Background: Encouragement of home blood pressure (BP) monitoring has a great potential to improve hypertension control rates. The purpose of this study was to test validation of the Omron M3 Intellisense (HEM-7051-E) upper arm BP measuring monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in stage 3-5 chronic kidney disease (CKD) patients. Methods: 66 patients having CKD stage 3-5 were included in the study. Nine consecutive measurements were made according to the ESH-IP2 protocol. Results: The Omron M3 Intellisense device fulfills the validation criteria of the ESH-1P2 for stage 3-5 CKD patients. Conclusion: Although arterial stiffness can affect accurate BP measurement, there are limited data regarding the use of automated oscillometric devices in CKD. To our knowledge, this is the first study investigating validation of an oscillometric device in stage 3-5 predialysis CKD patients. This study validates the Omron M3 Intellisense upper arm device for stage 3-5 CKD patients. New validation studies investigating other oscillometric sphygmomanometers for CKD patients and involvement of nephrologists in these studies have great potential to increase patient care in CKD. Copyright (C) 2011 S. Karger AG, Basel
引用
收藏
页码:82 / 88
页数:7
相关论文
共 25 条
[1]   Managing hypertension using home blood pressure monitoring among haemodialysis patients-a call to action [J].
Agarwal, Rajiv .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2010, 25 (06) :1766-1771
[2]   Out-of-office blood pressure monitoring in chronic kidney disease [J].
Agarwal, Rajiv ;
Peixoto, Aldo J. ;
Santos, Sergio F. F. ;
Zoccali, Carmine .
BLOOD PRESSURE MONITORING, 2009, 14 (01) :2-11
[3]   Home sphygmomanometers: what should a nephrologist know? [J].
Akpolat, Tekin .
JOURNAL OF NEPHROLOGY, 2011, 24 (03) :300-306
[4]   Obesity, hypertension and home sphygmomanometer cuffs [J].
Akpolat, Tekin .
EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2010, 21 (04) :338-340
[5]   Proposal of a method for the evaluation of inaccuracy of home sphygmomanometers [J].
Akpolat, Tekin .
BLOOD PRESSURE MONITORING, 2009, 14 (05) :208-215
[6]   Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E) [J].
Asmar, Roland ;
Khabouth, Jose ;
Topouchian, Jirar ;
El Feghali, Ramzi ;
Mattar, Jimmy .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :49-54
[7]  
Association for the Advancement of Medical Instrumentation, 1993, SP101993 ANSIAAMI
[8]   The Message for World Kidney Day 2009 [J].
Bakris, George L. ;
Ritz, Eberhard .
KIDNEY & BLOOD PRESSURE RESEARCH, 2009, 32 (01) :67-70
[9]   Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension [J].
Coleman, Andrew ;
Steel, Stephen ;
DeGreef, Annemarie ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (03) :163-166
[10]   Determining the accuracy of blood pressure measurement by the Omron HEM-907 before and after hemodialysis [J].
Czarkowski, Marek ;
Staszkow, Monika ;
Kostyra, Kacper ;
Shebani, Zuhier ;
Niemczyk, Stanislaw ;
Matuszkiewicz-Rowinska, Joanna .