Assessment of sample cleanup and matrix effects in the pesticide residue analysis of foods using postcolumn infusion in liquid chromatography-tandem mass spectrometry

被引:95
作者
Kittlaus, Stefan [1 ]
Schimanke, Julia [2 ]
Kempe, Guenther [3 ]
Speer, Karl [2 ]
机构
[1] Joint Analyt Syst GmbH, D-47445 Moers, Germany
[2] Tech Univ Dresden, D-01069 Dresden, Germany
[3] State Lab Hlth & Vet Affairs LUA Saxony, D-01217 Dresden, Germany
关键词
Pesticides; QuEChERS; Matrix effects; Postcolumn infusion; LC-MS/MS; Sample preparation; PRESSURE CHEMICAL-IONIZATION; SOLID-PHASE EXTRACTION; ELECTROSPRAY-IONIZATION; INTERNAL STANDARD; ION SUPPRESSION; LC-MS/MS; MULTIRESIDUE METHOD; HUMAN PLASMA; MS; BIOANALYSIS;
D O I
10.1016/j.chroma.2011.09.054
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Matrix effect profiles can be used to visualize the effect of the sample matrix to the data signals occurring in a chromatogram. In the present study these profiles were generated by postcolumn infusion of a standard pesticide mix with extracts of different food matrices prepared by the QuEChERS method. Complete raw extracts as well as individual clean-up steps were analyzed. This allowed for a detailed comparison of the interferences caused by the matrix effects from various food samples. It also gave an idea about the efficiency of matrix reduction processes. When analyzing the individual clean-up extracts of the QuEChERS method just a slight reduction of matrix effects could be observed from step to step. Matrices causing strong signal effects in the results of the raw extracts also have strong effects after the final clean-up step. Some of the components responsible for the matrix effects show an extremely high retention time. After the injection of extracts from rocket or different types of tea, significant ion suppressions occurred even after rinsing the analytical column for a long time. The experiments have shown that similar matrices can produce different matrix effect profiles. For example, for black teas and green teas significantly different matrix effect profiles were obtained, while the matrix effects of teas within one of these groups were exactly the same. Analogous results could be found for citrus fruits. In order to overcome interfering matrix effects, analytical systems equipped with different electrospray ion sources were tested. Furthermore, profiles of diluted food extracts were generated. Dilution led to a significant decrease in the matrix effects. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:8399 / 8410
页数:12
相关论文
共 35 条
[1]   The ECHO technique -: the more effective way of data evaluation in liquid chromatography-tandem mass spectrometry analysis [J].
Alder, L ;
Lüderitz, S ;
Lindtner, K ;
Stan, HJ .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1058 (1-2) :67-79
[2]   Residue analysis of 500 high priority pesticides: Better by GC-MS or LC-MS/MS? [J].
Alder, Lutz ;
Greulich, Kerstin ;
Kempe, Guenther ;
Vieth, Barbel .
MASS SPECTROMETRY REVIEWS, 2006, 25 (06) :838-865
[3]  
Anastassiades M, 2003, J AOAC INT, V86, P412
[4]   Ion suppression in mass spectrometry [J].
Annesley, TM .
CLINICAL CHEMISTRY, 2003, 49 (07) :1041-1044
[5]  
Belitz H.-D., 2009, Food Chemistry, V4th, P374
[6]   Evaluation of matrix effects in metabolite profiling based on capillary liquid chromatography electrospray ionization quadrupole time-of-flight mass spectrometry [J].
Boettcher, Christoph ;
v. Roepenack-Lahaye, Edda ;
Willscher, Edith ;
Scheel, Dierk ;
Clemens, Stephan .
ANALYTICAL CHEMISTRY, 2007, 79 (04) :1507-1513
[7]  
Bonfiglio R, 1999, RAPID COMMUN MASS SP, V13, P1175, DOI 10.1002/(SICI)1097-0231(19990630)13:12<1175::AID-RCM639>3.0.CO
[8]  
2-0
[9]   A rapid and reliable method for the quantitation of tricyclic antidepressants in serum using HPLC-MS/MS [J].
Breaud, Autumn R. ;
Harlan, Robert ;
Kozak, Marta ;
Clarke, William .
CLINICAL BIOCHEMISTRY, 2009, 42 (12) :1300-1307
[10]   Analysis of steroids in yeast-mediated cell culture by on-line solid-phase extraction coupled high-performance liquid chromatography electrospray-ionization/mass spectrometry and novel continuous postcolumn infusion of internal standard technique [J].
Cheng, Cheanyeh ;
Tsai, Hsiang-Rong .
ANALYTICA CHIMICA ACTA, 2008, 623 (02) :168-177