Optimization of a hybrid magnetic bearing for a magnetically levitated blood pump via 3-D FEA

被引:36
作者
Cheng, Shanbao [1 ]
Olles, Mark W. [1 ]
Burger, Aaron F. [1 ]
Day, Steven W. [1 ]
机构
[1] Rochester Inst Technol, Rochester, NY 14623 USA
关键词
Hybrid magnetic bearing; 3-D finite element analysis; Axial flow blood pump; Magnetic bearing test rig; DESIGN; SYSTEM; ROTOR;
D O I
10.1016/j.mechatronics.2011.07.010
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In order to improve the performance of a magnetically levitated (maglev) axial flow blood pump, three-dimensional (3-D) finite element analysis (FEA) was used to optimize the design of a hybrid magnetic bearing (HMB). Radial, axial, and current stiffness of multiple design variations of the HMB were calculated using a 3-D FEA package and verified by experimental results. As compared with the original design, the optimized HMB had twice the axial stiffness with the resulting increase of negative radial stiffness partially compensated for by increased current stiffness. Accordingly, the performance of the maglev axial flow blood pump with the optimized HMBs was improved: the maximum pump speed was increased from 6000 rpm to 9000 rpm (50%). The radial, axial and current stiffness of the HMB was found to be linear at nominal operational position from both 3-D FEA and empirical measurements. Stiffness values determined by FEA and empirical measurements agreed well with one another. The magnetic flux density distribution and flux loop of the HMB were also visualized via 3-D FEA which confirms the designers' initial assumption about the function of this HMB. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1163 / 1169
页数:7
相关论文
共 21 条
[1]   Active magnetic bearings for the identification and fault diagnosis in turbomachinery [J].
Aenis, M ;
Knopf, E ;
Nordmann, R .
MECHATRONICS, 2002, 12 (08) :1011-1021
[2]  
Allaire P, 1998, ARTIF ORGANS, V22, P475
[3]  
[Anonymous], 2005, MAGNETIC BEARINGS BE
[4]  
Cheng S., 2010, P 2010 IEEE ASME INT, P187
[5]   Miniaturization of a Magnetically Levitated Axial Flow Blood Pump [J].
Cheng, Shanbao ;
Olles, Mark W. ;
Olsen, Don B. ;
Joyce, Lyle D. ;
Day, Steven W. .
ARTIFICIAL ORGANS, 2010, 34 (10) :807-815
[6]  
COMSOL, AC DC MOD
[7]  
Gomez A. David, 2009, THESIS ROCHESTER I T
[8]   Basic Characteristics of an Active Thrust Magnetic Bearing With a Cylindrical Rotor Core [J].
Hijikata, Kimio ;
Kobayashi, Shunsuke ;
Takemoto, Masatsugu ;
Tanaka, Yasuhiro ;
Chiba, Akira ;
Fukao, Tadashi .
IEEE TRANSACTIONS ON MAGNETICS, 2008, 44 (11) :4167-4170
[9]   Third-generation blood pumps with mechanical noncontact magnetic bearings [J].
Hoshi, H ;
Shinshi, T ;
Takatani, S .
ARTIFICIAL ORGANS, 2006, 30 (05) :324-338
[10]   A New Structure for Permanent-Magnet-Biased Axial Hybrid Magnetic Bearings [J].
Jiancheng, Fang ;
Jinji, Sun ;
Yanliang, Xu ;
Xi, Wang .
IEEE TRANSACTIONS ON MAGNETICS, 2009, 45 (12) :5319-5325