Clinical safety of magnetic resonance imaging in patients with implanted SynchroMed EL infusion pumps

被引:7
作者
Diehn, Felix E. [1 ]
Wood, Christopher P. [1 ]
Watson, Robert E., Jr. [1 ]
Mauck, William D. [2 ]
Burke, Michelle M. [3 ]
Hunt, Christopher H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Anesthesiol, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Psychiat & Psychol, Rochester, MN 55905 USA
关键词
MRI; Safety; Spinal infusion pumps; Medtronic; SynchroMed; CHRONIC PAIN; TECHNICAL COMPLICATIONS; INTRATHECAL BACLOFEN; NEURAXIAL INFUSION; CANCER-PATIENTS; EFFICACY; MORPHINE; THERAPY; DEVICES; ORIGIN;
D O I
10.1007/s00234-010-0737-y
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Patients with implanted SynchroMed spinal infusion pumps (Medtronic, Inc., Minneapolis, MN) routinely undergo magnetic resonance imaging at our institution. In August 2008, Medtronic issued an urgent medical device correction report regarding several pumps. Because of the rare potential "for a delay in the return of proper drug infusion" and "for a delay in the logging of motor stall events," "a patient's pump must be interrogated after MRI exposure in order to confirm proper pump functionality." This is particularly important in patients receiving intrathecal baclofen, for whom a delay in return of proper pump infusion could lead to life-threatening baclofen withdrawal syndrome. The objective of this report is to present our experience and protocol of performing magnetic resonance imaging in patients with implanted SynchroMed EL pumps. We retrospectively reviewed records of 86 patients with implanted SynchroMed EL spinal infusion pumps who underwent 112 examinations on 1.5-T magnetic resonance imaging scanners from September 1, 1998 to July 7, 2004. No SynchroMed EL pumps were damaged by magnetic resonance imaging, and the programmable settings remained unchanged in all patients. Our data suggest that SynchroMed EL pump malfunction is indeed rare after routine clinical 1.5-T magnetic resonance imaging examinations. However, based on the Medtronic correction report, we perform pump interrogation before and after imaging.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 21 条
[11]   Spinal opioid infusions in the treatment of chronic pain of nonmalignant origin [J].
Maron, J ;
Loeser, JD .
CLINICAL JOURNAL OF PAIN, 1996, 12 (03) :174-179
[12]  
*MEDTR INC, 2000, EFF MAGN RES IM MRI
[13]  
*MEDTR INC, 2008, URG MED DEV CORR IMP
[14]  
ONOFRIO BM, 1981, MAYO CLIN PROC, V56, P516
[15]   INTRATHECAL BACLOFEN FOR SEVERE SPINAL SPASTICITY [J].
PENN, RD ;
SAVOY, SM ;
CORCOS, D ;
LATASH, M ;
GOTTLIEB, G ;
PARKE, B ;
KROIN, JS .
NEW ENGLAND JOURNAL OF MEDICINE, 1989, 320 (23) :1517-1521
[16]  
POWERS C, 2000, CURR ANESTHESIOL REP, V2, P128
[17]  
Schueler BA, 1999, JMRI-J MAGN RESON IM, V9, P596, DOI 10.1002/(SICI)1522-2586(199904)9:4<596::AID-JMRI14>3.3.CO
[18]  
2-K
[19]  
VONROEMELING R, 1991, JMRI-J MAGN RESON IM, V1, P77
[20]  
WALLACE M, 1999, SPINAL DRUG DELIVERY