Monte Carlo calculations and measurements of absorbed dose per monitor unit for the treatment of uveal melanoma with proton therapy

被引:38
作者
Koch, Nicholas [1 ,2 ]
Newhauser, Wayne D. [1 ,2 ]
Titt, Uwe [1 ,2 ]
Gombos, Dan [3 ]
Coombes, Kevin [2 ,4 ]
Starkschall, George [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas Hlth Sci Ctr, Grad Sch Biomed Sci, Houston, TX 77030 USA
[3] MDACC, Dept Head & Neck Surg, Sect Ophthalmol, Unit 441, Houston, TX USA
[4] Univ Texas MD Anderson Canc Ctr, Dept Bioinformat & Computat Biol, Houston, TX 77030 USA
关键词
D O I
10.1088/0031-9155/53/6/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The treatment of uveal melanoma with proton radiotherapy has provided excellent clinical outcomes. However, contemporary treatment planning systems use simplistic dose algorithms that limit the accuracy of relative dose distributions. Further, absolute predictions of absorbed dose per monitor unit are not yet available in these systems. The purpose of this study was to determine if Monte Carlo methods could predict dose per monitor unit ( D/MU) value at the center of a proton spread-out Bragg peak ( SOBP) to within 1% on measured values for a variety of treatment fields relevant to ocular proton therapy. The MCNPX Monte Carlo transport code, in combination with realistic models for the ocular beam delivery apparatus and a water phantom, was used to calculate dose distributions and D/MU values, which were verified by the measurements. Measured proton beam data included central-axis depth dose profiles, relative cross-field profiles and absolute D/MU measurements under several combinations of beam penetration ranges and range-modulation widths. The Monte Carlo method predicted D/MU values that agreed with measurement to within 1% and dose profiles that agreed with measurement to within 3% of peak dose or within 0.5 mm distance-to-agreement. Lastly, a demonstration of the clinical utility of this technique included calculations of dose distributions and D/MU values in a realistic model of the human eye. It is possible to predict D/MU values accurately for clinical relevant range-modulated proton beams for ocular therapy using the Monte Carlo method. It is thus feasible to use the Monte Carlo method as a routine absolute dose algorithm for ocular proton therapy.
引用
收藏
页码:1581 / 1594
页数:14
相关论文
共 33 条
[1]  
CASCIO EW, 2004, NPTC RAD EFF DAT WOR
[2]  
CONSTABLE IJ, 1974, INVEST OPHTH VISUAL, V13, P280
[3]   Precise modelling of the eye for proton therapy of intra-ocular tumours [J].
Dobler, B ;
Bendl, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (04) :593-613
[4]   Eye retention after proton beam radiotherapy for uveal melanoma [J].
Egger, E ;
Zografos, L ;
Schalenbourg, A ;
Beati, D ;
Böhringer, T ;
Chamot, L ;
Goitein, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (04) :867-880
[5]   PLANNING PROTON THERAPY OF THE EYE [J].
GOITEIN, M ;
MILLER, T .
MEDICAL PHYSICS, 1983, 10 (03) :275-283
[6]  
GRAGOUDAS ES, 1980, OPHTHALMOLOGY, V87, P571
[7]   PROTON IRRADIATION OF CHOROIDAL MELANOMAS - PRELIMINARY-RESULTS [J].
GRAGOUDAS, ES ;
GOITEIN, M ;
KOEHLER, A ;
CONSTABLE, IJ ;
WAGNER, MS ;
VERHEY, L ;
TEPPER, J ;
SUIT, HD ;
BROCKHURST, RJ ;
SCHNEIDER, RJ ;
JOHNSON, KN .
ARCHIVES OF OPHTHALMOLOGY, 1978, 96 (09) :1583-1591
[8]   PROTON-BEAM IRRADIATION OF UVEAL MELANOMAS - RESULTS OF 51/2-YEAR STUDY [J].
GRAGOUDAS, ES ;
GOITEIN, M ;
VERHEY, L ;
MUNZENREIDER, J ;
URIE, M ;
SUIT, H ;
KOEHLER, A .
ARCHIVES OF OPHTHALMOLOGY, 1982, 100 (06) :928-934
[9]   PROTON IRRADIATION OF SMALL CHOROIDAL MALIGNANT MELANOMAS [J].
GRAGOUDAS, ES ;
GOITEIN, M ;
KOEHLER, AM ;
VERHEY, L ;
TEPPER, J ;
SUIT, HD ;
BROCKHURST, R ;
CONSTABLE, IJ .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1977, 83 (05) :665-673
[10]  
Gragoudas Evangelos S, 2005, Ophthalmol Clin North Am, V18, P111, DOI 10.1016/j.ohc.2004.08.002