yim, Sehyuk [1 ]
Sitti, Metin [1 ,2 ]
机构
[1] Carnegie Mellon Univ, Dept Mech Engn, Pittsburgh, PA 15213 USA
[2] Carnegie Mellon Univ, Inst Robot, Pittsburgh, PA 15213 USA
关键词
Capsule endoscopy; magnetic actuation; medical robotics; miniature robotics; soft robotics; MANIPULATION; FABRICATION;
D O I
10.1109/TRO.2011.2163861
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
This paper proposes a magnetically actuated soft capsule endoscope (MASCE) as a tetherless miniature mobile robot platform for diagnostic and therapeutic medical applications inside the stomach. Two embedded internal permanent magnets and a large external magnet are used to actuate the robot remotely. The proposed MASCE has three novel features. First, its outside body is made of soft elastomer-based compliant structures. Such compliant structures can deform passively during the robot-tissue contact interactions, which makes the device safer and less invasive. Next, it can be actively deformed in the axial direction by using external magnetic actuation, which provides an extra degree of freedom that enables various advanced functions such as axial position control, drug releasing, drug injection, or biopsy. Finally, it navigates in three dimensions by rolling on the stomach surface as a new surface locomotion method inside the stomach. Here, the external attractive magnetic force is used to anchor the robot on a desired location, and the external magnetic torque is used to roll it to another location, which provides a stable, continuous, and controllable motion. The paper presents design and fabrication methods for the compliant structures of the robot with its axial deformation and position control capability. Rolling-based surface locomotion of the robot using external magnetic torques is modeled, and its feasibility is tested and verified on a synthetic stomach surface by using a magnetically actuated capsule endoscope prototype.
引用
收藏
页码:183 / 194
页数:12
相关论文
共 30 条
  • [11] Glass P., 2009, THESIS CARNEGIE MELL
  • [12] A Legged Anchoring Mechanism for Capsule Endoscopes Using Micropatterned Adhesives
    Glass, Paul
    Cheung, Eugene
    Sitti, Metin
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (12) : 2759 - 2767
  • [13] Analysis of off-axis performance of compliant mechanisms with applications to mobile millirobot design
    Hoover, A. M.
    Fearing, R. S.
    [J]. 2009 IEEE-RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS, 2009, : 2770 - 2776
  • [14] Implementation of multichannel sensors for remote biomedical measurements in a microsystems format
    Johannessen, EA
    Wang, L
    Cui, L
    Tang, TB
    Ahmadian, M
    Astaras, A
    Reid, SWJ
    Murray, AF
    Flynn, BW
    Beaumont, SP
    Cumming, DRS
    Cooper, JM
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (03) : 525 - 535
  • [15] Johnson K. L., 1987, CONTACT MECH
  • [16] Kyoung-chul Kong, 2005, 2005 IEEE/RSJ International Conference on Intelligent Robots and Systems, P1839
  • [17] Capsule endoscopy: progress update and challenges ahead
    Moglia, Andrea
    Menciassi, Arianna
    Dario, Paolo
    Cuschieri, Alfred
    [J]. NATURE REVIEWS GASTROENTEROLOGY & HEPATOLOGY, 2009, 6 (06) : 353 - 362
  • [18] Paddling based microrobot for capsule endoscopes
    Park, Hyunjun
    Park, Sungjin
    Yoon, Euisung
    Kim, Byungkyu
    Park, Jongoh
    Park, Sukho
    [J]. PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, 2007, : 3377 - 3382
  • [19] Park S., 2007, Journal of Micromechanics and Microengineering, V18, P250
  • [20] Design of a FUR-Sized 12-legged endoscopic capsule robot
    Quirini, Marco
    Webster, Robert J., III
    Menciassi, Axianna
    Dario, Paolo
    [J]. PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, 2007, : 1856 - +