An Implantable Sacral Nerve Root Recording and Stimulation System for Micturition Function Restoration

被引:3
作者
Wang, Yuan [1 ]
Zhang, Xu [1 ]
Liu, Ming [1 ]
Pei, Weihua [1 ]
Wang, Kaifeng [2 ]
Chen, Hongda [1 ]
机构
[1] Chinese Acad Sci, Inst Semicond, State Key Lab Integrated Optoelectron, Beijing 100083, Peoples R China
[2] Peking Univ, Peoples Hosp, Dept Spinal Surg, Beijing 100044, Peoples R China
来源
IEICE TRANSACTIONS ON INFORMATION AND SYSTEMS | 2014年 / E97D卷 / 10期
基金
中国国家自然科学基金;
关键词
neural prosthesis; functional electrical stimulation; closed-loop bladder control; acute animal experiment; spinal cord injury; BLADDER CONTRACTIONS; DIELECTRIC-PROPERTIES; BIOLOGICAL TISSUES; PRESSURE; ONSET; ENG;
D O I
10.1587/transinf.2013EDP7395
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper provides a prototype neural prosthesis system dedicated to restoring continence and micturition function for patients with lower urinary tract diseases, such as detrusor hyperreflexia and detrusor-sphincter dyssynergia. This system consists of an ultra low-noise electroneurogram (ENG) signal recording module, a bi-phasic electrical stimulator module and a control unit for closed-loop bladder monitoring and controlling. In order to record extremely weak ENG signal from extradural sacral nerve roots, the system provides a programmable gain from 80 dB to 117 dB. By combining of advantages of commercial-off-the-shelf (COTS) electronics and custom designed IC, the recording front-end acquires a fairly low input-referred noise (IRN) of 0.69 mu V-rms under 300 Hz to 3 kHz and high area-efficiency. An on-chip multi-steps single slope analog-to-digital converter (ADC) is used to digitize the ENG signals at sampling rate of 10 kSPS and achieves an effective number of bits (ENOB) of 12.5. A bi-phasic current stimulus generator with wide voltage supply range (+/-0.9 V to +/-12.5 V) and variable output current amplitude (0-500 mu A) is introduced to overcome patient-depended impedance between electrode and tissue electrolyte. The total power consumption of the entire system is 5.61 mW. Recording and stimulation function of this system is switched by control unit with time division multiplexing strategy. The functionality of this proposed prototype system has been successfully verified through in-vivo experiments from dogs extradural sacral nerve roots.
引用
收藏
页码:2790 / 2801
页数:12
相关论文
共 38 条
  • [1] A combined wireless neural stimulating and recording system for study of pain processing
    Ativanichayaphong, Thermpon
    He, Ji Wei
    Hagains, Christopher E.
    Peng, Yuan B.
    Chiao, J. -C.
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 2008, 170 (01) : 25 - 34
  • [2] Implantable selective stimulator to improve bladder voiding: Design and chronic experiments in dogs
    Boyer, S
    Sawan, M
    Abdel-Gawad, M
    Robin, S
    Elhilali, MM
    [J]. IEEE TRANSACTIONS ON REHABILITATION ENGINEERING, 2000, 8 (04): : 464 - 470
  • [3] Variable patterned pudendal nerve stimuli improves reflex bladder activation
    Bruns, Tim M.
    Bhadra, Narendra
    Gustafson, Kenneth J.
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2008, 16 (02) : 140 - 148
  • [4] Bruns TM, 2011, IEEE ENG MED BIO, P4239, DOI 10.1109/IEMBS.2011.6091052
  • [5] The management of neurogenic bladder and sexual dysfunction after spinal cord injury
    Burns, AS
    Rivas, DA
    Ditunno, JF
    [J]. SPINE, 2001, 26 (24) : S129 - S136
  • [6] Principles of sacral nerve stimulation (SNS) for the treatment of bladder and urethral sphincter dysfunctions
    Chancellor, MB
    Chartier-Kastler, EJ
    [J]. NEUROMODULATION, 2000, 3 (01): : 15 - 26
  • [7] MEASUREMENT OF URINARY-BLADDER EMPTYING USING ELECTRICAL IMPEDANCE
    DENNISTON, JC
    BAKER, LE
    [J]. MEDICAL & BIOLOGICAL ENGINEERING, 1975, 13 (02): : 305 - 306
  • [8] Switch-Matrix-Based High-Density Microelectrode Array in CMOS Technology
    Frey, Urs
    Sedivy, Jan
    Heer, Flavio
    Pedron, Rene
    Ballini, Marco
    Mueller, Jan
    Bakkum, Douglas
    Hafizovic, Sadik
    Faraci, Francesca D.
    Greve, Frauke
    Kirstein, Kay-Uwe
    Hierlemann, Andreas
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2010, 45 (02) : 467 - 482
  • [9] The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz
    Gabriel, S
    Lau, RW
    Gabriel, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) : 2251 - 2269
  • [10] The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues
    Gabriel, S
    Lau, RW
    Gabriel, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) : 2271 - 2293