Influence of the number and position of stripped screws on plate-screw construct biomechanical properties

被引:1
作者
Garces, Gerardo L. [1 ]
Yanez, Alejandro [2 ]
Cuadrado, Alberto [2 ]
Martel, Oscar [2 ]
机构
[1] Univ Las Palmas Gran Canaria, Orthopaed Surg, Dept Med & Surg Sci, Las Palmas Gran Canaria, Spain
[2] Univ Las Palmas Gran Canaria, Dept Mech Engn, Las Palmas Gran Canaria, Spain
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2017年 / 48卷
关键词
screw stripping; stiffness; plate-screw construct; DYNAMIC COMPRESSION PLATE; ANKLE FRACTURE FIXATION; OSTEOPOROTIC BONE; LOCKING ELEMENTS; STRAIN; STRENGTH; MODEL; STABILITY; STIFFNESS; SYSTEM;
D O I
10.1016/S0020-1383(17)30795-7
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Screw stripping is a common situation in fracture fixation, particularly in osteopenic bone treatment. Surgeons' perception of screw stripping is relatively poor and the real number of loose screws in every plate-screw construct is unknown. The biomechanical and clinical implications of the different possible screw-stripping situations are also unidentified. In this study, construct stiffness in different scenarios of stripped screws is investigated. Method: A bone surrogate comminuted osteoporotic fracture was fixed with four screws in both sides of the fracture gap in 75 specimens. In four groups, one or two screws closest or distal to the gap were over-tightened and left in place in one part of the construct and the remaining screws were tightened with 0.3 N m torque (four groups). In the fifth group (control), all the screws were tightened with 0.3 N m torque. Construct stiffness was tested in terms of compression, bending, and torsion for 1000 cycles. Results: When one or two screws closest to the gap were stripped, stiffness only decreased by, respectively, 5.7% or 7.6% under compression and 4.7% or 6.7% under bending; however, stiffness in torsion was 15.1% or 32%, respectively, lower than the initial stiffness. When a screw distal to the gap was stripped, the stiffness decreased by 28% under bending and 10% under compression; no change was noted under torsion. When two screws distal to the gap were stripped, the stiffness decreased by 11% in compression, collapsed under bending, and decreased by 8% under torsion. Conclusions: Position and number of stripped screws affect the biomechanical properties of a construct in different ways, depending on the acting forces. (c) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:S54 / S59
页数:6
相关论文
共 30 条
  • [1] Use of a synthetic bone void filler to augment screws in osteopenic ankle fracture fixation
    Andreassen, GS
    Hoiness, PR
    Skraamm, I
    Granlund, O
    Engebretsen, L
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2004, 124 (03) : 161 - 165
  • [2] Experimental evaluation of the holding power/stiffness of the self-tapping bone screws in normal and osteoporotic bone material
    Battula, Suneel
    Schoenfeld, Andrew
    Vrabec, Gregory
    Njus, Glen O.
    [J]. CLINICAL BIOMECHANICS, 2006, 21 (05) : 533 - 537
  • [3] Biomechanical Considerations for Surgical Stabilization of Osteoporotic Fractures
    Bogunovic, Ljiljana
    Cherney, Steven M.
    Rothermich, Marcus A.
    Gardner, Michael J.
    [J]. ORTHOPEDIC CLINICS OF NORTH AMERICA, 2013, 44 (02) : 183 - +
  • [4] Effects of surgical errors on small fragment screw fixation
    Collinge, Cory
    Hartigan, Brian
    Lautenschlager, Eugene P.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (06) : 410 - 413
  • [5] Cordey J., 1980, CURRENT CONCEPTS INT, P235
  • [6] Suitability of DCPs with Screw Locking Elements to allow sufficient interfragmentary motion to promote secondary bone healing of osteoporotic fractures
    Cuadrado, A.
    Yanez, A.
    Carta, J. A.
    Garces, G.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2013, 35 (06) : 852 - 859
  • [7] Inadvertent Screw Stripping During Ankle Fracture Fixation in Elderly Bone
    Dinah, A. Feroz
    Mears, Simon C.
    Knight, Trevor A.
    Soin, Sandeep P.
    Campbell, John T.
    Belkoff, Stephen M.
    [J]. GERIATRIC ORTHOPAEDIC SURGERY & REHABILITATION, 2011, 2 (03) : 86 - 89
  • [8] A new method of fixation in osteoporotic bone - A preliminary report
    Drew, T
    Allcock, P
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2002, 33 (08): : 685 - 689
  • [9] Stripping torque as a predictor of successful internal fracture fixation
    Edwards, TR
    Tevelen, G
    English, H
    Crawford, R
    [J]. ANZ JOURNAL OF SURGERY, 2005, 75 (12) : 1096 - 1099
  • [10] Biomechanics of locked plates and screws
    Egol, KA
    Kubiak, EN
    Fulkerson, E
    Kummer, FJ
    Koval, KJ
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) : 488 - 493