The measurement of free thyroxine by isotope dilution tandem mass spectrometry

被引:99
作者
Soldin, SJ
Soukhova, N
Janicic, N
Jonklaas, J
Soldin, OP
机构
[1] Georgetown Clin Res Ctr, Bioanalyt Core Lab, Washington, DC 20007 USA
[2] Georgetown Univ, Dept Med, Washington, DC USA
[3] Georgetown Univ, Dept Pharmacol, Washington, DC USA
[4] George Washington Univ, Sch Med, Childrens Natl Med Ctr, Dept Lab Med, Washington, DC USA
[5] George Washington Univ, Sch Med, Dept Pediat, Washington, DC USA
[6] George Washington Univ, Sch Med, Dept Pathol, Washington, DC USA
[7] Georgetown Univ, Lombardi Canc Ctr, Div Canc Genet & Epidemiol, Washington, DC USA
关键词
free thyroxine; thyroxine; thyroid hormones; isotope dilution tandem mass spectrometry; immunoassay;
D O I
10.1016/j.cccn.2005.02.010
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Most clinical chemistry laboratories measure free thyroxine (FT4) by an analogue (direct) method. Nevertheless, the validity of analogue FT4 immunoassays has been questioned and patient's results using this approach frequently do not fit in with the clinical presentation. Because of these problems we routinely send out all direct free T4's that are < 2.5th percentile and many that are > 97.5th percentile for measurement by equilibrium dialysis, the gold standard method. In approximately half of these cases, the FT4 by equilibrium dialysis was normal. We developed a rapid, reliable free T4 method employing isotope dilution tandem mass spectrometry and compared our results with both the analogue (direct) free T4 and equilibrium dialysis procedures. Methods: An API-4000 tandem mass spectrometer (Sciex, Toronto, Canada) equipped with TurbolonSpray and Agilent HPLC system was used employing isotope dilution with deuterium labeled internal standard (IS= L-thyroxine-dA Serum was filtered through the Centrifree YM-30 ultrafiltration device by centrifugation, IS added to the ultrafiltrate, and 400 mu L injected onto a C18 column. After washing, the switch valve is activated and a methanol gradient allows for elution of both T4 and the IS into the LC/MS/MS system. Quantitation was by MRM analysis in the negative mode. Equilibrium dialysis was performed by the Nichols method and analogue free T4 results were obtained on the Dade Dimension RxL. Results: The within-day and between-day CV's were < 7.1% at all concentrations tested. The results correlated well with equilibrium dialysis (Eq Dial=0.971 MS+0.041, n=68, Syx=1.381, r=0.954). A poor correlation was found with the analogue (direct) free T4 method (IA=0.326 MS +6.27, n = 154, Syx= 1.96, r=0.459). Conclusions: Samples can be processed in batches of 30. The free T4 tandem MS method has it rapid turn-around-time vs the equilibrium dialysis procedure, with a chromatographic run time of 8 min per sample. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:113 / 118
页数:6
相关论文
共 16 条
[1]   The war on heterophilic antibody interference [J].
Bjerner, J ;
Bormer, OP ;
Nustad, K .
CLINICAL CHEMISTRY, 2005, 51 (01) :9-11
[2]  
BOSCATO LM, 1986, CLIN CHEM, V32, P1491
[3]  
DEMERS EM, 2003, NATL ACAD CLIN BIOCH, V263, P1529
[4]  
EKINS R, 1987, CLIN CHEM, V33, P2137
[5]   Simultaneous determination of endogenous and 13C-labelled thyroid hormones in plasma by stable isotope dilution mass spectrometry [J].
Hantson, AL ;
De Meyer, M ;
Guérit, N .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 807 (02) :185-192
[6]   Reference methods for the measurement of free thyroid hormones in blood evaluation of potential reference methods for free thyroxine [J].
Holm, SS ;
Hansen, SH ;
Faber, J ;
Staun-Olsen, P .
CLINICAL BIOCHEMISTRY, 2004, 37 (02) :85-93
[7]  
Klee GG, 2000, ARCH PATHOL LAB MED, V124, P921
[8]  
Kricka LJ, 2000, CLIN CHEM, V46, P1037
[9]   Trimester-specific changes in maternal thyroid hormone, thyrotropin, and thyroglobulin concentrations during gestation: Trends and associations across trimesters in iodine sufficiency [J].
Soldin, OP ;
Tractenberg, RE ;
Hollowell, JG ;
Jonklaas, J ;
Janicic, N ;
Soldin, SJ .
THYROID, 2004, 14 (12) :1084-1090
[10]   Trimester-specific reference intervals for thyroxine and triiodothyronine in pregnancy in iodine-sufficient women using isotope dilution tandem mass spectrometry and immunoassays [J].
Soldin, OP ;
Hilakivi-Clarke, L ;
Weiderpass, E ;
Soldin, SJ .
CLINICA CHIMICA ACTA, 2004, 349 (1-2) :181-189