Validation of the Omron M3 Intellisense (HEM-7051-E) Upper Arm Blood Pressure Monitor, for Self-Measurement, according to the European Society of Hypertension International Protocol Revision 2010 in a Stage 3-5 Chronic Kidney Disease Population

被引:22
作者
Akpolat, Tekin [1 ]
Erdem, Emre [1 ]
Aydogdu, Turkan [1 ]
机构
[1] Ondokuz Mayis Univ, Sch Med, Dept Nephrol, TR-55139 Samsun, Turkey
关键词
Arterial stiffness; Chronic kidney disease; Home blood pressure monitoring; Hypertension; Oscillometric sphygmomanometers; Validation protocols; PRACTICE GUIDELINES; MEASURING DEVICES; ACCURACY; HEM-907; CALL;
D O I
10.1159/000330719
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Background: Encouragement of home blood pressure (BP) monitoring has a great potential to improve hypertension control rates. The purpose of this study was to test validation of the Omron M3 Intellisense (HEM-7051-E) upper arm BP measuring monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in stage 3-5 chronic kidney disease (CKD) patients. Methods: 66 patients having CKD stage 3-5 were included in the study. Nine consecutive measurements were made according to the ESH-IP2 protocol. Results: The Omron M3 Intellisense device fulfills the validation criteria of the ESH-1P2 for stage 3-5 CKD patients. Conclusion: Although arterial stiffness can affect accurate BP measurement, there are limited data regarding the use of automated oscillometric devices in CKD. To our knowledge, this is the first study investigating validation of an oscillometric device in stage 3-5 predialysis CKD patients. This study validates the Omron M3 Intellisense upper arm device for stage 3-5 CKD patients. New validation studies investigating other oscillometric sphygmomanometers for CKD patients and involvement of nephrologists in these studies have great potential to increase patient care in CKD. Copyright (C) 2011 S. Karger AG, Basel
引用
收藏
页码:82 / 88
页数:7
相关论文
共 25 条
[21]   Call to action on use and reimbursement for home blood pressure monitoring: A joint scientific statement from the American Heart Association, American Society of Hypertension, and Preventive Cardiovascular Nurses Association [J].
Pickering, Thomas G. ;
Miller, Nancy Houston ;
Ogedegbe, Gbenga ;
Krakoff, Lawrence R. ;
Artinian, Nancy T. ;
Goff, David .
JOURNAL OF CARDIOVASCULAR NURSING, 2008, 23 (04) :299-323
[22]  
Pickering Thomas G., 2005, Hypertension (Baltimore), V45, P142
[23]   Hypertension awareness, treatment, and control in chronic kidney disease [J].
Sarafidis, Pantelis A. ;
Li, Suying ;
Chen, Shu-Cheng ;
Collins, Allan J. ;
Brown, Wendy W. ;
Klag, Michael J. ;
Bakris, George L. .
AMERICAN JOURNAL OF MEDICINE, 2008, 121 (04) :332-340
[24]   Accuracy of oscillometric blood pressure monitoring with concurrent auscultatory blood pressure in hemodialysis patients [J].
Semret, M ;
Zidehsarai, M ;
Agarwal, R .
BLOOD PRESSURE MONITORING, 2005, 10 (05) :249-255
[25]   Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy [J].
Thompson, Aliza M. ;
Eguchi, Kazuo ;
Reznik, Michael E. ;
Shah, Samir S. ;
Pickering, Thomas G. .
BLOOD PRESSURE MONITORING, 2007, 12 (04) :227-232