Two dimensional scanning probe using off-axis magnetic force of single solenoid for 3D OCT imaging

被引:0
作者
Min, Eun Jung [1 ]
Shin, Jun Geun [1 ]
Kim, Yuri [2 ]
Lee, Byeong Ha [1 ,2 ]
机构
[1] GIST, Sch Informat & Commun, 261 Cheomdan Gwagiro, Kwangju 500712, South Korea
[2] GIST, Grad Program Med Syst Engn, Kwangju 500712, South Korea
来源
THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XVIII | 2011年 / 7904卷
关键词
fiber-based hand-held probe; ferromagnetic material; spectral domain OCT; real-time in vivo imaging; OPTICAL COHERENCE TOMOGRAPHY;
D O I
10.1117/12.874367
中图分类号
TH742 [显微镜];
学科分类号
摘要
We propose a single piece optical fiber-based two-dimensional scanning hand-held probe suitable for three-dimensional optical coherence tomography. The probe consists of only a single piece of optical fiber loaded with a bead of ferromagnetic material, which acts as a vibrating cantilever. The fiber cantilever is two dimensionally actuated with a single miniaturized solenoid. For effective beam focusing, a fiber lens is formed at the end of the fiber. The inductance and input current of the solenoid were 100 mu H and 216 mA, respectively. The iron-bead on the fiber is located at the off-axis of solenoid for two-dimensional scanning. Then, by modulating the input current to the solenoid, it was possible to mechanically oscillate the fiber cantilever in an elliptically spiral pattern. With the proposed probe, 2-dimensional scanning could be experimentally achieved in a rate of 4 s/vortex across a scanning area of approximately 30 mm(2), which could be controlled with the length of the fiber or/and the weight of the iron-bead. Three-dimensional tomographic image of a coin was successfully obtained with the spectral domain optical coherence tomography equipped with the proposed scanner. It is expected that the scheme of 2-dimentional scanning with a single actuator might be useful in various real-time imaging applications including OCT owing to the advantages of low cost, low power consumption, simple fabrication process and versatile design.
引用
收藏
页数:7
相关论文
共 14 条
  • [1] Comparison of three-dimensional optical coherence tomography and high resolution photography for art conservation studies
    Adler, Desmond C.
    Stenger, Jens
    Gorczynska, Iwona
    Lie, Henry
    Hensick, Teri
    Spronk, Ron
    Wolohojian, Stephan
    Khandekar, Narayan
    Jiang, James Y.
    Barry, Scott
    Cable, Alex E.
    Huber, Robert
    Fujimoto, James G.
    [J]. OPTICS EXPRESS, 2007, 15 (24) : 15972 - 15986
  • [2] Two-axis MEMS scanning catheter for ultrahigh resolution three-dimensional and en face imaging
    Aguirre, Aaron D.
    Herz, Paul R.
    Chen, Yu
    Fujimoto, James G.
    Piyawattanametha, Wibool
    Fan, Li
    Wu, Ming C.
    [J]. OPTICS EXPRESS, 2007, 15 (05): : 2445 - 2453
  • [3] Forward-imaging instruments for optical coherence tomography
    Boppart, SA
    Bouma, BE
    Pitris, C
    Tearney, GJ
    Fujimoto, JG
    Brezinski, ME
    [J]. OPTICS LETTERS, 1997, 22 (21) : 1618 - 1620
  • [4] Häusler G, 1998, J BIOMED OPT, V3, P21, DOI 10.1117/1.429899
  • [5] Harzie R. L., 2008, OPT EXPRESS, V16, P20588
  • [6] An optical scanner based on electromagnetically actuated optical fiber
    Hu, Hans P.
    Le, Kevin D.
    Chiao, J. -C.
    [J]. MICROMACHINING AND MICROFABRICATION PROCESS TECHNOLOGY XI, 2006, 6109
  • [7] OPTICAL COHERENCE TOMOGRAPHY
    HUANG, D
    SWANSON, EA
    LIN, CP
    SCHUMAN, JS
    STINSON, WG
    CHANG, W
    HEE, MR
    FLOTTE, T
    GREGORY, K
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. SCIENCE, 1991, 254 (5035) : 1178 - 1181
  • [8] Single-body lensed-fiber scanning probe actuated by magnetic force for optical imaging
    Min, Eun Jung
    Na, Jihoon
    Ryu, Seon Young
    Lee, Byeong Ha
    [J]. OPTICS LETTERS, 2009, 34 (12) : 1897 - 1899
  • [9] Electrostatic forward-viewing scanning probe for Doppler optical coherence tomography using a dissipative polymer catheter
    Munce, Nigel R.
    Mariampillai, Adrian
    Standish, Beau A.
    Pop, Mihaela
    Anderson, Kevan J.
    Liu, George Y.
    Luk, Tim
    Courtney, Brian K.
    Wright, Graham A.
    Vitkin, I. Alex
    Yang, Victor X. D.
    [J]. OPTICS LETTERS, 2008, 33 (07) : 657 - 659
  • [10] Reitz J.R., 1993, FDN ELECTROMAGNETIC, V4th