Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns

被引:45
作者
Freitas, Amilcar C., Jr. [2 ,3 ]
Bonfante, Estevam A. [1 ]
Rocha, Eduardo P. [3 ]
Silva, Nelson R. F. A. [4 ]
Marotta, Leonard [2 ]
Coelho, Paulo G. [2 ]
机构
[1] UNIGRANRIO Univ, Postgrad Program Dent, Sch Hlth Sci, BR-25071202 Rio De Janeiro, Brazil
[2] NYU, Coll Dent, Dept Biomat & Biomimet, New York, NY USA
[3] Sao Paulo State Univ, Dept Dent Mat & Prosthodont, Aracatuba Sch Dent, Sao Paulo, Brazil
[4] NYU, Coll Dent, Dept Prosthodont, New York, NY USA
关键词
cement-retained; fractography; implant-supported prostheses; reliability; screw-retained; Y-TZP; ABUTMENT CONNECTIONS; COMPLICATION RATES; SINGLE CROWNS; BITE FORCE; IN-VITRO; FATIGUE; RESISTANCE; SURVIVAL; SYSTEM;
D O I
10.1111/j.1600-0722.2011.00837.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The mechanical performance of cemented or screw-retained implant-supported crowns with an internal or external configuration is yet to be understood. This in vitro study evaluated the effect of screw-retained and cement-retained prostheses on internal and external implant-abutment connections. Thereby, the reliability and failure modes of crowns were investigated. Eighty-four implants (Emfils; Colosso Evolution system) were divided into four groups (n = 21 each): screw-retained and internal connection (Si), screw-retained and external connection (Se), cement-retained and internal connection (Ci), and cement-retained and external connection (Ce). Ti-6Al-4V abutments were torqued (30 Ncm) to the implants, and maxillary central incisor metal crowns were torqued (30 Ncm) or cemented (Rely X Unicem; 3M-ESPE) and subjected to accelerated life-testing in water. Use-level probability Weibull curves and reliability for 50,000 cycles at 150 N were calculated. The beta values for Si (1.72), Se (1.50), Ci (1.34), and Ce (1.77) groups indicated that fatigue/damage accumulation accelerated their failure. The Ci group presented the highest reliability, the Se group presented the lowest reliability, and Si and Ce groups presented intermediate reliability. Screw-retained restorations presented mainly abutment fracture. Cement-retained restorations resulted in failures of the screw in the Ce group, but implant/screw fracture in the Ci group.
引用
收藏
页码:323 / 330
页数:8
相关论文
共 40 条
[1]  
ALBREKTSSON T, 1986, J ORAL MAXILLOFAC IM, V1, P11
[2]  
[Anonymous], WEIB DISTR BET
[3]  
[Anonymous], 2003, PRACT FAIL ANAL, DOI DOI 10.1007/BF02715463
[4]  
*ASM INT, 1992, HDB FAIL AN PREV, V11
[5]  
Binon PP, 2000, INT J ORAL MAX IMPL, V15, P76
[6]   Reliability and Failure Modes of Implant-Supported Y-TZP and MCR Three-Unit Bridges [J].
Bonfante, Estevam Augusto ;
Coelho, Paulo Guilherme ;
Navarro, Jose Manuel, Jr. ;
Pegoraro, Luiz Fernando ;
Bonfante, Gerson ;
Thompson, Van Purdy ;
Silva, Nelson Renato Franc Alves .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (03) :235-243
[7]   Fatigue testing of two porcelain-zirconia all-ceramic crown systems [J].
Coelho, P. G. ;
Silva, N. R. ;
Bonfante, E. A. ;
Guess, P. C. ;
Rekow, E. D. ;
Thompson, V. P. .
DENTAL MATERIALS, 2009, 25 (09) :1122-1127
[8]   Laboratory Simulation of Y-TZP All-ceramic Crown Clinical Failures [J].
Coelho, P. G. ;
Bonfante, E. A. ;
Silva, N. R. F. ;
Rekow, E. D. ;
Thompson, V. P. .
JOURNAL OF DENTAL RESEARCH, 2009, 88 (04) :382-386
[9]   Influence of the Connector and Implant Design on the Implant-Tooth-Connected Prostheses [J].
da Silva, Edmar Ferreira ;
Pellizzer, Eduardo Piza ;
Quinelli Mazaro, Jose Vitor ;
Garcia, Idelmo Rangel, Jr. .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (03) :254-262
[10]  
Finger IM, 2003, PRACT PROCED AESTHET, V15, P634