Hammering sound frequency analysis and prevention of intraoperative periprosthetic fractures during total hip arthroplasty

被引:28
作者
Sakai, Rina [1 ]
Kikuchi, Aki
Morita, Towa
Takahira, Naonobu [2 ]
Uchiyama, Katsufumi [2 ]
Yamamoto, Takeaki [2 ]
Moriya, Mistutoshi [2 ]
Uchida, Kentaro [2 ]
Fukushima, Kensuke [2 ]
Tanaka, Kensei [3 ]
Takaso, Masashi [2 ]
Itoman, Moritoshi [4 ]
Mabuchi, Kiyoshi
机构
[1] Kitasato Univ, Sch Allied Hlth Sci, Dept Med Engn & Technol, Minami Ku, Sagamihara, Kanagawa 2520329, Japan
[2] Kitasato Univ, Sch Med, Dept Orthopaed Surg, Sagamihara, Kanagawa 2520329, Japan
[3] Kitasato Univ, Grad Sch Med Sci, Sagamihara, Kanagawa 2520329, Japan
[4] Kyushu Rosai Hosp, Kitakyushu, Fukuoka, Japan
关键词
Periprosthetic fracture; Total hip arthroplasty; Hammering sound; Frequency analysis; Yield stress; FEMORAL FRACTURES; RISK-FACTORS; PROSTHESIS; STABILITY; STEMS; FAILURE; FEMUR;
D O I
10.5301/HIP.2011.8823
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Adequate fixation at the time of cementless stem implantation depends on the operator's experience. An objective evaluation method to determine whether the stem has been appropriately implanted may be helpful. We studied the relationship between the hammering sound frequency during stem implantation and internal stress in a femoral model, and evaluated the possible usefulness of hammering sound frequency analysis for preventing intraoperative fracture. Three types of cementless stem (BiCONTACT (R), SL-PLUS (R), and AI-Hip (R)) were used. Surgeons performed stem insertion using a procedure similar to that employed in a routine operation. Stress was estimated by finite element analysis, the hammering force was measured, and frequency analysis of hammering sound data obtained using a microphone. Finite element analysis showed a decrease in the hammering sound frequency with an increase in the estimated maximum stress. When a decrease in frequency was observed, adequate hammering had occurred, and the continuation of hammering risked fracture. Based on the relationship between stress and frequency, the evaluation of changes in frequency may be useful for preventing the development of intraoperative fractures. Using our method, when a decrease in frequency is observed, the hammering force should be reduced. Hammering sound frequency analysis may allow the prediction of bone fractures that can be visually confirmed, and may be a useful objective evaluation method for the prevention of intraoperative periprosthetic fractures during stem insertion.
引用
收藏
页码:718 / 723
页数:6
相关论文
共 16 条
[1]  
Evans F G, 1969, Artif Limbs, V13, P37
[2]   Periprosthetic femoral fractures - a comparison between cemented and uncemented hemiarthroplasties [J].
Foster, AP ;
Thompson, NW ;
Wong, J ;
Charlwood, AP .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2005, 36 (03) :424-429
[3]   Primary stability in cementless femoral stems:: custom-made versus conventional femoral prosthesis [J].
Götze, C ;
Steens, W ;
Vieth, V ;
Poremba, C ;
Claes, L ;
Steinbeck, J .
CLINICAL BIOMECHANICS, 2002, 17 (04) :267-273
[4]  
GRUEN TA, 1979, CLIN ORTHOP RELAT R, P17
[5]   Intra-operative evaluation of cementless hip implant stability: A prototype device based on vibration analysis [J].
Lannocca, Maurizio ;
Varini, Elena ;
Cappello, Angelo ;
Cristofolini, Luca ;
Bialoblocka, Ewa .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (08) :886-894
[6]   Risk factors for failure after treatment of a periprosthetic fracture of the femur [J].
Lindahl, H ;
Malchau, H ;
Odén, A ;
Garellick, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (01) :26-30
[7]   Epidemiology of periprosthetic femur fracture around a total hip arthroplasty [J].
Lindahl, Hans .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (06) :651-654
[8]   Periprosthetic fractures evaluation and treatment [J].
Masri, BA ;
Meek, RMD ;
Duncan, CP .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (420) :80-95
[9]  
Moroni A, 2000, ANN CHIR GYNAECOL FE, V89, P113
[10]   In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prostheses [J].
Pastrav L.C. ;
Jaecques S.V. ;
Jonkers I. ;
Perre G.V.D. ;
Mulier M. .
Journal of Orthopaedic Surgery and Research, 4 (1)