Accuracy of Dual-Energy Virtual Monochromatic CT Numbers: Comparison between the Single-Source Projection-Based and Dual-Source Image-Based Methods

被引:14
|
作者
Ueguchi, Takashi [1 ,2 ]
Ogihara, Ryota [3 ]
Yamada, Sachiko [3 ]
机构
[1] Natl Inst Informat & Commun Technol, Ctr Informat & Neural Networks, 1-4 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Osaka Univ, Grad Sch Frontier Biosci, 1-4 Yamadaoka, Suita, Osaka 5650871, Japan
[3] Osaka Univ Hosp, Dept Radiol, Suita, Osaka, Japan
基金
日本学术振兴会;
关键词
Dual-energy CT; virtual monochromatic CT; CT number; phantom study; beam hardening; ATOMIC NUMBERS; DISCRIMINATION; ANGIOGRAPHY; REDUCTION;
D O I
10.1016/j.acra.2018.02.022
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To investigate the accuracy of dual-energy virtual monochromatic computed tomography (CT) numbers obtained by two typical hardware and software implementations: the single-source projection-based method and the dual-source image based method. Materials and Methods: A phantom with different tissue equivalent inserts was scanned with both single-source and dual-source scanners. A fast kVp-switching feature was used on the single-source scanner, whereas a tin filter was used on the dual-source scanner. Virtual monochromatic CT images of the phantom at energy levels of 60, 100, and 140 keV were obtained by both projection-based (on the single source scanner) and image-based (on the dual-source scanner) methods. The accuracy of virtual monochromatic CT numbers for all inserts was assessed by comparing measured values to their corresponding true values. Linear regression analysis was performed to evaluate the dependency of measured CT numbers on tissue attenuation, method, and their interaction. Results: Root mean square values of systematic error over all inserts at 60, 100, and 140 keV were approximately 53, 21, and 29 Houns-field unit (HU) with the single-source projection-based method, and 46, 7, and 6 HU with the dual-source image-based method, respectively. Linear regression analysis revealed that the interaction between the attenuation and the method had a statistically significant effect on the measured CT numbers at 100 and 140 keV. Conclusions: There were attenuation-, method-, and energy level-dependent systematic errors in the measured virtual monochromatic CT numbers. CT number reproducibility was comparable between the two scanners, and CT numbers had better accuracy with the dual source image-based method at 100 and 140 keV.
引用
收藏
页码:1632 / 1639
页数:8
相关论文
共 50 条
  • [1] Virtual monochromatic imaging in dual-source dual-energy CT: Radiation dose and image quality
    Yu, Lifeng
    Christner, Jodie A.
    Leng, Shuai
    Wang, Jia
    Fletcher, Joel G.
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2011, 38 (12) : 6371 - 6379
  • [2] Virtual Monochromatic Imaging in Dual-source and Dual-energy CT for Visualization of Acute Ischemic Stroke
    Hara, Hidetake
    Muraishi, Hiroshi
    Matsuzawa, Hiroki
    Inoue, Toshiyuki
    Nakajima, Yasuo
    Satoh, Hitoshi
    Abe, Shinji
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 103 - 107
  • [3] Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke
    Hidetake Hara
    Hiroshi Muraishi
    Hiroki Matsuzawa
    Toshiyuki Inoue
    Yasuo Nakajima
    Hitoshi Satoh
    Shinji Abe
    Journal of the Korean Physical Society, 2015, 67 : 103 - 107
  • [4] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Nakhostin, Dominik
    Sartoretti, Thomas
    Eberhard, Matthias
    Krauss, Bernhard
    Mueller, Daniel
    Alkadhi, Hatem
    Euler, Andre
    ABDOMINAL RADIOLOGY, 2021, 46 (05) : 2079 - 2089
  • [5] Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography
    Alikhani, Babak
    Werner, Martin
    Jamali, Leila
    Wacker, Frank
    Werncke, Thomas
    ACADEMIC RADIOLOGY, 2019, 26 (08) : 1095 - 1101
  • [6] Attenuation-based characterization of coronary atherosclerotic plaque: Comparison of dual source and dual energy CT with single-source CT and histopathology
    Henzler, Thomas
    Porubsky, Stefan
    Kayed, Hany
    Harder, Nils
    Krissak, U. Radko
    Meyer, Mathias
    Sueselbeck, Tim
    Marx, Alexander
    Michaely, Henrik
    Schoepf, U. Joseph
    Schoenberg, Stefan O.
    Fink, Christian
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (01) : 54 - 59
  • [7] Diagnosis of Pulmonary Artery Embolism: Comparison of Single-Source CT and 3rd Generation Dual-Source CT using a Dual-Energy Protocol Regarding Image Quality and Radiation Dose
    Petritsch, Bernhard
    Kosmala, Aleksander
    Gassenmaier, Tobias
    Weng, Andreas Max
    Veldhoen, Simon
    Kunz, Andreas Steven
    Bley, Thorsten Alexander
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2017, 189 (06): : 527 - 536
  • [8] Image Quality of Virtual Monochromatic Reconstructions of Noncontrast CT on a Dual-Source CT Scanner in Adult Patients
    van Ommen, Fasco
    Kauw, Frans
    Bennink, Edwin
    Heit, Jeremy J.
    Wolman, Dylan N.
    Dankbaar, Jan Willem
    de Jong, Hugo W. A. M.
    Wintermark, Max
    ACADEMIC RADIOLOGY, 2021, 28 (10) : E323 - E330
  • [9] Tube Current Modulation Between Single- and Dual-Energy CT With a Second-Generation Dual-Source Scanner: Radiation Dose and Image Quality
    Matsubara, Kosuke
    Takata, Tadanori
    Kobayashi, Masanao
    Kobayashi, Satoshi
    Koshida, Kichiro
    Gabata, Toshifumi
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 354 - 361
  • [10] Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT
    Ferrero, Andrea
    Chen, Baiyu
    Li, Zhoubo
    Yu, Lifeng
    McCollough, Cynthia
    MEDICAL PHYSICS, 2017, 44 (05) : 1655 - 1660