Porous Ti-6Al-4V alloy fabricated by spark plasma sintering for biornimetic surface modification

被引:31
作者
Kon, M [1 ]
Hirakata, LM [1 ]
Asaoka, K [1 ]
机构
[1] Univ Tokushima, Sch Dent, Dept Dent Engn, Tokushima 7708504, Japan
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS | 2004年 / 68B卷 / 01期
关键词
titanium and titanium alloy; sintering; surface modification; calcium-phosphate coating; biocompatibility;
D O I
10.1002/jbm.b.20004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Porous compacts with both biological and biomecbanical compatibilities and high strength were developed. Spherical powders of Ti-6Al-4V alloy, which were either as received or surface modified with the use of calcium ions by hydrothermal treatment (HTT), were fabricated by a spark plasma sintering process. The porous compacts of pure Ti were used as reference materials. Porosity was approximately 30%, and compressive strengths were 113 and 125 MPa for the as-received Ti alloy powders and those modified by the HTT process, respectively. The bending strength and elastic modulus of as-received Ti alloy powders were 128-178 MPa and 16-18 GPa, respectively. Each of the compacts was immersed in simulated body fluid (SBF). The amount of adsorption/precipitation of calcium phosphate through the compacts was measured by weight change and was observed by SEM. The compacts were covered with calcium phosphate after 2 weeks of immersion in SBF. The compacts of Ti alloy had plenty of precipitated apatite crystals, and modification by HTT accumulated more precipitation. Because calcium phosphate is a mineral component of bone, apatite, which is precipitated on the surface of the compacts, could adsorb proteins and/or drugs such as antibiotics. It is expected that a large amount of proteins and/or drugs could be impregnated when the porous compacts developed are used. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:88 / 93
页数:6
相关论文
共 39 条
[1]   MECHANICAL-PROPERTIES AND BIOMECHANICAL COMPATIBILITY OF POROUS TITANIUM FOR DENTAL IMPLANTS [J].
ASAOKA, K ;
KUWAYAMA, N ;
OKUNO, O ;
MIURA, I .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1985, 19 (06) :699-713
[2]  
ASAOKA K, 2002, T 28 ANN M SOC BIOM, P630
[3]  
ASAOKA K, 2001, T PRICM4, P245
[4]  
BENNISSAN B, 1997, BIOCERAMICS, V10, P175
[5]   INTERFACE MECHANICS OF POROUS TITANIUM IMPLANTS [J].
CLEMOW, AJT ;
WEINSTEIN, AM ;
KLAWITTER, JJ ;
KOENEMAN, J ;
ANDERSON, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (01) :73-82
[6]   PLASMA SPRAYED COATINGS OF HYDROXYLAPATITE [J].
DEGROOT, K ;
GEESINK, R ;
KLEIN, CPAT ;
SEREKIAN, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (12) :1375-1381
[7]   STRUCTURAL-ANALYSIS OF HYDROXYAPATITE COATINGS ON TITANIUM [J].
DUCHEYNE, P ;
VANRAEMDONCK, W ;
HEUGHEBAERT, JC ;
HEUGHEBAERT, M .
BIOMATERIALS, 1986, 7 (02) :97-103
[8]   A STUDY ON THE MECHANISM OF PROTEIN ADSORPTION TO TIO2 [J].
ELLINGSEN, JE .
BIOMATERIALS, 1991, 12 (06) :593-596
[9]   Characterization of surface oxide films on titanium and bioactivity [J].
Feng, B ;
Chen, JY ;
Qi, SK ;
He, L ;
Zhao, JZ ;
Zhang, XD .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2002, 13 (05) :457-464
[10]   Biomimetic hydroxyapatite coating on metal implants [J].
Habibovic, P ;
Barrère, F ;
van Blitterswijk, CA ;
de Groot, K ;
Layrolle, P .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 2002, 85 (03) :517-522