Field size dependent mapping of medical linear accelerator radiation leakage

被引:15
作者
Bezin, Jeremi Vu [1 ,2 ,3 ]
Veres, Attila [4 ]
Lefkopoulos, Dimitri [3 ]
Chavaudra, Jean [3 ]
Deutsch, Eric [3 ,5 ]
de Vathaire, Florent [1 ,2 ,3 ]
Diallo, Ibrahima [1 ,2 ,3 ]
机构
[1] Univ Paris 11, F-91405 Orsay, France
[2] INSERM, CESP Ctr Res Epidemiol & Populat Hlth, U1018, Radiat Epidemiol Team, F-94807 Villejuif, France
[3] Gustave Roussy, F-94805 Villejuif, France
[4] Equal Estro Lab, F-94805 Villejuif, France
[5] INSERM, UMR Radiotherapie Mol 1030, F-94807 Villejuif, France
基金
欧盟第七框架计划;
关键词
graphics library; multi-source modelling; leakage; radiotherapy; MONTE-CARLO MODEL; CANCER-RISK; DOSE CALCULATIONS; RADIOTHERAPY; CHILDHOOD; THERAPY; SCATTER; SCANS;
D O I
10.1088/0031-9155/60/5/2103
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the suitability of a graphics library based model for the assessment of linear accelerator radiation leakage. Transmission through the shielding elements was evaluated using the build-up factor corrected exponential attenuation law and the contribution from the electron guide was estimated using the approximation of a linear isotropic radioactive source. Model parameters were estimated by a fitting series of thermoluminescent dosimeter leakage measurements, achieved up to 100 cm from the beam central axis along three directions. The distribution of leakage data at the patient plane reflected the architecture of the shielding elements. Thus, the maximum leakage dose was found under the collimator when only one jaw shielded the primary beam and was about 0.08% of the dose at isocentre. Overall, we observe that the main contributor to leakage dose according to our model was the electron beam guide. Concerning the discrepancies between the measurements used to calibrate the model and the calculations from the model, the average difference was about 7%. Finally, graphics library modelling is a readily and suitable way to estimate leakage dose distribution on a personal computer. Such data could be useful for dosimetric evaluations in late effect studies.
引用
收藏
页码:2103 / 2116
页数:14
相关论文
共 35 条
[1]  
American National Standards Institute and American Nuclear Society, 1991, GAMM RAY ATT COEFF B
[2]  
[Anonymous], INT J SCI TECHNOL
[3]  
[Anonymous], DOSIMETRIE CURIETHER
[4]  
[Anonymous], MED EL EQ PART 2 1 P
[5]   Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validation [J].
Bednarz, Bryan ;
Xu, X. George .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04) :N43-N57
[6]   A multi-plane source model for out-of-field head scatter dose calculations in external beam photon therapy [J].
Benadjaoud, Mohamed Amine ;
Bezin, Jeremi ;
Veres, Attila ;
Lefkopoulos, Dimitri ;
Chavaudra, Jean ;
Bridier, Andre ;
de Vathaire, Florent ;
Diallo, Ibrahima .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22) :7725-7739
[7]   Radiotherapy out-of-field dosimetry: Experimental and computational results for photons in a water tank [J].
Bordy, J. M. ;
Bessieres, I. ;
d'Agostino, E. ;
Domingo, C. ;
d'Errico, F. ;
di Fulvio, A. ;
Knezevic, Z. ;
Miljanic, S. ;
Olko, P. ;
Ostrowsky, A. ;
Poumarede, B. ;
Sorel, S. ;
Stolarczyk, L. ;
Vermesse, D. .
RADIATION MEASUREMENTS, 2013, 57 :29-34
[8]   Internal scatter, the unavoidable major component of the peripheral dose in photon-beam radiotherapy [J].
Chofor, Ndimofor ;
Harder, Dietrich ;
Willborn, Kay C. ;
Poppe, Bjoern .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (06) :1733-1743
[9]   Experimental study on photon-beam peripheral doses, their components and some possibilities for their reduction [J].
Chofor, Ndimofor ;
Harder, Dietrich ;
Ruehmann, Antje ;
Willborn, Kay C. ;
Wiezorek, Tilo ;
Poppe, Bjoern .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (14) :4011-4027
[10]   Point/Counterpoint: Low-dose radiation is beneficial, not harmful [J].
Doss, Mohan ;
Little, Mark P. ;
Orton, Colin G. .
MEDICAL PHYSICS, 2014, 41 (07)