Ultrahigh-resolution imaging of human donor cornea using full-field optical coherence tomography

被引:38
作者
Akiba, Masahiro
Maeda, Naoyuki
Yumikake, Kazuhiko
Soma, Takeshi
Nishida, Kohji
Tano, Yasuo
Chan, Kin Pui
机构
[1] TOPCON Adv Biomed Imaging Lab, Paramus, NJ 07652 USA
[2] Yamagata Promot Org Ind Technol, Yamagata, Japan
[3] Osaka Univ, Sch Med, Dept Ophthalmol & Visual Sci, Suita, Osaka, Japan
[4] Tohoku Univ, Grad Sch Med, Dept Ophthalmol & Visual Sci, Sendai, Miyagi, Japan
关键词
optical coherence tomography; full-field; cornea; keratocyte; corneal epithelium;
D O I
10.1117/1.2764461
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A feasibility study of ultrahigh-resolution full-field optical coherence tomography (FF-OCT) for a subcellular-level imaging of human donor corneas is presented. The FF-OCT system employed in this experiment is based on a white light interference microscope, where the sample is illuminated by a thermal light source and a horizontal cross-sectional (en face) image is detected using a charge coupled device (CCD) camera. A conventional four-frame phase-shift detection technique is employed to extract the interferometric image from the CCD output. A 95-nm-broadband full-field illumination yields an axial resolution of 2.0 mu m, and the system covers an area of 850 mu m x 850 mu m with a transverse resolution of 2.4 mu m using a 0.3-NA microscope objective and a CCD camera with 512 x 512 pixels. Starting a measurement from the epithelial to the endothelial side, a series of en face images was obtained. From detected en face images, the epithelial cells, Bowman's layer, stromal keratocyte, nerve fiber, Descemet's membrane, and endothelial cell were clearly observed. Keratocyte cytoplasm, its nuclei, and its processes were also separately detected. Two-dimensional interconnectivity of the keratocytes is visualized, and the keratocytes existing between collagen lamellaes are separately extracted by exploiting a high axial resolution ability of FF-OCT. (C) 2007 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页数:7
相关论文
共 34 条
[1]   Full-field optical coherence tomography by two-dimensional heterodyne detection with a pair of CCD cameras [J].
Akiba, M ;
Chan, KP ;
Tanno, N .
OPTICS LETTERS, 2003, 28 (10) :816-818
[2]  
AKIBA M, 2007, 642958 SPIE
[3]   Confocal microscopy of the cornea [J].
Böhnke, M ;
Masters, BR .
PROGRESS IN RETINAL AND EYE RESEARCH, 1999, 18 (05) :553-628
[4]   SPECULAR MICROSCOPY OF HUMAN CORNEAL ENDOTHELIUM INVIVO [J].
BOURNE, WM ;
KAUFMAN, HE .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1976, 81 (03) :319-323
[5]  
Boyde A, 1986, CORNEA, V4, P205
[6]  
CAVANAGH HD, 1993, OPHTHALMOLOGY, V100, P1444
[7]   Spectral domain optical coherence tomography - Ultra-high speed, ultra-high resolution ophthalmic imaging [J].
Chen, TC ;
Cense, B ;
Pierce, MC ;
Nassif, N ;
Park, BH ;
Yun, SH ;
White, BR ;
Bouma, BE ;
Tearney, GJ ;
de Boer, JF .
ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (12) :1715-1720
[8]   Swept source optical coherence tomography using an all-fiber 1300-nm ring laser source [J].
Choma, MA ;
Hsu, K ;
Izatt, JA .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (04)
[9]   Ultrahigh-resolution ophthalmic optical coherence tomography [J].
Drexler, W ;
Morgner, U ;
Ghanta, RK ;
Kärtner, FX ;
Schuman, JS ;
Fujimoto, JG .
NATURE MEDICINE, 2001, 7 (04) :502-507
[10]   Ultrahigh-resolution optical coherence tomography [J].
Drexler, W .
JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) :47-74