Long-term wear failure analysis of uhmwpe acetabular cup in total hip replacement

被引:24
作者
Shahemi, N. [1 ]
Liza, S. [1 ]
Abbas, A. A. [2 ]
Merican, A. M. [2 ]
机构
[1] Univ Teknol Malaysia, MJIIT, TriPreM I Kohea, Dept Mech Precis Engn, Kuala Lumpur 59100, Malaysia
[2] Univ Malaya, Fac Med, Natl Orthopaed Ctr Excellence Res & Learning NOCE, Dept Orthopaed Surg, Kuala Lumpur 50603, Malaysia
关键词
UHMWPE; Wear; Total hip replacement; Oxidative degradation; Fatigue; JOINT REPLACEMENT; ARTHROPLASTY; MECHANISM; LINERS;
D O I
10.1016/j.jmbbm.2018.07.017
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A revision of a metal-on-ultra high molecular weight (UHMWPE) bearing couple for total hip replacement was performed due to aseptic loosening after 23 years in vivo. It is a major long-term failure identified from wear generation. This study includes performing failure analysis of retrieved polyethylene acetabular cup from Zimmer Trilogy Acetabular system. The UHMWPE acetabular cup was retrieved from a 61 years old male patient with ability to walk but limited leg movement when he presented to hospital in early 2016 with complaint left thigh pain. It was 23 years after his primary total hip replacement procedure. Surface roughness and morphology condition were measured using 3D laser microscope and Scanning Electron Microscope (SEM) to evaluate and characterize the wear features on polyethylene acetabular cup surface. ATR-Fourier Transform Infra-Red (ATR-FTIR), differential scanning calorimetry (DSC) and gel permeation chromatography (GPC) were used to characterize the chemical composition of carbon-oxygen bonding, crystallinity percentage and molecular weight of the polymer liner that might changes the mechanical properties of polyethylene. Nano indentation is to measure hardness and elasticity modulus where the ratio of hardness to elastic modulus value can be reflected as the degradation of mechanical properties. A prominent difference of thickness between two regions resulted from acentric loading concentration was observed and wear rate were measured. The linear wear rate for thin side and thick side were 0.33 mm/year and 0.05 mm/year respectively. Molecular weight reduction of 57.5% and relatively low ratio of hardness to elastic modulus (3.59 x 10(-3)) were the indicator of major mechanical properties degradation happened on UHMWPE acetabular cup. This major degradation was contributed by oxidation and polishing wear feature accompanied with delamination, craters, ripple and cracks were the indication of extensive usage of UHMWPE from the suggested life span of acetabular cup application.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 25 条
[1]   Tribology and total hip joint replacement: Current concepts in mechanical simulation [J].
Affatato, S. ;
Spinelli, A. ;
Zavalloni, M. ;
Mazzega-Fabbro, C. ;
Viceconti, A. .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (10) :1305-1317
[2]   Position of the acetabular component determines the fate of femoral head autografts in total hip replacement for acetabular dysplasia [J].
Atilla, B. ;
Ali, H. ;
Aksoy, M. C. ;
Caglar, O. ;
Tokgozoglu, A. M. ;
Alpaslan, M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (07) :874-878
[3]   Cracks, microcracks and fracture in polymer structures: Formation, detection, autonomic repair [J].
Awaja, Firas ;
Zhang, Shengnan ;
Tripathi, Manoj ;
Nikiforov, Anton ;
Pugno, Nicola .
PROGRESS IN MATERIALS SCIENCE, 2016, 83 :536-573
[4]   Wear Performance of UHMWPE and Reinforced UHMWPE Composites in Arthroplasty Applications: A Review [J].
Baena, Juan C. ;
Wu, Jingping ;
Peng, Zhongxiao .
LUBRICANTS, 2015, 3 (02) :413-436
[5]  
Bayer R.G., 1995, STRAIN, V31, P115
[6]   Failure analysis on retrieved ultra high molecular weight polyethylene (UHMWPE) acetabular cups [J].
Burger, N. D. L. ;
de Vaal, P. L. ;
Meyer, J. P. .
ENGINEERING FAILURE ANALYSIS, 2007, 14 (07) :1329-1345
[7]   Mechanical wear and oxidative degradation analysis of retrieved ultra high molecular weight polyethylene acetabular cups [J].
Choudhury, Dipankar ;
Ranusa, Matus ;
Fleming, Robert A. ;
Vrbka, Martin ;
Krupka, Ivan ;
Teeter, Matthew G. ;
Goss, Josh ;
Zou, Min .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 79 :314-323
[8]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[9]   Surface topography of retrieved PCA acetabular liners: proposal of a novel wear mechanism [J].
Elfick, AP ;
Hall, RM ;
Unsworth, A ;
Pinder, IM .
JOURNAL OF MATERIALS SCIENCE LETTERS, 1998, 17 (13) :1085-1088
[10]   A novel approach to model effects of surface roughness parameters on wear [J].
Ghosh, Arnab ;
Sadeghi, Farshid .
WEAR, 2015, 338 :73-94