PFC knee replacement - Osteolytic failures from extreme polyethylene degradation

被引:18
作者
Casey, David [1 ]
Cottrell, Jocelyn [1 ]
DiCarlo, Edward [1 ]
Windsor, Russell [1 ]
Wright, Timothy [1 ]
机构
[1] Hosp Special Surg, New York, NY 10021 USA
关键词
D O I
10.1097/BLO.0b013e3181591c62
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Despite the long-term success of press-fit condylar (PFC) knee prostheses, premature failures caused by aggressive rapid osteolysis have been reported. To investigate why patients experience such failures, we reviewed 48 retrieved implants and surrounding tissues together with demographic and radiographic data. Polyethylene degradation was determined from density profiles taken through the retrieved inserts. We compared the histology of tissues around PFC implants with that from around failed implants of similar designs from patients matched to length of implantation, body mass index, and age. The pathologic response in PFC patients showed more widespread, dense, sheet-like cellular infiltrate, whereas in the matched patients, the infiltrate was generally scattered discontinuously. The dominant wear mode of the PFC inserts was severe delamination on the articular surfaces. Wear damage was worse with increased length of implantation and was correlated with oxidative degradation and osteolysis. Degradation and osteolysis were more severe with inserts stored longer and sterilized by gamma radiation in air. These results underscore that degradation and increased shelf life lead to osteolysis and loosening. However, they raise questions concerning the cellular reaction to the debris from PFC implants that could lead to a better general understanding of osteolysis.
引用
收藏
页码:157 / 163
页数:7
相关论文
共 21 条
[1]   Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions [J].
Affatato, S ;
Bersaglia, G ;
Rocchi, M ;
Taddei, P ;
Fagnano, C ;
Toni, A .
BIOMATERIALS, 2005, 26 (16) :3259-3267
[2]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[3]  
BOHL JR, 1999, CLIN ORTHOP RELAT R, V367, P28
[4]  
BOSTROM MP, 1994, CLIN ORTHOPAEDICS, V309, P20
[5]   The press-fit condylar total knee system - 8-to 10-year results with a posterior cruciate-retaining design [J].
Buehler, KO ;
Venn-Watson, E ;
D'Lima, DD ;
Colwell, CW .
JOURNAL OF ARTHROPLASTY, 2000, 15 (06) :698-701
[6]   Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty [J].
Collier, MB ;
Engh, CA ;
Engh, GA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (04) :763-769
[7]   Backside wear of polyethylene tibial inserts: Mechanism and magnitude of material loss [J].
Conditt, MA ;
Thompson, MT ;
Usrey, MM ;
Ismaily, SK ;
Noble, PC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (02) :326-331
[8]   Bearing surface design changes affect contact patterns in total knee arthroplasty [J].
Cottrell, Jocelyn M. ;
Townsend, Elizabeth ;
Lipman, Joseph ;
Sculco, Thomas P. ;
Wright, Timothy M. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) :127-131
[9]   Polyethylene damage on the nonarticular surface of modular total knee prostheses [J].
Cuckler, JM ;
Lemons, J ;
Tamarapalli, JR ;
Beck, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :248-253
[10]   Factors influencing wear and osteolysis in press-fit condylar modular total knee replacements [J].
Fehring, TK ;
Murphy, JA ;
Hayes, TD ;
Roberts, DW ;
Pomeroy, DL ;
Griffin, WL .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :40-50