Fast determination of paraquat residues in water by an optical immunosensor and validation using capillary electrophoresis-ultraviolet detection

被引:83
作者
Mallat, E
Barzen, C
Abuknesha, R
Gauglitz, G
Barceló, D
机构
[1] CSIC, IIQAB, Dept Environm Chem, ES-08034 Barcelona, Spain
[2] Univ Tubingen, Inst Phys Chem, D-72076 Tubingen, Germany
[3] Kings Coll London, London W8 7AH, England
关键词
optical immunosensor; paraquat residues; capillary zone electrophoresis-ultraviolet detection;
D O I
10.1016/S0003-2670(00)01016-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A solid-phase fluoroimmunoassay combined with an optical transducer chemically modified with an analyte derivative coupled to a FIA system was used [A. Brecht, A. Klotz, C. Barzen, G. Gauglitz, R. Harris, G. Quigley, J. Wilkinson, P. Sztajnbok, R. Abuknesha, J. Gascon, A. Oubina, D. Barcelo, Anal. Chim. Acta 362 (1998) 69]. Excitation and collection of fluorescence from fluorescently labelled anti-paraquat antibodies locally bound at the planar interface allows the measurement of the fluorescent signal which is indirectly related to the paraquat concentration of the sample. Matrix effects on the immunosensor response were observed, thus leading to the following detection limits, 0.01 and 0.06 mug l(-1), when analyzing paraquat in MilliQ and in river water, respectively, The validation of the biosensor was carried out analyzing paraquat samples by capillary zone electrophoresis with ultraviolet detection (CZE-UV) at 214 nm. Preconcentration of the samples prior to their injection in the capillary electrophoresis were performed using the automated solid-phase extraction system (ASPEC-XL). Paraquat samples were adjusted at pH 9 and were percolated through a silica cartridge, subsequent elution was carried out using a mixture of hydrochloric acid and methanol and afterwards samples were evaporated and injected. This optical immunosensor, also called River ANAlyzer (RIANA) with an optical immunosensor can be applied for fast monitoring of paraquat in river water samples being the total analysis time of 15 min. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:165 / 171
页数:7
相关论文
共 24 条
[1]  
Barcelo D, 1993, OFFICIAL METHODS ANA, P149
[2]   Optical immunoprobe development for multiresidue monitoring in water [J].
Brecht, A ;
Klotz, A ;
Barzen, C ;
Gauglitz, G ;
Harris, RD ;
Quigley, GR ;
Wilkinson, JS ;
Sztajnbok, P ;
Abuknesha, R ;
Gascón, J ;
Oubiña, A ;
Barceló, D .
ANALYTICA CHIMICA ACTA, 1998, 362 (01) :69-79
[3]  
CAI J, 1992, J LIQ CHROMATOGR, V223, P1193
[4]   Spatial and temporal trends of paraquat, diquat, and difenzoquat contamination in water from marsh areas of the Valencian community (Spain) [J].
Fernandez, M ;
Ibanez, M ;
Pico, Y ;
Manes, J .
ARCHIVES OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 1998, 35 (03) :377-384
[5]   Separation of quaternary ammonium herbicides by capillary electrophoresis with indirect UV detection [J].
Galceran, MT ;
Carneiro, MC ;
Diez, M ;
Puignou, L .
JOURNAL OF CHROMATOGRAPHY A, 1997, 782 (02) :289-295
[6]   CAPILLARY ELECTROPHORESIS OF QUATERNARY AMMONIUM ION HERBICIDES - PARAQUAT, DIQUAT AND DIFENZOQUAT [J].
GALCERAN, MT ;
CARNEIRO, MC ;
PUIGNOU, L .
CHROMATOGRAPHIA, 1994, 39 (9-10) :581-586
[7]   Strengths and limitations of immunoassays for effective and efficient use for pesticide analysis in water samples: A review [J].
Hennion, MC ;
Barcelo, D .
ANALYTICA CHIMICA ACTA, 1998, 362 (01) :3-34
[8]   On-line liquid chromatographic trace enrichment and highperformance liquid chromatographic determination of diquat, paraquat and difenzoquat in water [J].
Ibanez, M ;
Pico, Y ;
Manes, J .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :325-331
[9]   Influence of organic matter and surfactants on solid-phase extraction of diquat, paraquat and difenzoquat from waters [J].
Ibanez, M ;
Pico, Y ;
Manes, J .
JOURNAL OF CHROMATOGRAPHY A, 1996, 727 (02) :245-252
[10]   River analyzer for chlorotriazines with a direct optical immunosensor [J].
Mallat, E ;
Barzen, C ;
Klotz, A ;
Brecht, A ;
Gauglitz, G ;
Barcelo, D .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (06) :965-971