Evaluation of Deformable Image Registration Between High Dose Rate Brachytherapy and Intensity Modulated Radiation Therapy for Prostate Cancer

被引:0
作者
Yokoyama, Noriomi [1 ]
Takemura, Akihiro [2 ]
Kojima, Hironori [3 ]
Tsukamoto, Kousuke [1 ]
Ueda, Shinichi [3 ]
Noto, Kimiya [3 ]
机构
[1] Kanazawa Univ, Grad Sch Med Sci, Div Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ, Fac Hlth Sci, Inst Med Pharmaceut & Hlth Sci, Kanazawa, Ishikawa, Japan
[3] Kanazawa Univ Hosp, Dept Radiol Technol, Kanazawa, Ishikawa, Japan
来源
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3 | 2019年 / 68卷 / 03期
关键词
Radiotherapy; Brachytherapy; Intensity-modulated radiation therapy; Deformable image registration;
D O I
10.1007/978-981-10-9023-3_91
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
High risk prostate cancer is treated with a combination of intensity-modulated radiation therapy (IMRT) and high dose rate brachytherapy (HDR-BT). Deformable image registration (DIR) techniques used for dose accumulation sums dose distributions. The accuracy of DIR would get worse when density of an organ in a pair of registering two images differs greatly each other. Needles and contrast medium are used in HDR-BT. In this study, the effect of needles and contrast medium for DIR accuracy was evaluated. Six patients with prostate cancer were enrolled, who were treated with the combination of HDR-BT and IMRT. In the HDR-BT plan, needleless image (NI) and needleless and no-contrast medium image (NCI) were created from the original HDR-BT plan image (OI) to investigate the influence of needles and contrast medium. Both DIR and rigid registration (RR) were performed on the OI, NIs and NCIs by using MIM Maestro ver. 6.7.6 (MIM software Inc, Cleveland, USA) and after that the dose distribution of HDR-BT (used as the reference image) and IMRT were accumulated. The Dice Similarity coefficient (DSC) between DIR and RR were analyzed and compared each other. The mean DSC values of the prostate with DIR on OI, NI and NCI were 0.51, 0.57 and 0.73, respectively. The DSC with DIR on NCI was higher than DSC with DIR on OI and NI. The DSC values improved by removing the contrast medium.
引用
收藏
页码:505 / 508
页数:4
相关论文
共 6 条
[1]   RESULTS OF A MULTI-INSTITUTION DEFORMABLE REGISTRATION ACCURACY STUDY (MIDRAS) [J].
Brock, Kristy K. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02) :583-596
[2]   MEASURES OF THE AMOUNT OF ECOLOGIC ASSOCIATION BETWEEN SPECIES [J].
DICE, LR .
ECOLOGY, 1945, 26 (03) :297-302
[3]  
Ikeda T., 2016, AAPM 58 ANN M EXH
[4]  
Kadoya N., 2017, J RADIAT RES, P1
[5]  
Tanner C, 2000, LECT NOTES COMPUT SC, V1935, P327
[6]  
van Heerden Laura E., 2017, RADIOTHERAPY ONCOLOG