Real-time monitoring in managing the decontamination of flexible gastrointestinal endoscopes

被引:47
作者
Obee, PC
Griffith, CJ
Cooper, RA
Cooke, RP
Bennion, NE
Lewis, M
机构
[1] Cardiff Univ, Inst Cardiff, Sch Appl Sci, Cardiff, Wales
[2] E Sussex Hlth Trust, Hellingly, England
[3] Biotrace Ltd, Bridgend, Wales
[4] Bro Taf Hlth Author, Cardiff, S Glam, Wales
关键词
D O I
10.1016/j.ajic.2004.07.008
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Objectives: To evaluate the efficacy of endoscope reprocessing using bacteriologic methods and adenosine triphosphate (ATP) bioluminescence and to evaluate the potential benefits of the latter in managing the process. Setting: An endoscopy unit in each of 2 acute district general hospitals in the United Kingdom. Methods: Following visual observation, 8 locations important in the decontamination of gastrointestinal (GI) endoscopes were monitored during reprocessing by bacteriologic sampling (paddle method) and ATP bioluminescence. Results: Endoscope reprocessing was conducted in accordance with British Society of Gastroenterology guidelines. Bacteriologic tests required 24 hours for results to be obtained, ATP tests were completed within 2 minutes. Of the 504 sites tested throughout the decontamination process, bacterial growth was recorded on 32 occasions, 3 from the end product. Throughout the decontamination process, ATP tests obtained 95 results above benchmark values and I from the end product. Conclusions: The bacteriologic results after terminal disinfection indicated that the reprocessing was carried out effectively overall. ATP results prior to disinfection indicated that manual cleaning might not have always been performed adequately and that this could have impaired the disinfection process. ATP provided a rapid means of assessing the efficacy of cleaning steps prior to terminal disinfection.
引用
收藏
页码:202 / 206
页数:5
相关论文
共 26 条
[1]  
Alvarado CJ, 2000, AM J INFECT CONTROL, V28, P138
[2]  
[Anonymous], 2003, GASTROINTEST ENDOSC, V58, P1
[3]  
[Anonymous], 1991, DISINFECTION STERILI
[4]  
[Anonymous], 2003, BSG GUID DEC EQ GAST
[5]   ENDOSCOPE DECONTAMINATION - WHERE DO WE GO FROM HERE [J].
BABB, JR ;
BRADLEY, CR .
JOURNAL OF HOSPITAL INFECTION, 1995, 30 :543-551
[6]   Patient-to-patient transmission of hepatitis C virus during colonoscopy [J].
Bronowicki, JP ;
Venard, V ;
Botte, C ;
Monhoven, N ;
Gastin, I ;
Chone, L ;
Hudziak, H ;
Rhin, B ;
Delanoe, C ;
LeFaou, A ;
Bigard, MA ;
Gaucher, P .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 337 (04) :237-240
[7]  
Casas JMB, 1999, REV ESP ENFERM DIG, V91, P111
[8]   Bacteria-free water for automatic washer-disinfectors: an impossible dream? [J].
Cooke, RPD ;
Whymant-Morris, A ;
Umasankar, RS ;
Goddard, SV .
JOURNAL OF HOSPITAL INFECTION, 1998, 39 (01) :63-65
[9]   The clinical risks of infection associated with endoscopy [J].
Cowen, AE .
CANADIAN JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2001, 15 (05) :321-331
[10]  
Davidson CA, 1999, LUMINESCENCE, V14, P33, DOI 10.1002/(SICI)1522-7243(199901/02)14:1<33::AID-BIO514>3.0.CO