Quality of treatment plans and accuracy of in vivo portal dosimetry in hybrid intensity-modulated radiation therapy and volumetric modulated arc therapy for prostate cancer

被引:9
作者
Bedford, James L.
Smyth, Gregory
Hanson, Ian M.
Tree, Alison C.
Dearnaley, David P.
Hansen, Vibeke N.
机构
[1] Inst Canc Res, London, England
[2] Royal Marsden NHS Fdn Trust, London, England
关键词
IMRT; VMAT; Prostate; Treatment planning; Portal dosimetry; CONFORMAL RADIOTHERAPY; INTRAFRACTION MOTION; HELICAL TOMOTHERAPY; EPID DOSIMETRY; VERIFICATION; IMRT; VMAT; TIME;
D O I
10.1016/j.radonc.2016.07.004
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Delivering selected parts of volumetric modulated arc therapy (VMAT) plans using step-and-shoot intensity modulated radiotherapy (IMRT) beams has the potential to increase plan quality by allowing specific aperture positioning. This study investigates the quality of treatment plans and the accuracy of in vivo portal dosimetry in such a hybrid approach for the case of prostate radiotherapy. Material and methods: Conformal and limited-modulation VMAT plans were produced, together with five hybrid IMRT/VMAT plans, in which 0%, 25%, 50%, 75% or 100% of the segments were sequenced for IMRT, while the remainder were sequenced for VMAT. Integrated portal images were predicted for the plans. The plans were then delivered as a single hybrid beam using an Elekta Synergy accelerator with Agility head to a water-equivalent phantom and treatment time, isocentric dose and portal images were measured. Results: Increasing the IMRT percentage improves dose uniformity to the planning target volume (p < 0.01 for 50% IMRT or more), substantially reduces the volume of rectum irradiated to 65 Gy (p = 0.02 for 25% IMRT) and increases the monitor units (p < 0.001). Delivery time also increases substantially. All plans show accurate delivery of dose and reliable prediction of portal images. Conclusions: Hybrid IMRT/VMAT can be efficiently planned and delivered as a single beam sequence. Beyond 25% IMRT, the delivery time becomes unacceptably long, with increased risk of intrafraction motion, but 25% IMRT is an attractive compromise. Integrated portal images can be used to perform in vivo dosimetry for this technique. (C) 2016 The Royal Marsden NHS Foundation Trust. Published by Elsevier Ireland Ltd.
引用
收藏
页码:320 / 326
页数:7
相关论文
共 33 条
  • [1] Portal dosimetry for VMAT using integrated images obtained during treatment
    Bedford, James L.
    Hanson, Ian M.
    Hansen, Vibeke Nordmark
    [J]. MEDICAL PHYSICS, 2014, 41 (02)
  • [2] Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator
    Bedford, James L.
    Thomas, Michael D. R.
    Smyth, Gregory
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 172 - 185
  • [3] Sinogram analysis of aperture optimization by iterative least-squares in volumetric modulated arc therapy
    Bedford, James L.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05) : 1235 - 1250
  • [4] Treatment planning for volumetric modulated arc therapy
    Bedford, James L.
    [J]. MEDICAL PHYSICS, 2009, 36 (11) : 5128 - 5138
  • [5] Speed versus accuracy in a fast convolution photon dose calculation for conformal radiotherapy
    Bedford, JL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (19) : 3475 - 3484
  • [6] A comparison of coplanar four-field techniques for conformal radiotherapy of the prostate
    Bedford, JL
    Khoo, VS
    Oldham, M
    Dearnaley, DP
    Webb, S
    [J]. RADIOTHERAPY AND ONCOLOGY, 1999, 51 (03) : 225 - 235
  • [7] British Institute of Radiology, 2008, SAF RAD
  • [8] The superiority of hybrid-volumetric arc therapy (VMAT) technique over double arcs VMAT and 3D-conformal technique in the treatment of locally advanced non-small cell lung cancer - A planning study
    Chan, Oscar S. H.
    Lee, Michael C. H.
    Hung, Albert W. M.
    Chang, Amy T. Y.
    Yeung, Rebecca M. W.
    Lee, Anne W. M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 101 (02) : 298 - 302
  • [9] Model-based prediction of portal dose images during patient treatment
    Chytyk-Praznik, K.
    VanUytven, E.
    vanBeek, T. A.
    Greer, P. B.
    McCurdy, B. M. C.
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [10] Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: preliminary safety results from the CHHiP randomised controlled trial
    Dearnaley, David
    Syndikus, Isabel
    Sumo, Georges
    Bidmead, Margaret
    Bloomfield, David
    Clark, Catharine
    Gao, Annie
    Hassan, Shama
    Horwich, Alan
    Huddart, Robert
    Khoo, Vincent
    Kirkbride, Peter
    Mayles, Helen
    Mayles, Philip
    Naismith, Olivia
    Parker, Chris
    Patterson, Helen
    Russell, Martin
    Scrase, Christopher
    South, Chris
    Staffurth, John
    Hall, Emma