Flow Analysis of Ventricular Assist Device Inflow and Outflow Cannula Positioning Using a Naturally Shaped Ventricle and Aortic Branch

被引:53
作者
Laumen, Marco [1 ]
Kaufmann, Tim [1 ]
Timms, Daniel [1 ]
Schlanstein, Peter [1 ]
Jansen, Sebastian [1 ]
Gregory, Shaun [2 ]
Wong, Kai Chun [1 ]
Schmitz-Rode, Thomas [1 ]
Steinseifer, Ulrich [1 ]
机构
[1] Rhein Westfal TH Aachen, Helmholtz Inst, Dept Cardiovasc Engn, Inst Appl Med Engn, D-52074 Aachen, Germany
[2] Queensland Univ Technol, Sch Engn Syst & Med Device Domain, Inst Hlth & Biomed Innovat, Brisbane, Qld 4001, Australia
关键词
Computational fluid dynamics; Inflow cannula; Particle image velocimetry; Rotary blood pump; Ventricular assist device; CARDIOPULMONARY BYPASS; SUCTION DETECTION; PULSATILE; SUPPORT; EXPERIENCE; PERFUSION; IMPACT; ARTERY; PUMPS;
D O I
10.1111/j.1525-1594.2010.01098.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Tip geometry and placement of rotary blood pump inflow and outflow cannulae influence the dynamics of flow within the ventricle and aortic branch. Cannulation, therefore, directly influences the potential for thrombus formation and end-organ perfusion during ventricular assist device (VAD) support or cardiopulmonary bypass (CPB). The purpose of this study was to investigate the effect of various inflow/outflow cannula tip geometries and positions on ventricular and greater vessel flow patterns to evaluate ventricular washout and impact on cerebral perfusion. Transparent models of a dilated cardiomyopathic ventricle and an aortic branch were reconstructed from magnetic resonance imaging data to allow flow measurements using particle image velocimetry (PIV). The contractile function of the failing ventricle was reproduced pneumatically, and supported with a rotary pump. Flow patterns were visualized around VAD inflow cannulae, with various tip geometries placed in three positions in the ventricle. The outflow cannula was placed in the subclavian artery and at several positions in the aorta. Flow patterns were measured using PIV and used to validate an aortic flow computational fluid dynamic study. The PIV technique indicated that locating the inflow tip in the left ventricular outflow tract improved complete ventricular washout while the tip geometry had a smaller influence. However, side holes in the inflow cannula improved washout in all cases. The PIV results confirmed that the positioning and orientation of the outflow cannula in the aortic branch had a high impact on the flow pattern in the vessels, with a negative blood flow in the right carotid artery observed in some cases. Cannula placement within the ventricle had a high influence on chamber washout. The positioning of the outflow cannula directly influences the flow through the greater vessels, and may be responsible for the occasional reduction in cerebral perfusion seen in clinical CPB.
引用
收藏
页码:798 / 806
页数:9
相关论文
共 24 条
[1]  
ADRIAN RJ, 1991, ANNU REV FLUID MECH, V23, P261, DOI 10.1146/annurev.fluid.23.1.261
[2]   Intraaortic Balloon Pump Counterpulsation and Cerebral Autoregulation: an observational study [J].
Bellapart, Judith ;
Geng, Shureng ;
Dunster, Kimble ;
Timms, Daniel ;
Barnett, Adrian G. ;
Boots, Rob ;
Fraser, John F. .
BMC ANESTHESIOLOGY, 2010, 10
[3]   Initial clinical experience with the Jarvik 2000 implantable axial-flow left ventricular assist system [J].
Frazier, OH ;
Myers, TJ ;
Gregoric, ID ;
Khan, T ;
Delgado, R ;
Croitoru, M ;
Miller, K ;
Jarvik, R ;
Westaby, S .
CIRCULATION, 2002, 105 (24) :2855-2860
[4]   The effects of pulsatile and nonpulsatile systemic perfusion on renal sympathetic nerve activity in anesthetized dogs [J].
Fukae, K ;
Tominaga, R ;
Tokunaga, S ;
Kawachi, Y ;
Imaizumi, T ;
Yasui, H .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 111 (02) :478-484
[5]   Impella Recover 100 microaxial left ventricular assist device: The Niguarda experience [J].
Garatti, A ;
Colombo, T ;
Russo, C ;
Lanfranconi, M ;
Bruschi, G ;
Milazzo, F ;
Catena, E ;
Vitali, E .
TRANSPLANTATION PROCEEDINGS, 2004, 36 (03) :623-626
[6]   Hydrodynamics of aortic arch vessels during perfusion through the right subclavian artery [J].
Gerdes, A ;
Joubert-Hübner, E ;
Esders, K ;
Sievers, HH .
ANNALS OF THORACIC SURGERY, 2000, 69 (05) :1425-1430
[7]   Changes in blood viscosity by heparin and argatroban [J].
Hitosugi, M ;
Niwa, M ;
Takatsu, A .
THROMBOSIS RESEARCH, 2001, 104 (05) :371-374
[8]   The impact of aortic manipulation on neurologic outcomes after coronary artery bypass surgery: A risk-adjusted study [J].
Kapetanakis, EI ;
Stamou, SC ;
Dullum, MKC ;
Hill, PC ;
Haile, E ;
Boyce, SW ;
Bafi, AS ;
Petro, KR ;
Corso, PJ .
ANNALS OF THORACIC SURGERY, 2004, 78 (05) :1564-1571
[9]   Flow Distribution During Cardiopulmonary Bypass in Dependency on the Outflow Cannula Positioning [J].
Kaufmann, Tim A. S. ;
Hormes, Marcus ;
Laumen, Marco ;
Timms, Daniel L. ;
Schmitz-Rode, Thomas ;
Moritz, Anton ;
Dzemali, Omer ;
Steinseifer, Ulrich .
ARTIFICIAL ORGANS, 2009, 33 (11) :988-992
[10]   The Impact of Aortic/Subclavian Outflow Cannulation for Cardiopulmonary Bypass and Cardiac Support: A Computational Fluid Dynamics Study [J].
Kaufmann, Tim A. S. ;
Hormes, Marcus ;
Laumen, Marco ;
Timms, Daniel L. ;
Linde, Torsten ;
Schmitz-Rode, Thomas ;
Moritz, Anton ;
Dzemali, Omer ;
Steinseifer, Ulrich .
ARTIFICIAL ORGANS, 2009, 33 (09) :727-732