Effects of joint alignment and type on mechanical properties of thermoplastic articulated ankle-foot orthosis

被引:28
作者
Gao, Fan [1 ]
Carlton, William [1 ]
Kapp, Susan [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Prosthet & Orthot Program, Dept Hlth Care Sci, Sch Hlth Profess, Dallas, TX 75390 USA
关键词
Ankle-foot orthosis; misalignment; dorsiflexion assist joint; VISCOELASTIC PROPERTIES; STIFFNESS CONTROL; GAIT; AFOS; FLEXIBILITY; APPARATUS; TRIMLINE; ASSIST; IMPACT; MOMENT;
D O I
10.1177/0309364611409617
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Articulated or hinged ankle-foot orthosis (AFO) allow more range of motion. However, quantitative investigation on articulated AFO is still sparse. Objective: The objective of the study was to quantitatively investigate effects of alignment and joint types on mechanical properties of the thermoplastic articulated AFO. Study design: Tamarack dorsiflexion assist flexure joints with three durometers (75, 85 and 95) and free motion joint were tested. The AFO joint was aligned with the center of the motor shaft (surrogate ankle joint), 10 mm superior, inferior, anterior and posterior with respect to the motor shaft center. Methods: The AFO was passively moved from 20 degrees plantar flexion to 15 degrees dorsiflexion at a speed of 10 degrees/s using a motorized device. Mechanical properties including index of hysteresis, passive resistance torque and quasi-static stiffness (at neutral, 5 degrees, 10 degrees and 15 degrees in plantar flexion) were quantified. Results: Significant effects of joint types and joint alignment on the mechanical properties of an articulated thermoplastic AFO were revealed. Specifically, center alignment showed minimum resistance and stiffness while anterior and posterior alignment showed significantly higher resistance and stiffness. The dorsiflexion assist torques at neutral position ranged from 0.69 +/- 0.09 to 1.88 +/- 0.10 Nm. Conclusions: Anterior and posterior alignment should be avoided as much as possible.
引用
收藏
页码:181 / 189
页数:9
相关论文
共 39 条
[1]  
Alimusaj M, 2007, ORTHOPADE, V36, P752, DOI 10.1007/s00132-007-1125-0
[2]  
[Anonymous], 1992, GAIT ANAL NORMAL PAT
[3]   The effect of hinged ankle-foot orthosis on gait and energy expenditure in spastic hemiplegic cerebral palsy [J].
Balaban, Birol ;
Yasar, Evren ;
Dal, Ugur ;
Yazicioglu, Kamil ;
Mohur, Haydar ;
Kalyon, Tunc Alp .
DISABILITY AND REHABILITATION, 2007, 29 (02) :139-144
[4]  
Bielby S., 2010, JPO, V22, P204, DOI [10.1097/JPO.0b013-3181f9082e, DOI 10.1097/JPO.0B013E3181F9082E]
[5]   Articulated external fixation of the ankle: minimizing motion resistance by accurate axis alignment [J].
Bottlang, M ;
Marsh, JL ;
Brown, TD .
JOURNAL OF BIOMECHANICS, 1999, 32 (01) :63-70
[6]   A new method for evaluating ankle foot orthosis characteristics: BRUCE [J].
Bregman, D. J. J. ;
Rozumalski, A. ;
Koops, D. ;
de Groot, V. ;
Schwartz, M. ;
Harlaar, J. .
GAIT & POSTURE, 2009, 30 (02) :144-149
[7]   A continuous loading apparatus for measuring three-dimensional stiffness of ankle-foot orthoses [J].
Cappa, P ;
Patanè, F ;
Di Rosa, G .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (06) :1025-1029
[8]   A novel device to evaluate the stiffness of ankle-foot orthosis devices [J].
Cappa, P ;
Patanè, F ;
Pierro, MM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (06) :913-917
[9]   3-DIMENSIONAL FINITE-ELEMENT STRESS-ANALYSIS OF THE POLYPROPYLENE, ANKLE-FOOT ORTHOSIS - STATIC ANALYSIS [J].
CHU, TM ;
REDDY, NP ;
PADOVAN, J .
MEDICAL ENGINEERING & PHYSICS, 1995, 17 (05) :372-379
[10]  
Chu TM, 1995, J REHABIL RES DEV, V32, P349