Characterization of heparin aerosols generated in jet and ultrasonic nebulizers

被引:13
作者
Bendstrup, KE [1 ]
Newhouse, MT [1 ]
Pedersen, OF [1 ]
Jensen, JI [1 ]
机构
[1] Esbjerg Hosp, Dept Internal Med, DK-6700 Esbjerg, Denmark
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 1999年 / 12卷 / 01期
关键词
heparin; aerosol; nebulizers; viscosity; surface tension; particle characterization;
D O I
10.1089/jam.1999.12.17
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Inhaled heparin has been used for asthma treatment, but results have been inconsistent, probably due to highly varying lung doses. We determined the output per unit time and the particle size distributions of sodium heparin, calcium heparin, and low molecular weight (LMW) heparin formulations in five concentrations from Sidestream jet nebulizers (Medic-Aid, Bognor Regis, England) and an Ultraneb 2000 ultrasonic nebulizer (DeVilbiss, Langen, Germany). We also determined the inhaled mass and the estimated respirable mass for some combinations. For the jet nebulizer, output per minute increased with increasing concentration and flow rate, and particle size decreased from 3.64 to 2.01 mu m (mass median diameter [MMD]). The percentage of particles less than 3 mu m ranged from 41% to 74%. For the ultrasonic nebulizer, maximum output per minute was achieved at a concentration of 7000 IU/mL; this maximum depended upon the viscosity and temperature of the solution. MMD was independent of formulation, temperature, or concentration and ranged from 5.61 to 7.03 mu m. Sodium heparin/calcium heparin in a concentration of 20,000 IU/mL in the jet nebulizer driven at 10 L/min produced the highest dose of heparin capable of reaching the lower respiratory tract. Mass balance was determined for these combinations with the jet nebulizer run until visible aerosol generation ceased. Of a loading dose of 80,000 IU of heparin, 45,000 IU remained in the dead space of the nebulizer, 20,000 IU was deposited on the exhalation filter, and 15,000 IU was captured on the inhalation filter (inhaled mass). This corresponds to a respirable mass of 10,000 IU of heparin with a high probability of reaching the lower respiratory tract in normal healthy adults.
引用
收藏
页码:17 / 25
页数:9
相关论文
共 17 条
  • [1] Ceyhan BB, 1997, INT J CLIN PHARM TH, V35, P208
  • [2] *COUNC EUR, 1996, EUR PHARM
  • [3] CRIMI N, 1995, AM J RESP CRIT CARE, V151, pA130
  • [5] JET AND ULTRASONIC NEBULIZER OUTPUT - USE OF A NEW METHOD FOR DIRECT MEASUREMENT OF AEROSOL OUTPUT
    DENNIS, JH
    STENTON, SC
    BEACH, JR
    AVERY, AJ
    WALTERS, EH
    HENDRICK, DJ
    [J]. THORAX, 1990, 45 (10) : 728 - 732
  • [6] Effect of inhaled heparin on allergen-induced early and late asthmatic responses in patients with atopic asthma
    Diamant, Z
    Timmers, MC
    vanderVeen, H
    Page, CP
    vanderMeer, FJ
    Sterk, PJ
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 153 (06) : 1790 - 1795
  • [7] FERRON GA, 1976, AM REV RESPIR DIS, V114, P899
  • [8] Time course of the protective effect of inhaled heparin on exercise-induced asthma
    Garrigo, J
    Danta, I
    Ahmed, T
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 153 (05) : 1702 - 1707
  • [9] Effect of pretreatment with heparin on pulmonary and cutaneous response
    Kalpaklioglu, AF
    Demirel, YS
    Saryal, S
    Misirligil, Z
    [J]. JOURNAL OF ASTHMA, 1997, 34 (04) : 337 - 343
  • [10] KUNA P, 1995, ALLERGY S26, V50, P25